Infant Flow SiPAP 675-CFG-005Recalls & reported problems

Carefusion Corporation · ASSY MK3 IF DRVR CONFIG US OVRLY/ PLUS/ EN · Class 2

All models of Infant Flow SiPAP (7)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Quick summary

This Vyaire SiPAP ventilator, made by CareFusion, is a continuous-use facility ventilator. The 29 reports mostly describe the unit not building pressure or not delivering breaths, gas leaking out of the exhaust, internal leaks, and a damaged display. Three reports involved injuries; no deaths were reported. No recalls name this device.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
29
Deaths reported
0
Injuries reported
3
Recalls naming it
0

Summary

The FDA's MAUDE database contains 29 adverse event reports that list this exact device (DI 10846446004847), received between Jul 9, 2019 and Apr 8, 2024 — classified as 3 injuries, 26 malfunctions.

The most frequently reported specific device problem is “Mechanical Problem” (11 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 1 reportsApr 2024: 2 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520262
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Mechanical Problem11
  • Leak/Splash5
  • Infusion or Flow Problem3
  • Pressure Problem3
  • No Pressure2
  • Defective Alarm1
  • Device Alarm System1
  • Erratic or Intermittent Display1

Patient problems

  • No Clinical Signs, Symptoms or Conditions26
  • No Patient Involvement2
  • No Information1

Recent adverse event reports

The 25 most recent of 29 reports, in the reporter's own words as published by the FDA.

  • MalfunctionApr 8, 2024No Device Output; No Device Output

    IT WAS REPORTED TO VYAIRE MEDICAL THE SIPAP IS IN BIPHASIC MODE, THE BREATHS WILL NOT DELIVER NO MATTER WHAT RATE IS SET. FURTHERMORE, THERE WAS NO PATIENT ASSOCIATED WITH THE REPORTED EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19059454 ↗
  • MalfunctionApr 4, 2024No Pressure; No Pressure

    IT WAS REPORTED TO VYAIRE MEDICAL THE VENT IS NOT PRESSURIZING.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19047222 ↗
  • MalfunctionMar 4, 2024No Pressure; No Pressure

    IT WAS REPORTED TO VYAIRE MEDICAL THE SIPAP WAS NOT PRESSURIZING AND THAT ALL OF THE GAS WAS COMING FROM THE EXHAUST. FURTHERMORE, THERE WAS NO PATIENT ASSOCIATED WITH THE REPORTED EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18826493 ↗
  • MalfunctionOct 3, 2023Leak/Splash; Leak/Splash

    IT WAS REPORTED TO VYAIRE MEDICAL THAT SIPAP VENTILATOR HAD AN EXCESSIVE LEAK FROM THE EXHAUST. THERE IS NO INFORMATION ABOUT PATIENT INVOLVEMENT ASSOCIATED WITH THIS REPORTED EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17865033 ↗
  • MalfunctionJul 11, 2023Mechanical Problem; Mechanical Problem

    IT WAS REPORTED TO VYAIRE MEDICAL THE SIPAP HAVE A DAMAGED DISPLAY. FURTHERMORE, THERE WAS NO PATIENT ASSOCIATED WITH THE REPORTED EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17302825 ↗
  • InjuryApr 28, 2023Mechanical Problem; Mechanical Problem

    IT WAS REPORTED TO VYAIRE MEDICAL THAT THE SIPAP VENTILATOR SHUT DOWN ON PATIENT. IT WAS ALSO STATED THAT AN E50 AND AN E40 ALARM HAS OCCURRED ON THE UNIT. FURTHERMORE, THE PATIENT WAS TRANSFERRED TO ANOTHER DEVICE AND THERE WAS NO PATIENT HARM ASSOCIATED WITH THE EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16834971 ↗
  • InjuryMar 27, 2023Leak/Splash; Leak/Splash

    JIT WAS REPORTED TO VYAIRE MEDICAL THAT UNIT HAS AN INTERNAL LEAK DURING USE ON A PATIENT ON THE SIPAP VENTILATOR. THE CUSTOMER REPORTED THE UNIT WAS REPLACED. THE CUSTOMER CONFIRMED THERE WAS NO PATIENT HARM ASSOCIATED WITH THE REPORTED EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16624812 ↗
  • MalfunctionFeb 8, 2023Leak/Splash; Leak/Splash

