Skip to content

Advance G30968Recalls & reported problems

Cook Incorporated · Advance, 18LP Low Profile PTA Balloon Dilatation Catheter · Class 2

Adverse event reports
1
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 10827002309682), received between Jan 31, 2024 and Jan 31, 2024 — classified as 1 malfunction.

The most frequently reported specific device problem is “Material Separation” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 1 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Material Separation1

Patient problems

  • No Clinical Signs, Symptoms or Conditions1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJan 31, 2024Material Separation; Material Separation

    AS REPORTED, DURING AN ANGIOGRAM WITH DILATION, AN ADVANCE 18 LP LOW PROFILE BALLOON CATHETER¿S BALLOON SEPARATED IN THE PATIENT UPON REMOVAL OF THE DEVICE. A COOK 6-FRENCH SHEATH WAS USED DURING THE PROCEDURE. AN UNSPECIFIED INFLATION DEVICE WAS USED TO INFLATE THE BALLOON ONE TIME TO EIGHT ATMOSPHERES, WITHIN A LESION IN THE CALCIFIED RIGHT SUPERFICIAL FEMORAL ARTERY. THE LESION WAS NOT TOTALLY OCCLUDED. AFTER INFLATION, THE DEVICE WAS DEFLATED AND ALLOWED TO RETURN TO AMBIENT PRESSURE, WITH NEGATIVE PRESSURE MAINTAINED. THE BALLOON CATHETER WAS PULLED BACK OVER A WIRE GUIDE; HOWEVER, UPON REMOVAL OF THE DEVICE, THE BALLOON RIPPED IN HALF AND DETACHED INSIDE THE PATIENT. IT IS UNKNOWN IF A COUNTERCLOCKWISE ROTATION OF THE BALLOON CATHETER WAS CONDUCTED UPON REMOVAL. THE SEPARATED BALLOON MATERIAL WAS REMOVED WITH A SNARE, WITH NO ADDITIONAL SURGICAL INTERVENTION REQUIRED. THE PATIENT¿S OUTCOME WAS GOOD. A SECTION OF THE DEVICE DID NOT REMAIN INSIDE THE PATIENT¿S BODY. THE PATIENT DID NOT REQUIRE ANY ADDITIONAL PROCEDURES OR EXPERIENCE ANY ADVERSE EFFECTS DUE TO THIS OCCURRENCE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18617964 ↗

Other Catheter, angioplasty, peripheral, transluminal devices

All Catheter, angioplasty, peripheral, transluminal devices →

Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.