Skip to content

Advance G34336Recalls & reported problems

Cook Incorporated · Advance, 18LP Low Profile PTA Balloon Dilatation Catheter · Class 2

All models of Advance (206)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
2
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 10827002343365), received between May 31, 2017 and May 13, 2019 — classified as 1 injury, 1 malfunction.

The most frequently reported specific device problem is “Material Separation” (2 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Most reported problems

Device problems

  • Material Separation2
  • Crack1
  • Material Rupture1

Patient problems

  • Foreign Body In Patient1
  • No Code Available1
  • No Known Impact Or Consequence To Patient1

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • InjuryMay 13, 2019Material Rupture; Material Separation; Material Rupture; Material Separation

    IT WAS REPORTED, A FEMALE PATIENT AGED (B)(6) WAS UNDERGOING A VENOGRAM OF THE RIGHT SUBCLAVIAN VEIN. ACCESS WAS GAINED VIA THE LEFT INTERNAL JUGULAR VEIN, LEFT FEMORAL VEIN, AND RIGHT BRACHIAL VEIN IN ORDER TO TREAT A FULLY OCCLUDED LESION IN THE RIGHT SUBCLAVIAN VEIN. THE PHYSICIAN WAS ABLE TO PASS THROUGH THE OCCLUSION AND ATTEMPTED TO BALLOON WITH OTHER VENDORS' BALLOONS, SEVERAL OF WHICH BURST. THE FIRST OF TWO ADVANCE 18 LP LOW PROFILE BALLOON CATHETERS (PTA4-18-150-6-2) WAS SUBSEQUENTLY USED AND BURST BY PHYSICIAN PREFERENCE FOR THE PURPOSE OF ACCESS. THIS DEVICE WAS REMOVED AND A SECOND ADVANCE 18 LP LOW PROFILE BALLOON CATHETER (PTA4-18-150-6-8) WAS USED WHICH ALSO BURST, UNDESIRABLY, IN HALF. ONE HALF OF THIS BALLOON WAS LEFT IN THE PATIENT'S ANATOMY, REPORTEDLY CAUSING "PROBLEMS" WHILE TREATING, AND STENTED AGAINST THE VESSEL WALL. ACCORDING TO THE INITIAL REPORTER, THE PROCEDURE WAS ORIGINALLY PLANNED TO INCLUDE STENTING OF THE OCCLUDED VESSEL. ALTHOUGH THESE OCCURRENCES INCREASED THE PROCEDURE TIME, THE PROCEDURE WAS COMPLETED SUCCESSFULLY WITH A GOOD RESULT. THE PATIENT DID NOT EXPERIENCE ANY ADVERSE EFFECTS. ADDITIONAL INFORMATION PROVIDED BY THE COOK REPRESENTATIVE

    Patient problems: Foreign Body In Patient; No Code Available

    MAUDE report 8605088 ↗
  • MalfunctionMay 31, 2017Crack; Material Separation; Crack; Material Separation

    IT WAS REPORTED THAT DURING A LOWER LEG ANGIOGRAM PROCEDURE USING AN ADVANCE 18 LOW PROFILE BALLOON CATHETER, THE HUB OF THE CATHETER SEPARATED. NO UNINTENDED SECTION OF THE DEVICE REMAINED INSIDE THE PATIENT'S BODY. ANOTHER DEVICE WAS USED TO COMPLETE THE PROCEDURE. THERE WERE NO ADVERSE EFFECTS ON THE PATIENT REPORTED DUE TO THIS OCCURRENCE.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 6602576 ↗

Other Catheter, angioplasty, peripheral, transluminal devices

All Catheter, angioplasty, peripheral, transluminal devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page