    IT WAS REPORTED TO VYAIRE MEDICAL THAT DEVICE IS CONSTANTLY LEAKING FROM IPX1 PORT ON THE BACK OF THE SIPAP VENTILATOR. THE CUSTOMER REPORTED THERE WAS NO PATIENT INVOLVEMENT ASSOCIATED WITH THE REPORTED EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16327235 ↗
  • MalfunctionSep 14, 2022Leak/Splash; Leak/Splash

    IT WAS REPORTED TO VYAIRE MEDICAL THAT THERE IS A LARGE LEAK IN EXHALATION BLOCK ON THE SIPAP VENTILATOR. AT THIS TIME, THERE IS NO INFORMATION REGARDING PATIENT INVOLVEMENT ASSOCIATED WITH THE REPORTED EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15420240 ↗
  • MalfunctionJul 25, 2022Mechanical Problem; Mechanical Problem

    IT WAS REPORTED TO VYAIRE MEDICAL THAT THE SIPAP VENTILATOR TOUCH SCREEN WAS NOT WORKING. AT THIS TIME, THERE IS NO INFORMATION ABOUT PATIENT INVOLVEMENT ASSOCIATED WITH THIS REPORTED EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15097160 ↗
  • MalfunctionJun 8, 2022Pressure Problem; Pressure Problem

    THE CUSTOMER REPORTED TO VYAIRE MEDICAL THAT A SIPAP VENTILATOR WAS OVER PRESSURIZING AND THEN POPPING OFF. THERE WAS NO PATIENT INVOLVEMENT ASSOCIATED WITH THE REPORTED EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 14640167 ↗
  • MalfunctionMar 15, 2022Infusion or Flow Problem; Infusion or Flow Problem

    THE CUSTOMER REPORTED FLUCTUATING POSITIVE END EXPIRATORY PRESSURE ON THE SIPAP VENTILATOR. AT THIS TIME, THERE IS NO INFORMATION REGARDING PATIENT INVOLVEMENT ASSOCIATED WITH THE REPORTED EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 13766404 ↗
  • MalfunctionMar 15, 2022Infusion or Flow Problem; Infusion or Flow Problem

    THE CUSTOMER REPORTED FLUCTUATING POSITIVE END EXPIRATORY PRESSURE DURING USE ON A PATIENT ON THE SIPAP VENTILATOR. AT THIS TIME, THERE IS NO INFORMATION REGARDING PATIENT HARM ASSOCIATED WITH THE REPORTED EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 13766306 ↗
  • MalfunctionMar 15, 2022Infusion or Flow Problem; Infusion or Flow Problem

    THE CUSTOMER REPORTED FLUCTUATING POSITIVE END EXPIRATORY PRESSURE ON THE SIPAP VENTILATOR. AT THIS TIME, THERE IS NO INFORMATION REGARDING PATIENT INVOLVEMENT ASSOCIATED WITH THE REPORTED EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 13766190 ↗
  • MalfunctionFeb 7, 2022Leak/Splash; Leak/Splash

    THE CUSTOMER REPORTED THE SIPAP WAS SOUNDING LIKE IT HAD A LEAK. THE CUSTOMER REPORTED THERE WAS NO PATIENT INVOLVEMENT ASSOCIATED WITH THE REPORTED EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 13476750 ↗
  • MalfunctionDec 24, 2021Pressure Problem; Pressure Problem

    THE CUSTOMER REPORTED TO VYAIRE MEDICAL THAT THE SIPAP VENTILATOR WAS NOT BUILDING UP PRESSURE. AT THIS TIME, THERE IS NO INFORMATION REGARDING PATIENT INVOLVEMENT ASSOCIATED WITH THE REPORTED EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 13083273 ↗
  • MalfunctionDec 1, 2021Mechanical Problem; Mechanical Problem

