Zilver 635 G50622Recalls & reported problems
Cook Ireland Ltd · Zilver 635 Biliary Stent, Expandable Metal Biliary Stent System · Class 2 · Implant
All models of Zilver 635 (55)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 14
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 14 adverse event reports that list this exact device (DI 10827002506227), received between Jun 20, 2016 and Jun 2, 2026 — classified as 1 injury, 13 malfunctions.
The most frequently reported specific device problem is “Activation, Positioning or Separation Problem” (5 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Reports received per month
Most reported problems
Device problems
- Activation, Positioning or Separation Problem5
- Break4
- Delivery System Failure1
- Dent in Material1
- Fracture1
- Improper or Incorrect Procedure or Method1
- Material Deformation1
- Material Integrity Problem1
Patient problems
- No Consequences Or Impact To Patient8
- No Clinical Signs, Symptoms or Conditions4
- No Known Impact Or Consequence To Patient3
Recent adverse event reports
The 14 most recent of 14 reports, in the reporter's own words as published by the FDA.
MalfunctionJun 2, 2026Dent in Material
THROUGH THE ENDOSCOPIC CHANNEL, THE STENT WAS GUIDED ALONG THE ACRO WIRE GUIDE TO THE COMMON BILE DUCT. AFTER A LONG EFFORT, IT WAS UNABLE TO CROSS THE STENOSIS AREA. USER RETRACTED THE DEVICE FROM PATIENT, IT WAS FOUND THAT THE TIP OF THE STENT WAS BENT AND UNUSABLE. DURING THE CLEANING PROCESS, THE STENT WAS ACCIDENTALLY RELEASED, AND THE KINK INSIDE THE STENT WERE CLEARLY VISIBLE. SUCCESSFULLY COMPLETED THE PROCEDURE BY CHANGING TO ANOTHER SAME RPN DEVICE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25358719 ↗MalfunctionJan 11, 2022Break; Material Integrity Problem; Break; Material Integrity Problem
USER PLANNED TO PLACE 2 STENTS IN PATIENT. USER PLACED 2 WIRE GUIDE RESPECTIVELY IN LEFT AND RIGHT HEPATIC DUCT, THEN OPENED THE DEVICE PACKAGE AND INJECT WATER, THEN ADVANCED THE DELIVERY SYSTEM THROUGH WIRE GUIDE IN ENDOSCOPY WHILE USER DETECTED THE WHITE TIP OF DELIVERY SYSTEM MISSING. USER THEN RETRACTED THE DELIVERY SYSTEM AND CHANGED TO ANOTHER PRODUCT FROM SAME LOT# BUT OPENED THE PACKAGE FOUND OUT THE WHITE TIP MISSING AS WELL. USER THEN CHANGED TO BOSTON SCIENTIFIC SIMILAR DEVICE TO COMPLETE THE PROCEDURE. "A SECTION OF THE DEVICE DID NOT REMAIN INSIDE THE PATIENT¿S BODY. THE PATIENT DID NOT REQUIRE ANY ADDITIONAL PROCEDURES DUE TO THIS OCCURRENCE. ACCORDING TO THE INITIAL REPORTER, THE PATIENT DID NOT EXPERIENCE ANY ADVERSE EFFECTS DUE TO THIS OCCURRENCE." ARE IMAGES OF THE DEVICE OR PROCEDURE AVAILABLE? NO. WAS BALLOON DILATION PERFORMED PRIOR TO USE OF THE STENT DEVICE? YES. WHAT WAS THE LENGTH AND DIAMETER OF THE STRICTURE? 2CM LONG, 3 DIAMETER. WAS THE PRODUCT INSPECTED FOR KINKS OR DAMAGE BEFORE USE? NO. WHAT WAS THE POSITION OF THE ELEVATOR DURING DEPLOYMENT? CLOSED. WAS THE DELIVERY SYSTEM TRACKED AROUND A TIGHT ANGLE IN THE PATIENT ANATOMY OR AROUND A TIGHT ANGLE
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 13224393 ↗MalfunctionSep 3, 2021Break; Improper or Incorrect Procedure or Method; Break; Improper or Incorrect Procedure or Method
DEVICE EVALUATED ON 14-OCT-21: "STENT STRUT PROTRUDING THROUGH OUTER SHEATH. COILED SECTION OF OUTER SHEATH SEPARATED."
