Solus G25673Recalls & reported problems
Cook Ireland Ltd · Solus, Double Pigtail Stent with Introducer · Class 2 · Implant
Data updated Oct 13, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 5
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 5 adverse event reports that list this exact device (DI 10827002256733), received between Feb 21, 2020 and Mar 24, 2026 — classified as 5 malfunctions.
The most frequently reported specific device problem is “Improper or Incorrect Procedure or Method” (3 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Reports received per month
Most reported problems
Device problems
- Improper or Incorrect Procedure or Method3
- Difficult to Advance2
- Physical Resistance/Sticking2
- Deformation Due to Compressive Stress1
- Failure to Advance1
- Material Twisted/Bent1
Patient problems
- No Clinical Signs, Symptoms or Conditions4
- No Consequences Or Impact To Patient2
- No Known Impact Or Consequence To Patient1
Recent adverse event reports
The 5 most recent of 5 reports, in the reporter's own words as published by the FDA.
MalfunctionMar 24, 2026Improper or Incorrect Procedure or Method; Physical Resistance/Sticking
WHEN PHYSICIAN PUSH STENT OVER THE STENOSIS, HE WANTED TO PULL BACK THE INNER CATHETER. HE FOUND THAT HE COULDN'T PULL BACK HE INNER CATHETER, HE USED FORCE TO IT AND LED THE CATHETER BROKEN. FINALLY, HE CHANGED TO COMPETITOR'S PLASTIC STENT, THEN IT'S OKAY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24672610 ↗MalfunctionJun 16, 2022Failure to Advance; Physical Resistance/Sticking; Failure to Advance; Physical Resistance/Sticking
USER DETECTED THE PUSHER CANNOT BE PUT THROUGH STENT AT PREPARATION STAGE. NO USE ON PATIENT. N/A ¿ THIS OBSERVATION WAS MADE PRIOR TO PATIENT CONTACT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 14712492 ↗MalfunctionFeb 25, 2021Improper or Incorrect Procedure or Method; Difficult to Advance; Improper or Incorrect Procedure or Method; Difficult to Advance
SUPPLEMENTAL FOLLOW-UP REPORT IS BEING SUBMITTED DUE TO THE COMPLETION OF THE INVESTIGATION AND AN UPDATE TO THE INVESTIGATION CONCLUSIONS ON THE (B)(6) 2022.
Patient problems: No Clinical Signs, Symptoms or Conditions; No Consequences Or Impact To Patient
MAUDE report 11378767 ↗MalfunctionApr 6, 2020Deformation Due to Compressive Stress; Difficult to Advance; Deformation Due to Compressive Stress; Difficult to Advance
AS REPORTED TO CUSTOMER RELATIONS: "2 EA DEVICES, DURING THE SAME PROCEDURE, HAD THE SAME FAILURE MODE. BOTH MADE PATIENT CONTACT BUT ONLY 1 EA IS AVAILABLE FOR RETURN (OTHER THROWN AWAY). DM EVALUATION OF RETURNABLE PRODUCT IS THAT THE PUSHING CATHETER KINKED DURING STENT ADVANCEMENT, AND WHEN THEY WERE TRYING TO DEPLOY THE GUIDING CATHETER THE WHITE CAP CAME OFF. PROCEDURE WAS COMPLETED WITH UNKNOWN DEVICE. NO ADD'L PROCEDURES REQUIRED, NO ADVERSE EFFECTS EXPERIENCED BY PATIENT, NOTHING REMAINING IN PATIENT, NO PROLONGED HOSPITALIZATION." .1 FOR ALL COMPLAINTS, ASK: 2.1.1 DOES THE COMPLAINT RELATE TO: DEVICE PLACEMENT (THIS ONE). DEVICE REMOVAL. OBSERVATION PRIOR TO PATIENT CONTACT. 2.1.2 PLEASE INDICATE WHERE THE DEVICE WAS STORED PRIOR TO USE. IN THE BOX, IN THE PROCEDURE ROOM IN A CABINET. 2.1.3 WHAT IS THE REORDER NUMBER, OUTER DIAMETER AND LENGTH OF THE WIRE GUIDE THAT WAS USED WITH THIS DEVICE IN THIS PROCEDURE? UNKNOWN, EITHER .035 (BOSTON HYDROJAG) OR .025 (COOK). 2.1.4 WERE PREVIOUS PROCEDURES I.E. SPHINCTEROTOMY ETC. CARRIED OUT PRIOR TO PLACING THE DEVICE OVER THE PARTICULAR WIRE GUIDE? ASKU. IF YES, PLEASE INDICATE THE PROCEDURE PERFORMED. 2.1.5 WAS THE WIRE GUIDE INS
Patient problems: No Consequences Or Impact To Patient
MAUDE report 9929091 ↗MalfunctionFeb 21, 2020Improper or Incorrect Procedure or Method; Material Twisted/Bent; Improper or Incorrect Procedure or Method; Material Twisted/Bent
SUPPLEMENTAL REPORT IS BEING SUBMITTED DUE TO THE COMPLETION OF THE INVESTIGATION.
Patient problems: No Clinical Signs, Symptoms or Conditions; No Known Impact Or Consequence To Patient
MAUDE report 9737810 ↗
Other CATHETER, BILIARY, DIAGNOSTIC devices
- Hurricane™ RX M00545900Boston Scientific Corporation · CATHETER, BILIARY, DIAGNOSTIC2 recalls1 death705 reports
- MUSTANG™ H74939171060470Boston Scientific Corporation · CATHETER, BILIARY, DIAGNOSTIC384 reports
- CRE™ PRO M00558710Boston Scientific Corporation · CATHETER, BILIARY, DIAGNOSTIC1 recall284 reports
- Hurricane™ RX M00545920Boston Scientific Corporation · CATHETER, BILIARY, DIAGNOSTIC2 recalls211 reports
- MUSTANG™ H74939171070470Boston Scientific Corporation · CATHETER, BILIARY, DIAGNOSTIC265 reports
- MUSTANG™ H74939171050470Boston Scientific Corporation · CATHETER, BILIARY, DIAGNOSTIC212 reports
- Hurricane™ RX M00545940Boston Scientific Corporation · CATHETER, BILIARY, DIAGNOSTIC2 recalls109 reports
- MUSTANG™ H74939171100470Boston Scientific Corporation · CATHETER, BILIARY, DIAGNOSTIC203 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 13, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 13, 2026
- Page last rebuilt: Oct 13, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page