Ti-Cron 88862947-53Recalls & reported problems

Covidien LP · Coated Braided Polyester · Class 2 · Implant

All models of Ti-Cron (104)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
9
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 9 adverse event reports that list this exact device (DI 10884521105263), received between Sep 16, 2020 and May 1, 2025 — classified as 9 malfunctions.

The most frequently reported specific device problem is “Break” (8 reports).

6 other recalls cover devices cleared under the same FDA submission (K930591); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 7 reportsMay 2025: 1 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520267
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Break8
  • Detachment of Device or Device Component1
  • Material Frayed1

Patient problems

  • No Clinical Signs, Symptoms or Conditions8
  • No Consequences Or Impact To Patient1

Recent adverse event reports

The 9 most recent of 9 reports, in the reporter's own words as published by the FDA.

  • MalfunctionMay 1, 2025Break; Material Frayed; Break; Material Frayed

    IT WAS REPORTED THAT THE SUTURE WAS NOT STRONG ENOUGH, IT BROKE BY SOME FORCE. THE TWO NEEDLES HAVE FRAYED SUTURES COMING FROM THE PROXIMAL ENDS. AN ADDITIONAL LENGTH OF SUTURE WAS ALSO HAS FRAYED AT BOTH ENDS. THE EVENT OCCURRED DURING LAB OR DEMO; THEREFORE, THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21941176 ↗
  • MalfunctionApr 30, 2025Break; Break

    ACCORDING TO THE REPORTER, DURING A KNEE ARTHROSCOPY PROCEDURE, THREE (3) SWAGED NEEDLES WITH SUTURE SNAPPED WHEN THE SURGEON TIED THE KNOT OVER THE CAPSULE. THE SUTURES WERE REMOVED FROM THE PATIENT. THE PROCEDURE WAS SUCCESSFULLY COMPLETED WITH NO DELAY USING A BACK-UP DEVICE. NO FURTHER COMPLICATIONS WERE REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21937252 ↗
  • MalfunctionApr 30, 2025Break; Break

    ACCORDING TO THE REPORTER, DURING A KNEE ARTHROSCOPY PROCEDURE, THREE (3) SWAGED NEEDLES WITH SUTURE SNAPPED WHEN THE SURGEON TIED THE KNOT OVER THE CAPSULE. THE SUTURES WERE REMOVED FROM THE PATIENT. THE PROCEDURE WAS SUCCESSFULLY COMPLETED WITH NO DELAY USING A BACK-UP DEVICE. NO FURTHER COMPLICATIONS WERE REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21937216 ↗
  • MalfunctionApr 30, 2025Break; Break

    ACCORDING TO THE REPORTER, DURING A KNEE ARTHROSCOPY PROCEDURE, THREE (3) SWAGED NEEDLES WITH SUTURE SNAPPED WHEN THE SURGEON TIED THE KNOT OVER THE CAPSULE. THE SUTURES WERE REMOVED FROM THE PATIENT. THE PROCEDURE WAS SUCCESSFULLY COMPLETED WITH NO DELAY USING A BACK-UP DEVICE. NO FURTHER COMPLICATIONS WERE REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21937158 ↗
  • MalfunctionApr 30, 2025Break; Break

    IT WAS REPORTED THAT DURING AN ARTHROSCOPY PROCEDURE, THE BLUE SUTURE OF THREE UNITS OF THE DEVICE BROKE WHILE TIGHTENING THE KNOT. THE PROCEDURE WAS SUCCESSFULLY COMPLETED WITH A COMPETITOR'S DEVICE AND A SURGICAL DELAY OF LESS THAN 30 MINUTES. NO FURTHER COMPLICATIONS WERE REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21937037 ↗
  • MalfunctionApr 30, 2025Break; Break