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12421825 ↗MalfunctionAug 17, 2020Material Deformation; Material Deformation
SUPPLEMENTAL REPORT BEING SUBMITTED DUE TO THE INVESTIGATION BEING COMPLETED ON 16-MAR-2021. AFTER THEY PLACE WHOLE DEVICE INTO THE PLACE AND THEY TRY TO DEPLOY THE DEVICE AND THEY CANNOT DEPLOY, SO THEY CHANGE A NEW ONE.
Patient problems: No Clinical Signs, Symptoms or Conditions; No Known Impact Or Consequence To Patient
MAUDE report 10417662 ↗MalfunctionSep 11, 2019Activation, Positioning or Separation Problem; Activation, Positioning or Separation Problem
USER ADVANCED THE DEVICE THROUGH ENDOSCOPY BUT FOUND OUT THE STENT PREMATURE RELEASED 0.5CM WHILE PASSING DUODENAL PAPILLA WHICH CAUSE THE DEVICE CANNOT ENTER BILE DUCT. USER RETRACTED THE DEVICE FROM PATIENT AND CHECKED THE DEVICE FOUND OUT THE STENT CANNOT BE RETRACTED BACK INTO DELIVERY SYSTEM. USER CHANGED ANOTHER SAME DEVICE TO COMPLETE THE PROCEDURE. FDA MDR REPORTING REQUIRED: EVENT IS FDA MDR REPORTABLE BASED ON THE DEVICE MALFUNCTION REPORTING PRECEDENCE FOR THIS DEVICE FAMILY FOR THE ISSUE OF ¿PREMATURE STENT DEPLOYMENT WITH SAFETY LOCK IN PLACE¿. NO ADVERSE EFFECTS TO THE PATIENT HAVE BEEN REPORTED AS OCCURRING.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 8988825 ↗MalfunctionApr 26, 2019Fracture; Fracture
USER ADVANCED THE DEVICE THROUGH WIRE GUIDE WHILE FOUND OUT THE DELIVERY SYSTEM DETACHED FROM OUTER SHEATH. USER CHANGED ANOTHER SAME DEVICE TO COMPLETE THE PROCEDURE.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 8554231 ↗MalfunctionNov 15, 2018Break; Break
AS REPORTED TO CUSTOMER RELATIONS: "WHEN THEY WERE GOING TO DEPLOY THE STENT, THEY HAD A LOT OF RESISTANCE AND THE OUTER SHEATH BY THE HANDLE BECAME SEPARATED FROM THE HANDLE. ABOUT 2-3 CM FROM THE TIP OF THE CATHETER, THERE IS A KINK ON THE CATHETER. THE STENT IS PARTIALLY DEPLOYED (ABOUT 1 CM OF IT). DEVICE WAS REMOVED, ANOTHER DEVICE WAS USED TO COMPLETE THE PROCEDURE SUCCESSFULLY."
