Zilver 635 G50616Recalls & reported problems

Cook Ireland Ltd · Zilver 635 Biliary Stent, Expandable Metal Biliary Stent System · Class 2 · Implant

All models of Zilver 635 (55)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
1
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 10827002506166), received between Aug 28, 2018 and Aug 28, 2018 — classified as 1 malfunction.

The most frequently reported specific device problem is “Activation, Positioning or Separation Problem” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Most reported problems

Device problems

  • Activation, Positioning or Separation Problem1

Patient problems

  • No Consequences Or Impact To Patient1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • MalfunctionAug 28, 2018Activation, Positioning or Separation Problem; Activation, Positioning or Separation Problem

    REPORT IS BEING SUBMITTED UNDER THE DEVICE MALFUNCTION REPORTING PRECEDENCE FOR THIS DEVICE FAMILY FOR THE ISSUE OF ¿PREMATURE STENT DEPLOYMENT WITH SAFETY LOCK IN PLACE¿. ERCP WAS COMPLETED. UPON PLACING TWO SIDE BY SIDE STENTS, THE SECOND STENT (EVEN AFTER LUBRICATION) THERE WAS RESISTANCE AND IT WAS NOTICED THAT THE STENT HAD PARTIALLY ADVANCED OUT OF THE DELIVERY SYSTEM. THE DELIVERY SYSTEM WAS REMOVED AND THE STENT CAME OUT WITH THE DELIVERY SYSTEM FULLY "INTACT." ANOTHER DEVICE WAS PLACED TO COMPLETE THE ORIGINALLY INTENDED PROCEDURE. NO PATIENT HARM.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 7821932 ↗