Zilver 635 G36239Recalls & reported problems
Cook Ireland Ltd · Zilver 635 Biliary Stent · Class 2 · Implant
All models of Zilver 635 (55)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 2
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 10827002362397), received between Apr 11, 2017 and Oct 14, 2024 — classified as 2 malfunctions.
The most frequently reported specific device problem is “Activation, Positioning or Separation Problem” (1 report).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Reports received per month
Most reported problems
Device problems
- Activation, Positioning or Separation Problem1
- Break1
- Off-Label Use1
Patient problems
- No Clinical Signs, Symptoms or Conditions1
- No Consequences Or Impact To Patient1
Recent adverse event reports
The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.
MalfunctionOct 14, 2024Off-Label Use; Off-Label Use
DEVICE TRACKING FORM RECEIVED FROM CINC REGULATORY INDICATING PATIENT IS DECEASED. PROCEDURE INDICATES TO HAVE DM CONTACT PHYSICIAN TO CONFIRM IF THE DEVICE CAUSED/CONTRIBUTED TO THE DEATH. THE PHYSICIAN INDICATED: THE PATIENT HAD A TYPE A DISSECTION COMPLICATED BY MESENTERIC, RENAL AND PERIPHERAL ARTERIAL MALPERFUSION I DID A TEVAR, DISSECTION STENT AND LEFT RENAL ARTERY STENT CT SURGERY DELAYED FIXING THE TYPE DISSECTION BECAUSE THE PATIENT WAS STILL ON ANTICOAGULANTS WE FEEL SHE RETROGRADE, DISSECTED, EVEN MORE RUPTURED THEY HAD NOTHING TO DO WITH THE STENT PER SE. PATIENT OUTCOME: PATIENT EXPIRED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20438354 ↗MalfunctionApr 11, 2017Break; Activation, Positioning or Separation Problem; Break; Activation, Positioning or Separation Problem
IT WAS REPORTED THAT THE USER ADVANCED THE DELIVERY SHEATH INTO BODY WITH STENT AND IT WAS A CALCIFIED BIFURCATION. NO TROUBLE ADVANCING SYSTEM INTO THE BODY. THEY PULLED THE RED SAFETY AND BEGAN TO PIN AND PULL THE STENT DEPLOYMENT SYSTEM. THERE WAS EXTREME RESISTANCE WHEN TRYING TO PULL BACK ON THE SYSTEM. THEY PULLED HARD ENOUGH AND THE HUB SNAPPED BACK AND THE STENT DID NOT DEPLOY. THEY TOOK THE DEPLOYMENT SYSTEM OUT OF THE BODY. THEY BELIEVED THE STENT EITHER GOT STUCK ON SHEATH OR WHEN TAKING OFF THE RED SAFETY, THE SAFETY MAY NOT HAVE UNLOCKED. THE PHYSICIAN CONSIDERED THE CASE DONE AND THE PATIENT DID NOT RECEIVE THE STENT. THE DEVICE RELATED TO THIS COMPLAINT WAS EVALUATED IN THE LAB. THE ENGINEERS DETERMINED THAT THE FAILURE MODE WAS: "HANDLE/FLEXOR SEPARATION". EVENT IS FDA MDR REPORTABLE BASED ON THE DEVICE MALFUNCTION REPORTING PRECEDENCE FOR THIS DEVICE FAMILY FOR THE ISSUE OF 'FLEXOR BREAKAGE DURING DEPLOYMENT (INCL. FLEXOR/HANDLE SEPARATION)'. NO ADVERSE EFFECTS TO THE PATIENT HAVE BEEN REPORTED AS OCCURRING.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 6481038 ↗
Other CATHETER, BILIARY, DIAGNOSTIC devices
- Hurricane™ RX M00545900Boston Scientific Corporation · CATHETER, BILIARY, DIAGNOSTIC2 recalls1 death705 reports
- MUSTANG™ H74939171060470Boston Scientific Corporation · CATHETER, BILIARY, DIAGNOSTIC384 reports
- CRE™ PRO M00558710Boston Scientific Corporation · CATHETER, BILIARY, DIAGNOSTIC1 recall284 reports
- Hurricane™ RX M00545920Boston Scientific Corporation · CATHETER, BILIARY, DIAGNOSTIC2 recalls211 reports
- MUSTANG™ H74939171070470Boston Scientific Corporation · CATHETER, BILIARY, DIAGNOSTIC265 reports
- MUSTANG™ H74939171050470Boston Scientific Corporation · CATHETER, BILIARY, DIAGNOSTIC212 reports
- Hurricane™ RX M00545940Boston Scientific Corporation · CATHETER, BILIARY, DIAGNOSTIC2 recalls109 reports
- MUSTANG™ H74939171100470Boston Scientific Corporation · CATHETER, BILIARY, DIAGNOSTIC203 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page