Zilver G25730Recalls & reported problems
Cook Incorporated · Zilver Expandable Metal Biliary Stent System · Class 2 · Implant
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 4
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 10827002257303), received between Apr 12, 2016 and Jun 20, 2025 — classified as 1 injury, 3 malfunctions.
The most frequently reported specific device problem is “Malposition of Device” (2 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Malposition of Device2
- Unintended Ejection2
- Difficult to Remove1
- Material Separation1
- Separation Failure1
Patient problems
- No Clinical Signs, Symptoms or Conditions2
- No Consequences Or Impact To Patient2
Recent adverse event reports
The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.
MalfunctionJun 20, 2025Unintended Ejection; Malposition of Device; Unintended Ejection; Malposition of Device
DURING AN UNKNOWN BILE DUCT PROCEDURE, THE PHYSICIAN USED A COOK ZILVER EXPANDABLE METAL BILIARY STENT SYSTEM. IT WAS REPORTED THAT THE USER DETECTED THE INNER CORE DETACHMENT, THEN IMMEDIATELY CHANGED TO A NEW DEVICE. THE PROCEDURE WAS COMPLETED WITH ANOTHER OF THE SAME DEVICE. A SECTION OF THE DEVICE DID NOT REMAIN INSIDE THE PATIENT¿S BODY. THE PATIENT DID NOT REQUIRE ANY ADDITIONAL PROCEDURES DUE TO THIS OCCURRENCE. ACCORDING TO THE INITIAL REPORTER, THE PATIENT DID NOT EXPERIENCE ANY ADVERSE EFFECTS DUE TO THIS OCCURRENCE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22302051 ↗MalfunctionOct 21, 2021Unintended Ejection; Malposition of Device; Unintended Ejection; Malposition of Device
DURING A METAL STENT PLACEMENT PROCEDURE, THE PHYSICIAN USED A COOK ZILVER EXPANDABLE METAL BILIARY STENT SYSTEM. THE STENT SLIPPED OUT OF THE CATHETER AND COULD NOT BE DEPLOYED. A SECTION OF THE DEVICE DID NOT REMAIN INSIDE THE PATIENT¿S BODY. THE PATIENT DID NOT REQUIRE ANY ADDITIONAL PROCEDURES DUE TO THIS OCCURRENCE. ACCORDING TO THE INITIAL REPORTER, THE PATIENT DID NOT EXPERIENCE ANY ADVERSE EFFECTS DUE TO THIS OCCURRENCE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12678084 ↗InjurySep 15, 2020Difficult to Remove; Material Separation; Difficult to Remove; Material Separation
DURING AN ENDOSCOPIC RETROGRADE CHOLANGIOPANCREATOGRAPHY (ERCP), THE PHYSICIAN USED A ZILVER EXPANDABLE METAL BILIARY STENT SYSTEM. THE USER DEPLOYED THE STENT WITHOUT AN ISSUE. THE TIP OF THE INNER SHEATH DETACHED AND REMAINED IN THE PATIENT. ONE PART OF THE BROKEN TIP IS IN THE HILAR BILE DUCT AND THE OTHER PART IS IN THE PATIENT'S STOMACH. ADDITIONAL INFORMATION WAS RECEIVED ON 14-SEPTEMBER-2020 WHICH STATES THAT THE USER REMOVED THE REMAINING PART OF DEVICE BY FORCEPS UNDER AN ADDITIONAL ENDOSCOPIC PROCEDURE. A SECTION OF THE DEVICE DETACHED INITIALLY, BUT ULTIMATELY DID NOT REMAIN INSIDE THE PATIENT¿S BODY. THE DETACHED INNER SHEATH WAS RETRIEVED 4 DAYS AFTER THE PROCEDURE DUE TO THIS OCCURRENCE DURING ANOTHER ENDOSCOPIC PROCEDURE. ACCORDING TO THE INITIAL REPORTER, THE PATIENT DID NOT EXPERIENCE ANY ADVERSE EFFECTS DUE TO THIS OCCURRENCE.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 10538002 ↗MalfunctionApr 12, 2016Separation Failure; Separation Failure
DURING AN ENDOSCOPY PROCEDURE, THE PHYSICIAN USED A COOK ZILVER EXPANDABLE METAL BILIARY STENT SYSTEM. PER THE INFORMATION PROVIDED ON THE COOK COMPLAINT REPORTING FORM: ON THE PROCEDURE, THE STENT CANNOT TO DEPLOY AND RETURN [UNABLE TO DEPLOY]. PER CLARIFICATION RECEIVED ON 03/09/2016: THE STENT CAN DEPLOY ABOVE 3 MM [STENT WILL ONLY DEPLOY 3MM] AND METAL STENT WILL ADHERE THE INTRODUCER CANNOT PUSH TO DEPLOY [STENT CANNOT DEPLOY]. WHEN THE DEVICE WAS RECEIVED FOR EVALUATION ON 03/18/2016, IT WAS NOTED THAT THE STENT WAS DEPLOYED MORE THAN 1 CM. CLARIFICATION WAS REQUESTED. ON 03/22/2016, THE FOLLOWING WAS PROVIDED FROM THE COOK SALES REPRESENTATIVE: THIS DEVICE IS SINGLE USE DEVICE, SO I CANNOT CHECK WHETHER IT CAN DEPLOY [BE DEPLOYED] BEFORE USING. IN THIS CASE, THE DEVICE ENTERED TO THE BODY OF PATIENT AND DEPLOYED MORE THAN 1 CM. HOWEVER, IT COULD NOT DEPLOY COMPLETELY. SO THE DOCTOR TOOK IT OUT FROM THE PATIENT'S BODY AND RETURNED IT TO US. DUE TO THE HYGIENE REASON, THE NURSE HAD CLEANED UP THE DEVICE BEFORE RETURNING. THEREFORE, IT CANNOT SEE ANY BODILY FLUID ON THE DEVICE [BODILY FLUID CANNOT BE SEEN ON THE DEVICE] BUT IT COULD BE CONFIRMED THAT THIS DEVICE WAS USED INSIDE
Patient problems: No Consequences Or Impact To Patient
MAUDE report 5570920 ↗
Other CATHETER, BILIARY, DIAGNOSTIC devices
- Hurricane™ RX M00545900Boston Scientific Corporation · CATHETER, BILIARY, DIAGNOSTIC2 recalls1 death705 reports
- MUSTANG™ H74939171060470Boston Scientific Corporation · CATHETER, BILIARY, DIAGNOSTIC384 reports
- CRE™ PRO M00558710Boston Scientific Corporation · CATHETER, BILIARY, DIAGNOSTIC1 recall284 reports
- Hurricane™ RX M00545920Boston Scientific Corporation · CATHETER, BILIARY, DIAGNOSTIC2 recalls211 reports
- MUSTANG™ H74939171070470Boston Scientific Corporation · CATHETER, BILIARY, DIAGNOSTIC265 reports
- MUSTANG™ H74939171050470Boston Scientific Corporation · CATHETER, BILIARY, DIAGNOSTIC212 reports
- Hurricane™ RX M00545940Boston Scientific Corporation · CATHETER, BILIARY, DIAGNOSTIC2 recalls109 reports
- MUSTANG™ H74939171100470Boston Scientific Corporation · CATHETER, BILIARY, DIAGNOSTIC203 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page