Skip to content

Advance Micro G26098Recalls & reported problems

Cook Incorporated · Advance Micro 14 Ultra Low-Profile PTA Balloon Catheter · Class 2

All models of Advance Micro (25)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
1
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 10827002260983), received between Sep 30, 2024 and Sep 30, 2024 — classified as 1 malfunction.

The most frequently reported specific device problem is “Break” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 1 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Break1

Patient problems

  • No Clinical Signs, Symptoms or Conditions1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • MalfunctionSep 30, 2024Break; Break

    AS REPORTED, DURING A PROCEDURE INVOLVING ANGIOPLASTY OF THE ANTERIOR TIBIAL ARTERY, AN ADVANCE MICRO 14 ULTRA LOW-PROFILE PTA BALLOON CATHETER SEPARATED. ACCESS WAS OBTAINED IN THE PEDAL ARTERY, USING A COOK PEDAL ACCESS SET. SINGLE FEMORAL ACCESS WAS ALSO OBTAINED. THE BALLOON CATHETER TRACKED NORMALLY THROUGH THE LESION, AND THE BALLOON WAS INFLATED AND DEFLATED, COMPLETING THE PROCEDURE. WHEN THE USER REPORTEDLY ¿PINCHED¿ THE SHAFT OF THE DEVICE BETWEEN HIS FINGERS TO REMOVE IT, THE SHAFT BROKE WHERE IT WAS ¿PINCHED¿. THE USER PULLED SOME OF THE SHAFT OUT AND READJUSTED HIS FINGERS; HOWEVER, THE SHAFT BROKE AGAIN. REPORTEDLY, THIS HAPPENED SEVERAL TIMES BEFORE THE TIP SEPARATED IN THE PATIENT. PER THE REPORTER, RESISTANCE WAS NOT ENCOUNTERED UPON REMOVAL OF THE CATHETER UNTIL THE SHAFT SEPARATED AND CONTINUED TO BREAK. THE SEPARATED PORTION OF THE CATHETER REMAINED ON THE WIRE GUIDE; THEREFORE, THE USER EXTERNALIZED THE WIRE GUIDE FROM A FEMORAL APPROACH, AND A COOK CXI CATHETER WAS ADVANCED FROM THE FEMORAL SITE AND USED TO PUSH THE TIP OF THE DEVICE INTO THE PEDAL ACCESS SHEATH. THE SEPARATED PORTION OF THE COMPLAINT DEVICE WAS SUCCESSFULLY REMOVED. THE HOSPITAL LENGTH OF STA

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20340436 ↗

Other Catheter, angioplasty, peripheral, transluminal devices

All Catheter, angioplasty, peripheral, transluminal devices →

Sources and updates