    IT WAS REPORTED THAT DURING AN ARTHROSCOPY PROCEDURE, THE BLUE SUTURE OF THREE UNITS OF THE DEVICE BROKE WHILE TIGHTENING THE KNOT. THE PROCEDURE WAS SUCCESSFULLY COMPLETED WITH A COMPETITOR'S DEVICE AND A SURGICAL DELAY OF LESS THAN 30 MINUTES. NO FURTHER COMPLICATIONS WERE REPORTED.KL 4/30/2025

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21937013 ↗
  • MalfunctionApr 30, 2025Break; Break

    IT WAS REPORTED THAT DURING AN ARTHROSCOPY PROCEDURE, THE BLUE SUTURE OF THREE UNITS OF THE DEVICE BROKE WHILE TIGHTENING THE KNOT. THE PROCEDURE WAS SUCCESSFULLY COMPLETED WITH A COMPETITOR'S DEVICE AND A SURGICAL DELAY OF LESS THAN 30 MINUTES. NO FURTHER COMPLICATIONS WERE REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21936806 ↗
  • MalfunctionApr 10, 2025Break; Break

    IT WAS REPORTED THAT DURING AN ARTHROSCOPY PROCEDURE, THE BLUE SUTURE OF THREE (3) UNITS OF THE TI-CRON BROKE WHILE TIGHTENING THE KNOT. THE PROCEDURE WAS SUCCESSFULLY COMPLETED WITH A COMPETITOR'S DEVICE AND A SURGICAL DELAY OF LESS THAN 30 MINUTES. NO FURTHER COMPLICATIONS WERE REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21820673 ↗
  • MalfunctionSep 16, 2020Detachment of Device or Device Component; Detachment of Device or Device Component

    ACCORDING TO THE REPORTER, DURING KNEE ARTHROPLASTY, UPON PERFORMING MENISCUS SUTURE, THREAD DROPPED FROM THE NEEDLE. ANOTHER THREAD WAS USED TO COMPLETE THE CASE. THERE WAS NO PATIENT INJURY.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 10540231 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K930591. One submission often covers many models, so check the product and lot details in each record.

  • Z-2830-2011 · Jun 3, 2011 · Terminated
    Syneture Ti-Cron 0 5X18" BLUE T-40/HGS-22 D-TACH, non-absorbable polyester sutures Material 8886337962 Ti-Cron is intended for use in general soft tissue approximation and/or lig
  • Z-2827-2011 · Jun 3, 2011 · Terminated
    Syneture Ti-Cron 1 Blue 5x18 HGS-22 D-Tach non-absorbable polyester sutures Material 8886326062 Ti-Cron is intended for use in general soft tissue approximation and/or ligation, i
  • Z-2825-2011 · Jun 3, 2011 · Terminated
    Syneture Ti-Cron 0 BLUE 6X18" PRE-CUT non-absorbable polyester sutures Material 8886300162 Ti-Cron is intended for use in general soft tissue approximation and/or ligation, inclu
  • Z-2828-2011 · Jun 3, 2011 · Terminated
    Syneture Ti-Cron 1 Blue 5x18 HGS-22 D-Tach non-absorbable polyester sutures Material 8886337272 Ti-Cron is intended for use in general soft tissue approximation and/or ligation, i
  • Z-2826-2011 · Jun 3, 2011 · Terminated
    Syneture Ti-Cron 0 5X18" BLUE T-19/GS-22 D-TACH non-absorbable polyester sutures Material 8886325962 Ti-Cron is intended for use in general soft tissue approximation and/or ligat
  • Z-2829-2011 · Jun 3, 2011 · Terminated
    Syneture Ti-Cron 2 BLUE 5X18" HGS-22 non-absorbable polyester sutures Material 8886337282 Ti-Cron is intended for use in general soft tissue approximation and/or ligation, includi

Other SUTURE, NONABSORBABLE, SYNTHETIC, POLYETHYLENE devices

All SUTURE, NONABSORBABLE, SYNTHETIC, POLYETHYLENE devices →