Skip to content

Hall 50 PRO7400BRecalls & reported problems

Conmed Corporation · HALL 50 RECIPROCATING SAW BATTERY HANDPIECE · Class 1

Data updated Oct 13, 2026 from official FDA records · Editorial policy

Adverse event reports
1
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 10845854040058), received between Apr 23, 2025 and Apr 23, 2025 — classified as 1 malfunction.

The most frequently reported specific device problem is “Detachment of Device or Device Component” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 1 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Detachment of Device or Device Component1

Patient problems

  • No Clinical Signs, Symptoms or Conditions1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • MalfunctionApr 23, 2025Detachment of Device or Device Component; Detachment of Device or Device Component

    A SALES REPRESENTATIVE REPORTED ON BEHALF OF A CUSTOMER THAT THE PRO7400B, HALL 50 RECIPROCATING SAW BATTERY HANDPIECE DEVICE WAS USED IN A TOTAL HIP PROCEDURE ON AN UNKNOWN DATE AND ¿THE SAW BLADE DETACHED FROM THE HANDPIECE TWO TIMES DURING ONE SURGERY. THE NURSE CAREFULLY ATTACHED THE SAW BLADE AFTER THE FIRST TIME BUT IT DETACHED AGAIN DURING ACTION DESPITE THAT. THE HANDPIECE WILL BE SENT FOR REPAIR. NO HARM CAME TO THE PATIENT AND THEY FINISHED THE SURGERY WITH ANOTHER HANDPIECE.". THE PROCEDURE WAS COMPLETED WITH ¿ANOTHER PRO7400B¿ AND A DELAY OF 5 MINUTES. FURTHER ASSESSMENT FOUND "THE BLADE CAME LOOSE DURING THE CUT OF COLLUM FEMORIS. THERE WERE NO PATIENT IMPACT OR INJURY. THERE WAS A MINIMAL DELAY IN THE SURGERY WHEN THEY OPENED A NEW HANDPIECE. THE TECHNICIANS IN THE HOSPITAL HAVE INVESTIGATED THE HANDPIECE AND CAN¿T FIND ANYTHING WRONG WITH IT. THEY HAVE DECIDED TO TRY THE SAW AGAIN DURING SURGERY AND NOT RETURNING IT TO US IF DOES NOT HAPPEN AGAIN. THE BLADE DETACHED ¿ THEY HAD TO STOP AND ATTACH THE BLADE AGAIN. THE SAME THING HAPPENED AGAIN WHEN THEY TRIED TO COMPLETE THE CUT. THEY THEN BROUGHT IN A DIFFERENT SAW AND COMPLETED THE OPERATION WITHOUT COMPLICATIONS. TH

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21889937 ↗

Other SAW, POWERED, AND ACCESSORIES devices

All SAW, POWERED, AND ACCESSORIES devices →

Sources and updates