Skip to content

Novafil 8886 4027-21Recalls & reported problems

Covidien LP · Monofilament Polybutester · Class 2 · Implant

All models of Novafil (39)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
1
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 10884521099494), received between Apr 10, 2024 and Apr 10, 2024 — classified as 1 malfunction.

The most frequently reported specific device problem is “Material Twisted/Bent” (1 report).

2 other recalls cover devices cleared under the same FDA submission (K990952); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 1 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Adverse Event Without Identified Device or Use Problem1
  • Material Twisted/Bent1

Patient problems

  • No Clinical Signs, Symptoms or Conditions1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • MalfunctionApr 10, 2024Material Twisted/Bent; Adverse Event Without Identified Device or Use Problem; Material Twisted/Bent; Adverse Event Without Identified Device or Use Problem

    ACCORDING TO THE REPORTER, DURING RHINOPLASTY, THE NEEDLE OF THE THREAD BECAME DISTORTED WHEN TAKING THE NEEDLE WITH THE NEEDLE HOLDER BY THE SURGEON. ANOTHER SUTURE WAS USED TO RESOLVE THE ISSUE. THERE WAS NO PATIENT INJURY. IT WAS REPORTED THAT THE SURGICAL TIME WAS INCREASED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19079535 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K990952. One submission often covers many models, so check the product and lot details in each record.

  • Z-1175-2023 · Dec 2, 2022 · Open, Classified
    Novafil Monofilament Polybutester Suture Product Description: 8886 4002-41 NOVAFIL 3-0 BLU 90CM C14 8886 4400-13 NOVAFIL 6-0 BLU 45CM SBE2 8886 4400-23 NOVAFIL 5-0 BLU 45CM SBE2 8
  • Z-1181-2023 · Dec 2, 2022 · Open, Classified
    Ti-Cron Coated Braided Polyester Suture Product Description 88862775-31 TICRON* 4-0 BLU 75CM CV327DA 88862880-41 TICRON* 3-0 WHI 75CM Y31DA 88863012-61 TICRON* 0 BLU 90CM C16 X36 8

Other SUTURE, NONABSORBABLE, SYNTHETIC, POLYETHYLENE devices

All SUTURE, NONABSORBABLE, SYNTHETIC, POLYETHYLENE devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page