Zilver 635 G50618Recalls & reported problems
Cook Ireland Ltd · Zilver 635 Biliary Stent, Expandable Metal Biliary Stent System · Class 2 · Implant
All models of Zilver 635 (55)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 5
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 5 adverse event reports that list this exact device (DI 10827002506180), received between Jan 19, 2017 and Jan 19, 2023 — classified as 5 malfunctions.
The most frequently reported specific device problem is “Premature Activation” (3 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Most reported problems
Device problems
- Premature Activation3
- Improper or Incorrect Procedure or Method2
- Break1
- Fracture1
- Material Deformation1
Patient problems
- No Clinical Signs, Symptoms or Conditions2
- No Consequences Or Impact To Patient2
- No Known Impact Or Consequence To Patient2
- Unintended Radiation Exposure1
Recent adverse event reports
The 5 most recent of 5 reports, in the reporter's own words as published by the FDA.
MalfunctionJan 19, 2023Break; Improper or Incorrect Procedure or Method; Break; Improper or Incorrect Procedure or Method
SUPPLEMENTAL FOLLOW-UP MDR REPORT IS BEING SUBMITTED DUE TO UPDATES TO THE DESCRIPTION OF EVENTS AND LAB EVALUATION NOTES ON 19-APR-2023THE STENT DELIVERY WAS ADVANCED INTO THE BILIARY TRACT VIA A DUODENAL PAPILLA APPROACH AND WAS POSITIONED AT THE TARGET SITE.WHEN THE USER RELEASED THE STENT, THE DELIVERY SHAFT CAME OFF THE HANDLE WITHOUT ANY RESISTANCE .THE STENT WAS NOT DEPLOYED AT ALL AND WAS RETRIEVED INTACT.THE USER WANTED TO USE ZILBS-635-10-4 10MM X 4CM, HOWEVER THERE WAS NO SPARE IN THE HOSPITAL INVENTORY, SO 10MM X 6CM WAS USED INSTEAD TO COMPLETE THE PROCEDURE.DESCRIPTION OF EVENT HAS BEEN REVISED BY SALE'S REP AS FOLLOWS ON 11/JAN/2023:THE STENT DELIVERY WAS ADVANCED INTO THE BILIARY TRACT VIA A DUODENAL PAPILLA APPROACH AND WAS POSITIONED AT THE TARGET SITE.WHEN THE USER TRIED TO RELEASE THE STENT, THE SHEATH WAS DISCONNECTED AT THE POINT WHERE THE COLOR CHANGED WHITE WITHOUT ANY RESISTANCE .THE STENT WAS NOT DEPLOYED AT ALL AND WAS RETRIEVED INTACT.THE USER WANTED TO USE ZILBS-635-10-4 10MM X 4CM, HOWEVER THERE WAS NO SPARE IN THE HOSPITAL INVENTORY, SO 10MM X 6CM WAS USED INSTEAD TO COMPLETE THE PROCEDURE.NOTE: USER ERROR INCORRECT SIZE 0.025IN WIREGUIDE USED BOTH FO
Patient problems: No Clinical Signs, Symptoms or Conditions; Unintended Radiation Exposure
MAUDE report 16205728 ↗MalfunctionJul 16, 2020Fracture; Improper or Incorrect Procedure or Method; Material Deformation; Fracture; Improper or Incorrect Procedure or Method; Material Deformation
SUPPLEMENTAL REPORT IS BEING SUBMITTED DUE TO THE COMPLETION OF THE INVESTIGATION.
