Advance G34330Recalls & reported problems
Cook Incorporated · Advance, 18LP Low Profile PTA Balloon Dilatation Catheter · Class 2
All models of Advance (206)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 2
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 10827002343303), received between Jun 6, 2019 and Jul 31, 2020 — classified as 1 injury, 1 malfunction.
The most frequently reported specific device problem is “Material Rupture” (2 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Most reported problems
Device problems
- Material Rupture2
- Material Separation1
Patient problems
- No Code Available1
- No Consequences Or Impact To Patient1
Recent adverse event reports
The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.
InjuryJul 31, 2020Material Rupture; Material Separation; Material Rupture; Material Separation
AS REPORTED, DURING AN INTERVENTIONAL PROCEDURE OF A 100% OCCLUDED LESION IN THE DISTAL SUPERFICIAL FEMORAL ARTERY, AN ADVANCE 18 LP LOW PROFILE BALLOON CATHETER RUPTURED CIRCUMFERENTIALLY AND SEPARATED. ACCESS WAS OBTAINED IN THE LEFT COMMON FEMORAL ARTERY AND AN ANTEGRADE APPROACH WAS USED. A COOK SHEATH AND OTHER MANUFACTURER'S WIRE GUIDE WERE ALSO USED DURING THE PROCEDURE .THE VESSEL WAS MODERATELY CALCIFIED; HOWEVER, ANGULATION AND TORTUOSITY WERE NOT REPORTED. THE BALLOON WAS INFLATED ONE TIME, USING AN UNKNOWN INFLATION DEVICE, WITH A 70/30 RATIO OF VISIPAQUE CONTRAST TO SALINE. THE INFLATION PRESSURE AT THE TIME OF RUPTURE IS UNKNOWN. BLOOD WAS NOTED IN THE INFLATION DEVICE AFTER THE DEVICE RUPTURED. THE BALLOON WAS NOT INFLATED WITHIN A STENT. THE SEPARATED PORTION OF THE BALLOON WAS REMOVED WITH A SNARE IN APPROXIMATELY TEN MINUTES. ANOTHER BALLOON WAS THEN USED TO COMPLETE THE PROCEDURE. A SECTION OF THE DEVICE DID NOT REMAIN INSIDE THE PATIENT¿S BODY. THE PATIENT DID NOT REQUIRE ANY ADDITIONAL PROCEDURES DUE TO THIS OCCURRENCE. ACCORDING TO THE INITIAL REPORTER, THE PATIENT DID NOT EXPERIENCE ANY ADVERSE EFFECTS DUE TO THIS OCCURRENCE.
Patient problems: No Code Available
MAUDE report 10351291 ↗MalfunctionJun 6, 2019Material Rupture; Material Rupture
ADDITIONAL INFORMATION WAS RECEIVED ON 31MAY2019 NOTING: AN UNKNOWN 6 FRENCH SHEATH WAS UTILIZED DURING THE PROCEDURE. THE COMPLAINT DEVICE WAS INFLATED TWO TIMES TO 8 ATMOSPHERES FOR 60 SECONDS EACH TIME IN THE RIGHT SUPERFICIAL FEMORAL ARTERY, WITH A 50%/50% RATIO OF VISIPAQUE CONTRAST TO SALINE. THE PATIENT¿S ANATOMY WAS NOT REPORTED TO BE TORTUOUS OR ANGULATED; HOWEVER SEVERE CALCIFICATION WAS REPORTED. THE BALLOON WAS NOT INFLATED INSIDE OF A STENT AND WAS NOT REMOVED BETWEEN INFLATIONS. IT IS UNKNOWN IF THE BALLOON RUPTURED CIRCUMFERENTIALLY OR LONGITUDINALLY. BLOOD WAS NOTED IN THE INFLATION DEVICE UPON RUPTURE.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 8675978 ↗
Other Catheter, angioplasty, peripheral, transluminal devices
- Coyote™ ES H74939135152010Boston Scientific Corporation · Catheter, angioplasty, peripheral, transluminal879 reports
- Coyote™ ES H74939135204010Boston Scientific Corporation · Catheter, angioplasty, peripheral, transluminal585 reports
- Coyote™ ES H74939135202010Boston Scientific Corporation · Catheter, angioplasty, peripheral, transluminal437 reports
- Coyote™ ES H74939135304010Boston Scientific Corporation · Catheter, angioplasty, peripheral, transluminal384 reports
- Coyote™ ES H74939135404010Boston Scientific Corporation · Catheter, angioplasty, peripheral, transluminal293 reports
- Coyote™ ES H74939134152010Boston Scientific Corporation · Catheter, angioplasty, peripheral, transluminal279 reports
- Coyote™ H74939185202210Boston Scientific Corporation · Catheter, angioplasty, peripheral, transluminal200 reports
- Coyote™ ES H74939135254010Boston Scientific Corporation · Catheter, angioplasty, peripheral, transluminal127 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page