Patient problems: No Consequences Or Impact To Patient
MAUDE report 8073138 ↗MalfunctionMay 2, 2018Premature Activation; Premature Activation
INITIAL MDR IS BEING SUBMITTED BASED ON THE DEVICE MALFUNCTION PRECEDENCE: 'PREMATURE STENT DEPLOYMENT WITH THE SAFETY LOCK IN PLACE.' USER PLAN TO IMPLANT 2 STENT IN BOTH SIDE OF HEPATIC DUCT. ONE STENT WAS ADVANCED TO RIGHT HEPATIC DUCT SUCCESSFULLY, ANOTHER STENT WAS ADVANCED INTO ENDOSCOPY CHANNEL BUT WITH PART DEPLOYED WHILE EMERGED FROM ENDOSCOPY IN PATIENT. RED LOCK WAS IN PLACE AND SECURED AT THE MOMENT. USER DECIDED DEPLOYED THE ONE IN RIGHT HEPATIC DUCT WHICH TURNS OUT SUCCESSFULLY. AND THEN WITHDRAW THE PART DEPLOYED STENT ALONG WITH ENDOSCOPY FROM PATIENT. USER RELEASE THE STENT IN ENDOSCOPY CHANNEL AFTER WITHDRAW. USER WITHDRAW THE DEVICE FROM ENDOSCOPY AT LAST.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 7479575 ↗MalfunctionDec 18, 2017Delivery System Failure; Delivery System Failure
PER REP - "DEVICE WAS PLACED OVER GUIDE WIRE INTO LEFT INTRAHEPATIC DUCT. DEVICE WAS FLUSHED IN BOTH PORTS PRIOR TO INSERTION. UPON DEPLOYMENT, RESISTANCE WAS FELT AND THE PHYSICIAN RELAXED THE ELEVATOR ON THE SCOPE. ASSISTANT ATTEMPTED TO DEPLOY STENT WITH RESISTANCE AND THE CATHETER SEPARATED FROM THE LUER LOCK HANDLE RENDERING THE DEVICE USELESS. DEVICE HAD NOT DEPLOYED AND THE CATHETER WAS REMOVED LEAVING GUIDE WIRE IN PLACE AND A REPLACEMENT LIKE DEVICE WAS USED TO COMPLETE PROCEDURE."
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 7126845 ↗MalfunctionDec 14, 2017Activation, Positioning or Separation Problem; Activation, Positioning or Separation Problem
THIS FOLLOW UP REPORT IS BEING SUBMITTED DUE TO THE EVALUATION OF THE DEVICE INVOLVED IN THIS EVENT AND THE CONCLUSION OF THIS INVESTIGATION. INITIAL REPORT DETAILS: THE SHEATH OF THE ZILVER STENT DISCONNECTED FROM THE HANDLE. THE DEVICE WAS REMOVED AND ANOTHER DEVICE WAS USED TO COMPLETE THE PROCEDURE.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 7116434 ↗MalfunctionMar 14, 2017Break; Break
DURING THE STENT PLACEMENT PROCEDURE, THE CATHETER OF THE STENT BROKE AWAY FROM THE HANDLE AND DEPLOYMENT WAS UNABLE TO BE SUCCESSFUL. ANOTHER DEVICE WAS USED TO COMPLETE THE PROCEDURE WITH NO HARM TO THE PATIENT.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 6402749 ↗MalfunctionFeb 7, 2017Activation, Positioning or Separation Problem; Activation, Positioning or Separation Problem
DURING ERCP, THE USER ADVANCED TWO STENTS ALONG WIRE GUIDE UNDER ENDOSCOPE THROUGH LEFT AND RIGHT HEPATIC DUCT. A -REPRESENT SENT THROUGH LEFT HEPATIC DUCT. B- -REPRESENT SENT THROUGH RIGHT HEPATIC DUCT. THE USER ADVANCED B INTO DESIRED POSITION SUCCESSFULLY AND TRIED TO DEPLOY TWO STENTS AT THE SAME TIME. THEN THEY FOUND THAT THE B STENT COULD NOT DEPLOY BECAUSE THE HANDLE PORTION WAS DETACHED. WITHDRAW THE STENT IMMEDIATELY. AFTER WITHDRAWING, ABOUT 55 CM DISTAL FROM THE END OF HANDLE OF OUTER SHEATH WAS PEELING. THE PEELED PORTION SHEATH WAS OPERATED OUTSIDE, SO NO SECTION WAS LEFT INSIDE PATIENT BODY.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 6307215 ↗InjuryFeb 2, 2017Activation, Positioning or Separation Problem; Activation, Positioning or Separation Problem
