Zilver 518 G31352Recalls & reported problems
Cook Ireland Ltd · Zilver 518 Biliary Stent · Class 2 · Implant
All models of Zilver 518 (20)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 3
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 10827002313528), received between Feb 14, 2017 and Nov 25, 2020 — classified as 3 malfunctions.
The most frequently reported specific device problem is “Detachment of Device or Device Component” (1 report).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Most reported problems
Device problems
- Detachment of Device or Device Component1
- Device Damaged Prior to Use1
- Fracture1
- Improper or Incorrect Procedure or Method1
- Material Deformation1
- Material Separation1
- Premature Activation1
Patient problems
- No Clinical Signs, Symptoms or Conditions2
- No Known Impact Or Consequence To Patient2
- No Consequences Or Impact To Patient1
- No Patient Involvement1
Recent adverse event reports
The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.
MalfunctionNov 25, 2020Fracture; Device Damaged Prior to Use; Detachment of Device or Device Component; Fracture; Device Damaged Prior to Use; Detachment of Device or Device Component
SUPPLEMENTAL REPORT IS BEING SUBMITTED DUE TO A CORRECTIONS TO THE COMPLETED INVESTIGATION AND ANNEX CODES.
Patient problems: No Clinical Signs, Symptoms or Conditions; No Known Impact Or Consequence To Patient; No Patient Involvement
MAUDE report 10904657 ↗MalfunctionNov 5, 2020Material Separation; Improper or Incorrect Procedure or Method; Material Deformation; Material Separation; Improper or Incorrect Procedure or Method; Material Deformation
STENT WOULD NOT UNSHEATH AND DEPLOY. DEVICE WAS ABLE TO BE REMOVED WITH NO PROBLEMS AND HAS BEEN BAGGED IN ORDER TO BE SENT BACK. DEVICE EVALUATED 09-OCT-2020: "OUTER SHEATH AT WHITE CAP ON DISTAL END OF HANDLE - SEPARATED. INNER CATHETER AT APPROX. 2.5CM FROM WHITE CAP - KINKED. RED SAFETY LOCK - NOT RETURNED." DID ANY UNINTENDED SECTION OF THE DEVICE REMAIN INSIDE THE PATIENT¿S BODY? NO IF YES, PLEASE DESCRIBE. WAS THE PATIENT HOSPITALIZED OR WAS THERE PROLONGED HOSPITALIZATION? NO. DID THE PATIENT REQUIRE ANY ADDITIONAL PROCEDURES DUE TO THIS OCCURRENCE? NO. IF YES, PLEASE DESCRIBE. DID THE PRODUCT CAUSE OR CONTRIBUTE TO THE NEED FOR ADDITIONAL PROCEDURES? NO. IF YES, PLEASE SPECIFY ADDITIONAL PROCEDURES AND PROVIDE DETAILS. HAS THE COMPLAINANT REPORTED ANY ADVERSE EFFECTS ON THE PATIENT DUE TO THIS OCCURRENCE? NO. HAS THE COMPLAINANT REPORTED THAT THE PRODUCT CAUSED OR CONTRIBUTED TO THE ADVERSE EFFECTS? NO. PLEASE SPECIFY ADVERSE EFFECTS AND PROVIDE DETAILS. WAS THE DEVICE USED PERCUTANEOUSLY? YES. WHERE ON THE PATIENT WAS THE PERCUTANEOUS ACCESS SITE? CONTRALATERAL GROIN. DETAILS OF ACCESS SHEATH USED (NAME, FR SIZE, LENGTH)? COOK ANSEL 6FRX55CM. WHAT WAS THE TARGET LOCATION
Patient problems: No Clinical Signs, Symptoms or Conditions; No Known Impact Or Consequence To Patient
MAUDE report 10792603 ↗MalfunctionFeb 14, 2017Premature Activation; Premature Activation
IT WAS REPORTED WHILST ATTEMPTING TO PUT CATHETER OVER THE WIRE TIP OF CATHETER, STENT CAME OFF EXPOSING THE STENT. DID NOT ATTEMPT TO USE AFTER.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 6328718 ↗
Other CATHETER, BILIARY, DIAGNOSTIC devices
- Hurricane™ RX M00545900Boston Scientific Corporation · CATHETER, BILIARY, DIAGNOSTIC2 recalls1 death705 reports
- MUSTANG™ H74939171060470Boston Scientific Corporation · CATHETER, BILIARY, DIAGNOSTIC384 reports
- CRE™ PRO M00558710Boston Scientific Corporation · CATHETER, BILIARY, DIAGNOSTIC1 recall284 reports
- Hurricane™ RX M00545920Boston Scientific Corporation · CATHETER, BILIARY, DIAGNOSTIC2 recalls211 reports
- MUSTANG™ H74939171070470Boston Scientific Corporation · CATHETER, BILIARY, DIAGNOSTIC265 reports
- MUSTANG™ H74939171050470Boston Scientific Corporation · CATHETER, BILIARY, DIAGNOSTIC212 reports
- Hurricane™ RX M00545940Boston Scientific Corporation · CATHETER, BILIARY, DIAGNOSTIC2 recalls109 reports
- MUSTANG™ H74939171100470Boston Scientific Corporation · CATHETER, BILIARY, DIAGNOSTIC203 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.