ILA 3973Recalls & reported problems
Covidien LP · Loading Unit · Class 2 · Implant
All models of ILA (5)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 4
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 10884521063181), received between Jun 14, 2018 and Sep 18, 2019 — classified as 1 injury, 3 malfunctions.
The most frequently reported specific device problem is “Detachment of Device or Device Component” (2 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Most reported problems
Device problems
- Detachment of Device or Device Component2
- Failure to Form Staple2
- Misfire2
- Adverse Event Without Identified Device or Use Problem1
- Failure to Fire1
- Mechanics Altered1
Patient problems
- No Code Available2
- No Known Impact Or Consequence To Patient2
- Cardiac Arrest1
- Death1
- Device Embedded In Tissue or Plaque1
Recent adverse event reports
The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.
InjurySep 18, 2019Mechanics Altered; Adverse Event Without Identified Device or Use Problem; Mechanics Altered; Adverse Event Without Identified Device or Use Problem
ACCORDING TO THE REPORTER, LAPAROSCOPIC WEDGE RESECTION, AN UNKNOWN ISSUE OCCURRED WITH THE DEVICE. SEWING WAS DONE TO RESOLVE THE ISSUE ON ORDER TO COMPLETE THE CASE.
Patient problems: No Code Available
MAUDE report 9085081 ↗MalfunctionDec 3, 2018Misfire; Failure to Form Staple; Detachment of Device or Device Component; Misfire; Failure to Form Staple; Detachment of Device or Device Component
ACCORDING TO THE REPORTER, DURING A VETERINARY PROCEDURE, WHILE BEING APPLIED ON ANASTOMOSIS ON AN OPEN JEJUNOSTOMY, THERE WAS A POOR STAPLE FORMATION WHERE STAPLES DID NOT FORM COMPLETE B SHAPE AND NOTICED AN INCOMPLETE STAPLE LINE. IT WAS ALSO NOTED THAT STAPLES FELL INTO LUMEN. THEY PERFORMED COLIC SURGERY AS SURGICAL INTERVENTION AND THE SURGICAL TIME WAS EXTENDED FOR 25 MINUTES. LATER, THEY CONVERTED TO HAND SEW TO COMPLETE THE CASE. THE PATIENT EXPERIENCED COLIC POST-OPERATIVELY AND EVENTUALLY LED TO CARDIAC TROPONIN. THE PATIENT IS DECEASED.
Patient problems: Cardiac Arrest; Death; Device Embedded In Tissue or Plaque; No Code Available
MAUDE report 8126518 ↗MalfunctionNov 28, 2018Misfire; Failure to Form Staple; Detachment of Device or Device Component; Misfire; Failure to Form Staple; Detachment of Device or Device Component
ACCORDING TO THE REPORTER, DURING A VETERINARY PROCEDURE, WHILE BEING APPLIED ON ANASTOMOSIS ON AN OPEN JEJUNOSTOMY, THERE WAS A POOR STAPLE FORMATION WHERE STAPLES DID NOT FORM COMPLETE B SHAPE AND NOTICED AN INCOMPLETE STAPLE LINE. IT WAS ALSO NOTED THAT STAPLES FELL INTO LUMEN AND WERE PASSED. THEY HAD TO OVER SEW THE STAPLE LINE TO COMPLETE THE CASE. THERE WAS NO PATIENT INJURY.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 8113833 ↗MalfunctionJun 14, 2018Failure to Fire; Failure to Fire
ACCORDING TO THE REPORTER, DURING A SUB TOTAL GASTRECTOMY PROCEDURE, AT THE FIRST FIRING, THE RELOAD WAS CONNECTED TO THE STAPLER, THE SURGEON TRIED TO FIRE THE DEVICE, HOWEVER THE PLASTIC KNOB DID NOT ADVANCE. A NEW RELOAD WAS USED AND CONNECTED THE SAME STAPLER, AND THE PROCEDURE WAS CONTINUED. THERE WAS NO REPORTED PATIENT INJURY.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 7601466 ↗
Other STAPLE, IMPLANTABLE devices
- Endo GIA EGIA60AMTCovidien LP · STAPLE, IMPLANTABLE19 deaths10,680 reports
- Endo GIA Ultra EGIAUSTNDCovidien LP · STAPLE, IMPLANTABLE5 deaths9,730 reports
- Signia SIGPHANDLECovidien LP · STAPLE, IMPLANTABLE1 recall3 deaths5,356 reports
- ECHELON FLEX PSEE60AEthicon Endo-Surgery, LLC · STAPLE, IMPLANTABLE1 recall13 deaths3,294 reports
- Endo GIA Ultra EGIAUSHORTCovidien LP · STAPLE, IMPLANTABLE2 deaths3,011 reports
- Endo GIA EGIA45AMTCovidien LP · STAPLE, IMPLANTABLE10 deaths2,965 reports
- Endo GIA Ultra EGIAUXLCovidien LP · STAPLE, IMPLANTABLE2,909 reports
- ENDOPATH ATS45Ethicon Endo-Surgery, LLC · STAPLE, IMPLANTABLE2 deaths2,046 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.