GRAFTMAX KS-ALBRecalls & reported problems
Conmed Corporation · GraftMax Button ALB (Adjustable Loop Button) · Class 2 · Implant
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 2
- Deaths reported
- 0
- Injuries reported
- 2
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 10845854042229), received between Mar 19, 2019 and Jun 16, 2020 — classified as 2 injuries.
The most frequently reported specific device problem is “Detachment of Device or Device Component” (1 report).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Most reported problems
Device problems
- Detachment of Device or Device Component1
- Osseointegration Problem1
- Patient Device Interaction Problem1
Patient problems
- Device Embedded In Tissue or Plaque1
- Failure of Implant1
Recent adverse event reports
The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.
InjuryJun 16, 2020Detachment of Device or Device Component; Detachment of Device or Device Component
THE DISTRIBUTOR REPORTED ON BEHALF OF THEIR CUSTOMER THAT THE KS-ALB, GRAFTMAX ADJUSTABLE LOOP BUTTON, WAS USED IN AN ACL RECONSTRUCTION WITH HAMSTRING GRAFT ON (B)(6) 2020. THE PLATE (BUTTON) BECAME DISLODGED FROM THE BONE WHEN PULLING THE THREADS (SUTURES) TO RAISE THE GRAFT. THE THREADS (SUTURES) BROKE THE PLATE (BUTTON) AND THE BUTTON MIGRATED TO THE THIGH MUSCLE. THE DOCTOR ATTEMPTED TO RETRIEVE THE BUTTON BUT WAS NOT SUCCESSFUL. THE PROCEDURE WAS COMPLETED USING AN INTERFERENCE SCREW, BRAND UNKNOWN. THERE WAS A 40-MINUTE DELAY. ALTHOUGH THE PATIENT IS FINE AND THERE IS NO REPORTED INJURY OR INDICATION ANOTHER SURGERY WILL BE REQUIRED TO REMOVE THE PLATE(BUTTON) THIS IS CONSIDERED AN INJURY FOR CONMED. THIS REPORT IS BEING RAISED DUE TO PATIENT INJURY.
Patient problems: Device Embedded In Tissue or Plaque
MAUDE report 10157319 ↗InjuryMar 19, 2019Osseointegration Problem; Patient Device Interaction Problem; Osseointegration Problem; Patient Device Interaction Problem
THE CUSTOMER REPORTED THE KS-ALB, GRAFTMAX BUTTON FAILED DURING AN ACL REPAIR ON (B)(6) 2019. AFTER THE PATIENT TOOK HER FIRST STEPS, THERE WAS A "SNAP" AND SHE LOST HER BALANCE. THE PATIENT REMAINED IN THE HOSPITAL UNTIL HER REVISION ACL REPAIR ON (B)(6) 2019. IT WAS DETERMINED THAT THE LOCKING MECHANISM SLIPPED OPEN. THE REVISION SURGERY WAS COMPLETED SUCCESSFULLY USING ANOTHER OF THE SAME GRAFTMAX BUTTONS. THIS REPORT IS BEING RAISED BASED ON PATIENT INJURY AND REVISION SURGERY.
Patient problems: Failure of Implant
MAUDE report 8434500 ↗
Other FASTENER, FIXATION, NONDEGRADABLE, SOFT TISSUE devices
- TRUESPAN ORTHOCORD 228151Medos International Sàrl · FASTENER, FIXATION, NONDEGRADABLE, SOFT TISSUE1,686 reports
- TRUESPAN ORTHOCORD 228152Medos International Sàrl · FASTENER, FIXATION, NONDEGRADABLE, SOFT TISSUE1,398 reports
- OMNISPAN ORTHOCORD 228142Medos International Sàrl · FASTENER, FIXATION, NONDEGRADABLE, SOFT TISSUE932 reports
- FOOTPRINT 72202901Smith & Nephew, Inc. · FASTENER, FIXATION, NONDEGRADABLE, SOFT TISSUE721 reports
- RIGIDLOOP 232447Medos International Sàrl · FASTENER, FIXATION, NONDEGRADABLE, SOFT TISSUE633 reports
- Arthrex® AR-3636Arthrex, Inc. · Fastener, fixation, nondegradable, soft tissue624 reports
- OMNISPAN ORTHOCORD 228141Medos International Sàrl · FASTENER, FIXATION, NONDEGRADABLE, SOFT TISSUE609 reports
- TRUESPAN ORTHOCORD 228162Medos International Sàrl · FASTENER, FIXATION, NONDEGRADABLE, SOFT TISSUE460 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page