Skip to content

Advance G34339Recalls & reported problems

Cook Incorporated · Advance, 18LP Low Profile PTA Balloon Dilatation Catheter · Class 2

All models of Advance (206)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
2
Deaths reported
0
Injuries reported
2
Recalls naming it
0

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 10827002343396), received between Mar 16, 2020 and Oct 3, 2023 — classified as 2 injuries.

The most frequently reported specific device problem is “Material Rupture” (2 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Material Rupture2
  • Difficult to Remove1
  • Material Separation1

Patient problems

  • Foreign Body In Patient1
  • No Clinical Signs, Symptoms or Conditions1
  • No Code Available1
  • Thrombus1

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • InjuryOct 3, 2023Difficult to Remove; Material Rupture; Difficult to Remove; Material Rupture

    AS REPORTED, DURING AN ANGIOGRAM, AN ADVANCE 18 LP LOW PROFILE BALLOON CATHETER'S BALLOON RUPTURED. THE ANATOMY WAS TORTUOUS AND HIGHLY CALCIFIED. THE BALLOON CATHETER WAS ADVANCED THROUGH AN UNSPECIFIED COOK 6-FRENCH, 45-CENTIMETER ANSEL SHEATH. ANOTHER MANUFACTURER'S INFLATION DEVICE WAS USED TO INFLATE THE BALLOON ONE TIME TO FOURTEEN ATMOSPHERES, WITHIN A 70% OCCLUDED LESION IN THE LEFT COMMON FEMORAL ARTERY, AT WHICH POINT THE BALLOON RUPTURED CIRCUMFERENTIALLY. BLOOD WAS NOTED WITHIN THE INFLATION DEVICE AT THE TIME OF RUPTURE. THE BALLOON WAS NOT INFLATED WITHIN A STENT. UPON ATTEMPTED REMOVAL OF THE BALLOON, IT WRAPPED AROUND THE SHEATH AND THE DEVICES BECAME STUCK IN THE PATIENT'S GROIN. THE PATIENT WAS TAKEN TO SURGERY, AND AN OPEN SURGICAL PROCEDURE WAS PERFORMED TO REMOVE THE BALLOON CATHETER AND SHEATH. THE PATIENT IS REPORTEDLY DOING WELL. THERE HAS BEEN NO ALLEGED MALFUNCTION OF THE COOK SHEATH. ADDITIONAL INFORMATION HAS BEEN REQUESTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17865105 ↗
  • InjuryMar 16, 2020Material Rupture; Material Separation; Material Rupture; Material Separation

    ADDITIONAL INFORMATION WAS RECEIVED. THE PHYSICIAN ATTEMPTED REMOVAL OF THE SEPARATED PORTION OF THE BALLOON AT THE TIME OF THE EVENT. THE REPORTER CONFIRMED THAT THE PATIENT HAD A BLOOD CLOT. THE PATIENT IS REPORTEDLY DOING WELL.

    Patient problems: Foreign Body In Patient; No Code Available; Thrombus

    MAUDE report 9834960 ↗

Other Catheter, angioplasty, peripheral, transluminal devices

All Catheter, angioplasty, peripheral, transluminal devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page