    THE CUSTOMER REPORTED TO VYAIRE MEDICAL THAT THE BLOCK ABOVE THE WATER TRAP IS INTERNALLY LEAKING OF THE SIPAP VENTILATOR. FURTHERMORE, THERE WAS NO PATIENT INVOLVEMENT ASSOCIATED WITH REPORTED EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12909883 ↗
  • MalfunctionNov 16, 2021Pressure Problem; Pressure Problem

    THE CUSTOMER REPORTED TO VYAIRE MEDICAL THAT THE SIPAP VENTILATOR UNIT JUST FINISHED PM, NOW AT FLOW OF 8 GENERATING PRESSURE GREATER THAN 10 CAUSING UNIT TO DUMP. THE REPORTER CONFIRMED THAT THERE IS NO PATIENT INVOLVEMENT ASSOCIATED ON THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12823868 ↗
  • MalfunctionNov 9, 2021Mechanical Problem; Mechanical Problem

    THE CUSTOMER REPORTED CANNOT GET THE DEVICE TO PRESSURIZE ABOVE 1.7CENTIMETERS ON THE SIPAP VENTILATOR. THE CUSTOMER REPORTED THERE WAS NO PATIENT INVOLVEMENT ASSOCIATED WITH THE REPORTED EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12782848 ↗
  • MalfunctionNov 7, 2021Mechanical Problem; Mechanical Problem

    THE CUSTOMER REPORTED TO VYAIRE MEDICAL THAT THE SIPAP VENTILATOR UNIT WAS HAVING FIO2/FRACTION OF INSPIRED OXYGEN FLUCTUATE. AT THIS TIME, THERE IS NO INFORMATION REGARDING PATIENT INVOLVEMENT ASSOCIATED WITH THE REPORTED EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12756143 ↗
  • MalfunctionSep 16, 2021Mechanical Problem; Mechanical Problem

    THE CUSTOMER REPORTED TO VYAIRE MEDICAL THAT THE LCD (LIQUID CRYSTAL DISPLAY) IS DIM AND NOT RESPONDING. AT THIS TIME, THERE IS NO INFORMATION REGARDING PATIENT INVOLVEMENT ASSOCIATED WITH THE REPORTED EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12483294 ↗
  • MalfunctionJul 20, 2021Mechanical Problem; Mechanical Problem

    THE CUSTOMER REPORTED HE WAS HAVING ISSUE WITH THE OXYGEN ON THE SIPAP VENTILATOR. THE CUSTOMER REPORTED THERE WAS NO PATIENT INVOLVEMENT ASSOCIATED WITH THE REPORTED EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12200129 ↗
  • InjuryJul 15, 2021Mechanical Problem; Mechanical Problem

    THE CUSTOMER REPORTED THAT THE SIPAP VENTILATOR EXPERIENCED WENT INTO DEMO MODE WHILE ON PATIENT. THE DISPLAY SHOWED UNIT WAS IN LIVE DEMO MODE BUT CONTINUED TO FUNCTION PROPERLY. AS AN INTERVENTION, THEY REMOVED UNIT FROM PATIENT. RE-SET UNIT INTO NORMAL MODE AND PUT THE PATIENT BACK. THE CUSTOMER CONFIRMED THAT THERE WAS NO PATIENT HARM ASSOCIATED WITH THE REPORTED EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12178141 ↗
  • MalfunctionJul 15, 2021Improper Flow or Infusion; Improper Flow or Infusion

    THE CUSTOMER REPORTED FIO2 (FRACTION OF INSPIRED OXYGEN) INACCURACY WITH THE LOW FLOW BLENDER. IT WAS STATED THAT THE DEVIATION IS TOO BIG. THERE IS NO INFORMATION REGARDING PATIENT INVOLVEMENT WITH THE INCIDENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12174938 ↗
  • MalfunctionApr 6, 2021Mechanical Problem; Mechanical Problem

    THE CUSTOMER REPORTED BLENDER FOR REPAIR FROM THE SIPAP VENTILATOR. THE CUSTOMER REPORTED THERE IS NO PATIENT INVOLVEMENT ASSOCIATED WITH THE REPORTED EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11624971 ↗

Other Ventilator, Continuous, Facility Use devices