THE CATHETER BROKE / SNAPPED IN HALF. THE ENTIRE CATHETER WAS PULLED OUT. A DIFFERENT STENT WAS USED AND PLACED TO COMPLETE THE PROCEDURE. ADDITIONAL INFORMATION RECEIVED ON 06-JAN-2017 AS FOLLOWS: WILL THE COMPLAINT DEVICE STILL BE RETURNED FOR INVESTIGATION? YES SENT INFORMATION. AT WHAT POINT OF THE PROCEDURE THE CATHETER BROKE? WAS IT DURING DEPLOYMENT? STENT DEPLOYMENT. HOW DID THEY PULL THE DEVICE OUT FROM THE PATIENT? DID THEY USE FORCEPS OR ANY TOOLS TO RETRIEVE THE DEVICE? DEVICE STAYED IN PATIENT. ON RETRACTION OF STENT CATHETER, STENT DEPLOYED IN CBD. WHERE WAS THE POINT OF BREAKAGE??? WAS THE CATHETER PULLED FROM THE HANDLE MAYBE? I HAVE NO CLUE WHY THE CATHETER WAS PULLED FROM WHAT HANDLE? CAN THEY CONFIRM THAT THEY REFER TO THE OUTER CATHETER NOT THE INNER CATHETER? NO THAT SHOULD BE DETERMINED BY INVESTIGATING REPORT. BASED ON THE ADDITIONAL INFORMATION THE STENT WAS DEPLOYED IN THE CBD. ADDITIONAL INFORMATION CONFIRMED THIS WAS UNINTENTIONAL DEPLOYMENT. NO ADVERSE EFFECTS TO THE PATIENT ARE REPORTED. HOWEVER EVENT MEETS FDA 'MALFUNCTION' REPORTING CRITERIA. THE RISK IS NOT REMOTE.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 6295875 ↗MalfunctionJun 20, 2016Activation, Positioning or Separation Problem; Activation, Positioning or Separation Problem
DURING THE ERCP BOTH SIDE OF HEPATIC DUCT METAL ZILBS STENT PLACEMENT, THE FIRST STENT WAS PLACED WELL AND ADVANCE THE SECOND STENT THROUGH THE ENDOSCOPE. THE CONNECTION BETWEEN THE TIP AND OUTER SHEATH WAS PARTLY SEPARATED. THE RED SAFETY LOCK HAD NOT BEEN OPENED. THE USER WITHDREW THE STENT AND REPLACED WITH A NEW DEVICE TO FINISH THE PROCEDURE SUCCESSFULLY.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 5735425 ↗
Other CATHETER, BILIARY, DIAGNOSTIC devices
- Hurricane™ RX M00545900Boston Scientific Corporation · CATHETER, BILIARY, DIAGNOSTIC2 recalls1 death705 reports
- MUSTANG™ H74939171060470Boston Scientific Corporation · CATHETER, BILIARY, DIAGNOSTIC384 reports
- CRE™ PRO M00558710Boston Scientific Corporation · CATHETER, BILIARY, DIAGNOSTIC1 recall284 reports
- Hurricane™ RX M00545920Boston Scientific Corporation · CATHETER, BILIARY, DIAGNOSTIC2 recalls211 reports
- MUSTANG™ H74939171070470Boston Scientific Corporation · CATHETER, BILIARY, DIAGNOSTIC265 reports
- MUSTANG™ H74939171050470Boston Scientific Corporation · CATHETER, BILIARY, DIAGNOSTIC212 reports
- Hurricane™ RX M00545940Boston Scientific Corporation · CATHETER, BILIARY, DIAGNOSTIC2 recalls109 reports
- MUSTANG™ H74939171100470Boston Scientific Corporation · CATHETER, BILIARY, DIAGNOSTIC203 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page