Skip to content

Advance Micro G26641Recalls & reported problems

Cook Incorporated · Advance Micro 14 Ultra Low-Profile PTA Balloon Catheter · Class 2

All models of Advance Micro (25)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
5
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 5 adverse event reports that list this exact device (DI 10827002266411), received between Jun 12, 2018 and Dec 9, 2024 — classified as 5 malfunctions.

The most frequently reported specific device problem is “Material Separation” (2 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 1 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Material Separation2
  • Device Operates Differently Than Expected1
  • Difficult to Remove1
  • Fluid/Blood Leak1
  • Material Rupture1
  • Stretched1

Patient problems

  • No Consequences Or Impact To Patient2
  • No Known Impact Or Consequence To Patient2
  • No Clinical Signs, Symptoms or Conditions1

Recent adverse event reports

The 5 most recent of 5 reports, in the reporter's own words as published by the FDA.

  • MalfunctionDec 9, 2024Fluid/Blood Leak; Fluid/Blood Leak

    NO ADDITIONAL INFORMATION REGARDING THE PATIENT AND/OR EVENT HAS BEEN RECEIVED SINCE THE PREVIOUS MEDWATCH REPORT WAS SENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20884561 ↗
  • MalfunctionApr 14, 2020Material Separation; Material Separation

    NO ADDITIONAL INFORMATION REGARDING THE PATIENT AND/OR EVENT HAS BEEN RECEIVED SINCE THE PREVIOUS MEDWATCH REPORT WAS SENT.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 9957087 ↗
  • MalfunctionMar 1, 2019Material Separation; Material Separation

    NO ADDITIONAL INFORMATION REGARDING THE PATIENT AND/OR EVENT HAS BEEN RECEIVED SINCE THE PREVIOUS MEDWATCH REPORT WAS SENT.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 8382811 ↗
  • MalfunctionJul 13, 2018Difficult to Remove; Material Rupture; Difficult to Remove; Material Rupture

    DURING AN ANGIOPLASTY PROCEDURE TREATING A LESION IN THE DISTAL ANTERIOR TIBIAL ARTERY, THE BALLOON PORTION OF THE ADVANCE MICRO ¿ 14 ULTRA LOW-PROFILE PTA BALLOON CATHETER RUPTURED LONGITUDINALLY AFTER INFLATION, AT "NO MORE THAN" 16 ATMOSPHERES (ATM). THE COMPLAINT DEVICE WAS REMOVED WITH SOME DIFFICULTY, REPORTEDLY CATCHING ON CALCIFIED PORTION OF THE PATIENT'S VASCULATURE. ANOTHER BALLOON CATHETER WAS INSERTED TO AID WITH REMOVAL. THE COMPLAINT DEVICE WAS COMPLETELY REMOVED, AND THE PROCEDURE WAS COMPLETED SUCCESSFULLY. A SECTION OF THE DEVICE DID NOT REMAIN INSIDE THE PATIENT¿S BODY. THE PATIENT DID NOT REQUIRE ANY ADDITIONAL PROCEDURES DUE TO THIS OCCURRENCE. ACCORDING TO THE INITIAL REPORTER, THE PATIENT DID NOT EXPERIENCE ANY ADVERSE EFFECTS DUE TO THIS OCCURRENCE.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 7686886 ↗
  • MalfunctionJun 12, 2018Stretched; Device Operates Differently Than Expected

    THE COMPLAINANT REPORTED A CROSS-OVER PTA (PERCUTANEOUS TRANSLUMINAL ANGIOPLASTY) IN THE CRURAL VESSELS. ANOTHER MANUFACTURER'S WIRE GUIDE WAS STUCK INSIDE THE ADVANCE MICRO 14 ULTRA LOW-PROFILE PTA BALLOON CATHETER DURING RECANALIZATION OF THE CRURAL VESSELS. IT WAS FURTHER NOTED THAT THE WIRE GUIDE REMAINED INSIDE THE BALLOON CATHETER, AND A PORTION OF THE CATHETER FELT STRETCHED. THERE WERE NO REPORTED SERIOUS INJURIES TO THE PATIENT AS A RESULT OF THE PRODUCT PROBLEM.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 7591460 ↗

Other Catheter, angioplasty, peripheral, transluminal devices

All Catheter, angioplasty, peripheral, transluminal devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page