Skip to content

Ti-Cron VCD20316MFERecalls & reported problems

Covidien LP · Coated Braided Polyester, CV-316, D/A, 4Blue/4White Strands with Firm PTFE Pledget 3x6x1.85 · Class 2 · Implant

All models of Ti-Cron (104)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
1
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 10884521550582), received between Jul 11, 2025 and Jul 11, 2025 — classified as 1 malfunction.

The most frequently reported specific device problem is “Detachment of Device or Device Component” (1 report).

6 other recalls cover devices cleared under the same FDA submission (K930591); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 1 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Detachment of Device or Device Component1

Patient problems

  • No Clinical Signs, Symptoms or Conditions1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJul 11, 2025Detachment of Device or Device Component; Detachment of Device or Device Component

    IT WAS REPORTED THAT DURING THE MITRAL VALVE REPLACEMENT, WHILE SUTURING, THE NEEDLE DETACHED FROM THE SUTURE AND FELL INTO THE PATIENT CAVITY. THE NEEDLE WAS FOUND AND RETRIEVED USING A LAPAROSCOPE. AN ADDITIONAL NEW SUTURE WAS USED WITHOUT ISSUE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22485621 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K930591. One submission often covers many models, so check the product and lot details in each record.

  • Z-2830-2011 · Jun 3, 2011 · Terminated
    Syneture Ti-Cron 0 5X18" BLUE T-40/HGS-22 D-TACH, non-absorbable polyester sutures Material 8886337962 Ti-Cron is intended for use in general soft tissue approximation and/or lig
  • Z-2828-2011 · Jun 3, 2011 · Terminated
    Syneture Ti-Cron 1 Blue 5x18 HGS-22 D-Tach non-absorbable polyester sutures Material 8886337272 Ti-Cron is intended for use in general soft tissue approximation and/or ligation, i
  • Z-2829-2011 · Jun 3, 2011 · Terminated
    Syneture Ti-Cron 2 BLUE 5X18" HGS-22 non-absorbable polyester sutures Material 8886337282 Ti-Cron is intended for use in general soft tissue approximation and/or ligation, includi
  • Z-2827-2011 · Jun 3, 2011 · Terminated
    Syneture Ti-Cron 1 Blue 5x18 HGS-22 D-Tach non-absorbable polyester sutures Material 8886326062 Ti-Cron is intended for use in general soft tissue approximation and/or ligation, i
  • Z-2826-2011 · Jun 3, 2011 · Terminated
    Syneture Ti-Cron 0 5X18" BLUE T-19/GS-22 D-TACH non-absorbable polyester sutures Material 8886325962 Ti-Cron is intended for use in general soft tissue approximation and/or ligat
  • Z-2825-2011 · Jun 3, 2011 · Terminated
    Syneture Ti-Cron 0 BLUE 6X18" PRE-CUT non-absorbable polyester sutures Material 8886300162 Ti-Cron is intended for use in general soft tissue approximation and/or ligation, inclu

Other SUTURE, NONABSORBABLE, SYNTHETIC, POLYETHYLENE devices

All SUTURE, NONABSORBABLE, SYNTHETIC, POLYETHYLENE devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page