Zilver Flex 35 G50170Recalls & reported problems
Cook Incorporated · Zilver Flex 35 Biliary Self-Expanding Stent · Class 2 · Implant
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 3
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 10827002501703), received between Jul 25, 2017 and May 20, 2022 — classified as 3 malfunctions.
The most frequently reported specific device problem is “Off-Label Use” (2 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Most reported problems
Device problems
- Off-Label Use2
- Difficult to Advance1
- Premature Activation1
- Sticking1
Patient problems
- No Clinical Signs, Symptoms or Conditions2
- No Consequences Or Impact To Patient1
- No Patient Involvement1
Recent adverse event reports
The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.
MalfunctionMay 20, 2022Premature Activation; Off-Label Use; Premature Activation; Off-Label Use
AS REPORTED, DURING A PROCEDURE INVOLVING PLACEMENT OF A STENT WITHIN A LESION IN THE SUPERFICIAL FEMORAL ARTERY, A ZILVER FLEX 35 BILIARY SELF-EXPANDING STENT PARTIALLY DEPLOYED. THE HEAVILY CALCIFIED LESION WAS CROSSED AND PRE-DILATED WITH AN UNKNOWN BALLOON. THE COMPLAINT DEVICE WAS THEN INTRODUCED INTO THE PATIENT OVER AN UNKNOWN 0.035-INCH WIRE AND THROUGH AN UNSPECIFIED 6 FRENCH, 45-CENTIMETER SHEATH; HOWEVER, THE USER REPORTED THAT THE DELIVERY DEVICE SEEMED TO BE CAUGHT UP IN THE SUB-INTIMA. THE DEVICE WAS REMOVED FROM THE PATIENT, AT WHICH POINT THE USER NOTICED THAT THE STENT WAS PARTIALLY DEPLOYED. AT NO TIME WAS ANY PART OF THE COMPLAINT DEVICE/STENT LEFT IN THE PATIENT. ANOTHER STENT OF THE SAME SIZE WAS THEN DEPLOYED AND THE PROCEDURE WAS COMPLETED SUCCESSFULLY. A SECTION OF THE DEVICE DID NOT REMAIN INSIDE THE PATIENT¿S BODY. THE PATIENT DID NOT REQUIRE ANY ADDITIONAL PROCEDURES DUE TO THIS OCCURRENCE. ACCORDING TO THE INITIAL REPORTER, THE PATIENT DID NOT EXPERIENCE ANY ADVERSE EFFECTS DUE TO THIS OCCURRENCE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 14455488 ↗MalfunctionJan 11, 2021Off-Label Use; Difficult to Advance; Off-Label Use; Difficult to Advance
NO NEW PATIENT OR EVENT INFORMATION TO REPORT.
Patient problems: No Clinical Signs, Symptoms or Conditions; No Consequences Or Impact To Patient
MAUDE report 11147863 ↗MalfunctionJul 25, 2017Sticking; Sticking
IT WAS REPORTED THAT THE ZILVER FLEX 35 BILIARY SELF-EXPANDING STENT GOT STUCK AND PULLED OUT ON THE WIRE. THE CUSTOMER REPORTED THAT IT LOOKED FINE OUT OF THE PACKAGING. THE DEVICE DID NOT MAKE PATIENT CONTACT. THE PROCEDURE WAS COMPLETED SUCCESSFULLY WITH ANOTHER STENT. THIS PROBLEM DID NOT CREATE THE NEED FOR ANY INTERVENTION OR ADDITIONAL PROCEDURES. THERE WERE NO ADVERSE EFFECTS TO THE PATIENT.
Patient problems: No Patient Involvement
MAUDE report 6740944 ↗
Other CATHETER, BILIARY, DIAGNOSTIC devices
- Hurricane™ RX M00545900Boston Scientific Corporation · CATHETER, BILIARY, DIAGNOSTIC2 recalls1 death705 reports
- MUSTANG™ H74939171060470Boston Scientific Corporation · CATHETER, BILIARY, DIAGNOSTIC384 reports
- CRE™ PRO M00558710Boston Scientific Corporation · CATHETER, BILIARY, DIAGNOSTIC1 recall284 reports
- Hurricane™ RX M00545920Boston Scientific Corporation · CATHETER, BILIARY, DIAGNOSTIC2 recalls211 reports
- MUSTANG™ H74939171070470Boston Scientific Corporation · CATHETER, BILIARY, DIAGNOSTIC265 reports
- MUSTANG™ H74939171050470Boston Scientific Corporation · CATHETER, BILIARY, DIAGNOSTIC212 reports
- Hurricane™ RX M00545940Boston Scientific Corporation · CATHETER, BILIARY, DIAGNOSTIC2 recalls109 reports
- MUSTANG™ H74939171100470Boston Scientific Corporation · CATHETER, BILIARY, DIAGNOSTIC203 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page