Skip to content

Edge SDK4450-FTRecalls & reported problems

Covidien LP · Firm Tip; Endobronchial Procedure Kit;45 Catheter · Class 2

All models of Edge (43)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
2
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 10884521545564), received between Apr 3, 2019 and Nov 6, 2024 — classified as 2 malfunctions.

The most frequently reported specific device problem is “Material Fragmentation” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 1 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Material Fragmentation1
  • Patient Data Problem1
  • Use of Device Problem1

Patient problems

  • No Clinical Signs, Symptoms or Conditions1
  • No Consequences Or Impact To Patient1

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • MalfunctionNov 6, 2024Use of Device Problem; Patient Data Problem; Use of Device Problem; Patient Data Problem

    ACCORDING TO THE REPORTER, DURING A PROCEDURE, THE LOCATABLE GUIDE HAD INABILITY TO DEFLECT TIP. DUE TO THIS, THE PATIENT COULD NOT REGISTER DURING THE WORKFLOW. THE PHYSICIAN WAS ABLE TO COMPLETE THE ELECTROMAGNETIC NAVIGATIONAL BRONCHOSCOPY (ENB) PORTION OF THECASE. THERE WAS NO PATIENT INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20617039 ↗
  • MalfunctionApr 3, 2019Material Fragmentation; Material Fragmentation

    ACCORDING TO THE REPORTER, DURING A PROCEDURE, WHEN PHYSICIAN WENT TO SAMPLE USING A NEEDLE A PLASTIC STRINGY SUBSTANCE WAS EXPELLED FROM DISTAL END OF CATHETER. PHYSICIAN RETRIEVED THE STRINGY SUBSTANCE AND REMOVED THE EXTENDED WORKING CHANNEL (EWC) ALL TOGETHER. CASE WAS COMPLETED AND THERE WAS NO PATIENT INJURY.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 8477789 ↗

Other System, x-ray, tomography, computed devices

All System, x-ray, tomography, computed devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page