Patient problems: No Clinical Signs, Symptoms or Conditions; No Known Impact Or Consequence To Patient
MAUDE report 10281733 ↗MalfunctionMar 23, 2017Premature Activation; Premature Activation
THE DEVICE WAS USED FOR EUS-GUIDED ANTEGRADE STENTING. WHEN ADVANCING THE DELIVERY SYSTEM, THE PHYSICIAN NOTED THAT THE STENT STARTED DEPLOYMENT BEFORE PASSING THROUGH THE STENOSED SITE ALTHOUGH THE RED SAFETY LOCK WAS STILL IN PLACE. THEN, THE DEVICE WITH THE STENT PARTIALLY DEPLOYED WAS REMOVED FROM THE PATIENT. ANOTHER MANUFACTURER'S DEVICE WAS USED INSTEAD AFTER THE EVENT.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 6428245 ↗MalfunctionMar 1, 2017Premature Activation; Premature Activation
THE PATIENT HAD PREVIOUSLY HAD RECONSTRUCTION(ROUX-EN-Y) AFTER GASTRECTOMY AND THE DOCTOR SELECTED ZILVER 635 BILIARY SELF EXPANDING METAL STENT TO TREAT HIS MALIGNANT STRICTURE OF THE INTRAHEPATIC BILE DUCT. EST WAS NOT PERFORMED. ZILBS-635-8-4 WAS INSERTED AND ADVANCED TO THE TARGET SITE USING A DOUBLE BALLOON ENDOSCOPE (FUJI FILM'S EI-530B). HOWEVER THE ANATOMY WAS SEVERE AS THE SCOPE POSITION WAS AS 'LOOPED' AND THE ADVANCING THE DELIVERY SYSTEM WAS DIFFICULT WITH STRONG RESISTANCE. THE USER SAW ONLY THE INNER CATHETER WAS ADVANCED FROM THE SHEATH, SO HE DECIDED NOT TO CONTINUE TO USE THIS DEVICE. HE REMOVED THE DELIVERY SYSTEM AND REPLACED IT WITH ZILBS-635-10-4, BUT HE EXPERIENCED THE SAME DIFFICULTY. HE DISCONTINUED THE PROCEDURE WITHOUT PLACING ANY STENTS AND PLANNED PLACING STENTS PERCUTANEOUSLY FOR THE NEXT TIME. WHEN WASHING THE ENDOSCOPE AFTER THE PROCEDURE, A STENT WAS FOUND IN THE SCOPE CHANNEL. THE USER WAS NOT SURE WHEN IT WAS DEPLOYED FROM THE DELIVERY SYSTEM OR IF IT WAS FROM 8-4 OR 10-4, BUT THEY DISCARDED BOTH DEVICES INCLUDING THIS DEPLOYED STENT. THE RED SAFETY TAB WAS NOT REMOVED. AS THERE ARE TWO SUSPECTED DEVICES INVOLVED IN THIS EVENT A SEPARATE REPORT HAS
Patient problems: No Consequences Or Impact To Patient
MAUDE report 6367567 ↗MalfunctionJan 19, 2017Premature Activation; Premature Activation
A FOLLOW-UP MDR IS BEING SUBMITTED TO INCLUDE THE DEVICE EVALUATION DETAILS. INITIAL REPORT DETAILS: METALIC STENTS HAD BEEN PLACED IN THE RIGHT AND LEFT HEPATIC DUCTS WITH SIDE-BY-SIDE TECHNIQUE BEFORE. THE ZILBS- STENT WAS TO BE ADDITIONALLY PLACED IN THE RIGHT HEPATIC DUCT AS REINTERVENTION. DURING ADVANCEMENT OF THE DELIVERY SYSTEM, THE STENT WAS PARTIALLY DEPLOYED. THEREFORE, THE DELIVERY SYSTEM WAS REMOVED FROM THE PATIENT AND THE PROCEDURE WAS FINISHED WITHOUT ANY FURTHER TREATMENT.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 6263432 ↗
Other CATHETER, BILIARY, DIAGNOSTIC devices
- Hurricane™ RX M00545900Boston Scientific Corporation · CATHETER, BILIARY, DIAGNOSTIC2 recalls1 death705 reports
- MUSTANG™ H74939171060470Boston Scientific Corporation · CATHETER, BILIARY, DIAGNOSTIC384 reports
- CRE™ PRO M00558710Boston Scientific Corporation · CATHETER, BILIARY, DIAGNOSTIC1 recall284 reports
- Hurricane™ RX M00545920Boston Scientific Corporation · CATHETER, BILIARY, DIAGNOSTIC2 recalls211 reports
- MUSTANG™ H74939171070470Boston Scientific Corporation · CATHETER, BILIARY, DIAGNOSTIC265 reports
- MUSTANG™ H74939171050470Boston Scientific Corporation · CATHETER, BILIARY, DIAGNOSTIC212 reports
- Hurricane™ RX M00545940Boston Scientific Corporation · CATHETER, BILIARY, DIAGNOSTIC2 recalls109 reports
- MUSTANG™ H74939171100470Boston Scientific Corporation · CATHETER, BILIARY, DIAGNOSTIC203 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page