Solus G25670Recalls & reported problems

Cook Ireland Ltd · Solus, Double Pigtail Stent with Introducer · Class 2 · Implant

All models of Solus (7)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
4
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 10827002256702), received between Feb 28, 2018 and Dec 1, 2025 — classified as 1 injury, 3 malfunctions.

The most frequently reported specific device problem is “Difficult to Advance” (2 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 1 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 1 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Difficult to Advance2
  • Failure to Advance2
  • Device Markings/Labelling Problem1

Patient problems

  • No Clinical Signs, Symptoms or Conditions2
  • No Known Impact Or Consequence To Patient2

Recent adverse event reports

The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.

  • MalfunctionDec 1, 2025Failure to Advance; Failure to Advance

    SUPPLEMENTAL REPORT IS BEING SUBMITTED DUE TO THE COMPLETION OF THE INVESTIGATION ON 23-JAN-2026.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23689591 ↗
  • InjuryDec 15, 2023Failure to Advance; Difficult to Advance; Failure to Advance; Difficult to Advance

    ON (B)(6) 2023 USER WAS CONDUCTING GASTROINTESTINAL ENDOSCOPY EXAMINATION AND TREATMENT. USER TOOK 10FR STENT WITH 3.7MM ENDOSCOPE TO CONDUCT BILE DRAINAGE ACCORDING TO IFU INSTRUCTION. DURING THE PROCEDURE, USER ADVANCED THE STENT INTO ENDOSCOPE WORKING CHANNEL SUCCESSFULLY , BUT CANNOT ADVANCED THE STENT THROUGH THE DISTAL END OF WORKING CHANNEL, AND THE STENT PUSHER WAS STUCK INSIDE THE ENDOSCOPE WORKING CHANNEL AND CANNOT MOVE FORWARD. USER HAS 2 CHOICES AT THE MOMENT, ONE IS CHANGE A SMALLER SIZE OF STENT (7FR), OR CHANGED A BIGGER DIAMETER ENDOSCOPE. BUT USER CONSIDER THE 3.7MM ENDOSCOPE IS MORE CONVENIENT TO PROCEED THE PROCEDURE, AND CHANGED ENDOSCOPE REQUIRES MORE TIME AND MORE COMPLICATED, USER DECIDED TO CHANGE THE SENT TO CONTINUE THE PROCEDURE. THIS KIND OF PROCEDURE NORMALLY COMPLETED IN 2 HOURS, BUT THIS PROCEDURE WAS COMPLETED USING 7 HOURS. USER OBSERVED THE PATIENT OUTCOME THROUGH THE PROCEDURE WHICH IS WELL. AND THE PATIENT RECOVERS WELL AFTER PROCEDURE, NO HARM TO PATIENT. USER REPORTED THE AE TO NATIONAL AE SYSTEM AS IT IS PROLONG THE PROCEDURE SERIOUSLY. PATIENT OUTCOME: THE PROCEDURE PROLONG 5 HOURS THAN USUAL. PATIENT IS WELL DURING PROCEDURE AND RECOVERS WE

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18335200 ↗
  • MalfunctionOct 31, 2019Difficult to Advance; Difficult to Advance

    FDA MDR REPORTING REQUIRED - EVENT IS FDA MDR REPORTABLE BASED ON THE DEVICE MALFUNCTION REPORTING PRECEDENCE FOR THIS DEVICE FAMILY FOR THE ISSUE OF ¿DIFFICULT ADVANCEMENT'. AS REPORTED TO CUSTOMER RELATIONS VIA PHONE CONVERSATION "THE DEVICE FAILED TO DEPLOY WHEN THE PHYSICIAN TRIED TO DEPLOY IT. A SECOND G25670-ZSS-10-3-RB WAS OPENED AND USED TO COMPLETED THE PROCEDURE SUCCESSFULLY."

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 9260394 ↗
  • MalfunctionFeb 28, 2018Device Markings/Labelling Problem; Device Markings/Labelling Problem

    THIS MDR IS BEING SUBMITTED AS A CORRECTION, BASED ON ADDITIONAL INFORMATION AND CLINICAL INPUT RECEIVED THERE WAS NO HARM TO THE PATIENT. THIS REPORT HAS BEEN RE-CATEGORIZED AS A MALFUNCTION REPORT ONLY. CONFIRMATION WAS RECEIVED THAT THE MARKERS FELL INTO THE STOMACH AND WERE EXCRETED FROM THE BODY NATURALLY. CLINICAL INPUT WAS RECEIVED (B)(6) 2018: WOULD YOU BE ABLE TO GIVE YOUR CLINICAL INPUT ON THE POTENTIAL HARM TO THE PATIENT BY EXCRETING THE RADIOPAQUE MARKERS? THERE IS NO HARM IF THE MARKERS WERE EXPELLED NATURALLY. AS REPORTED TO CUSTOMER RELATIONS, "THE RADIOPAQUE MARKERS CAME OFF THE SOLUS STENT DURING INSERTION. THE PHYSICIAN LEFT THE STENT IN PLACE AND THE MARKERS FELL INSIDE THE STOMACH. THE PHYSICIAN SPRAYED THE INTRODUCER WITH SILICONE BEFORE HAND."

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 7303289 ↗

Other CATHETER, BILIARY, DIAGNOSTIC devices

All CATHETER, BILIARY, DIAGNOSTIC devices →

Sources and updates