Skip to content

superDimension AAS00161-20RRecalls & reported problems

Covidien LP · Navigation System · Class 2

All models of superDimension (20)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
1
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 10884521199996), received between Oct 27, 2023 and Oct 27, 2023 — classified as 1 injury.

The most frequently reported specific device problem is “Application Program Problem” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Application Program Problem1
  • Failure to Power Up1

Patient problems

  • Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • InjuryOct 27, 2023Failure to Power Up; Application Program Problem; Failure to Power Up; Application Program Problem

    ACCORDING TO THE REPORTER, DURING A PROCEDURE, THE LOCATABLE SUB-SYSTEM (LSS) WAS NOT POWERING ON. THERE WAS POWER TO THE INDUSTRIAL PERSONAL COMPUTER (IPC) AND MONITOR. THE USER COULD NOT OPEN THE BACK PANEL TO CHECK THE POWER OF LSS. THE PHYSICIAN WAS NOT ABLE TO COMPLETE THE ELECTROMAGNETIC NAVIGATIONAL BRONCHOSCOPY (ENB) PORTION OF THE CASE EVEN BY OTHER MEANS. THE PATIENT WAS UNDER ANESTHESIA. IT WAS NOTED THAT SINCE THE SYSTEM WAS NOT FIXABLE AND ANOTHER DEVICE WAS CHOSEN BY THE FACILITY, RESCHEDULED PROCEDURE WAS UNCERTAIN.

    Patient problems: Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available

    MAUDE report 18022937 ↗

Other System, x-ray, tomography, computed devices

All System, x-ray, tomography, computed devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page