EEA TRIEEAXL25XTRecalls & reported problems
Covidien LP · Black Extra Thick XL Circular Stapler with Tri-Staple Technology · Class 2
All models of EEA (39)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 8
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 8 adverse event reports that list this exact device (DI 10884521593114), received between Apr 7, 2026 and Sep 18, 2026 — classified as 1 injury, 7 malfunctions.
The most frequently reported specific device problem is “Device Difficult to Setup or Prepare” (3 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Reports received per month
Most reported problems
Device problems
- Device Difficult to Setup or Prepare3
- Detachment of Device or Device Component1
- Failure to Fire1
- Incomplete or Inadequate Connection1
- Mechanics Altered1
- Misfire1
Patient problems
- No Clinical Signs, Symptoms or Conditions6
- Failure to Anastomose1
- Insufficient Information1
- Unintended Radiation Exposure1
Recent adverse event reports
The 8 most recent of 8 reports, in the reporter's own words as published by the FDA.
MalfunctionSep 18, 2026Detachment of Device or Device Component
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50829223 ↗MalfunctionSep 15, 2026Failure to Fire
It was reported that during an end-to-end anastomosis procedure, the instrument did not fire. The device was replaced with a new device to complete the procedure. No patient injury was reported.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50766507 ↗MalfunctionAug 5, 2026Misfire
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50136935 ↗MalfunctionMay 1, 2026Incomplete or Inadequate Connection
IT WAS REPORTED THAT DURING ROUX EN Y PROCEDURE USING THE CIRCULAR STAPLER, THE TROCAR TIP FELL OFF THE ANVIL WHILE BEING PASSED THROUGH THE GASTRIC TISSUE, AND THE SURGICAL TEAM HAD TO LOCATE THE TROCAR TIP AND RETRIEVED BY USING ANOTHER BAG. IT WAS ALSO NOTED THAT THE SURGICAL TIME WAS EXTENDED. THE PATIENT WAS REPORTED TO BE ALIVE WITH NO INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25054844 ↗MalfunctionApr 21, 2026Device Difficult to Setup or Prepare
IT WAS REPORTED THAT DURING THE LAPAROSCOPIC ROUX-EN-Y PROCEDURE, THE TROCAR TIP CAME OFF WHILE PUNCTURING THROUGH GASTRIC TISSUE. THE SURGICAL TIME WAS EXTENDED FOR LESS THAN 30 MINUTES WHILE THE DETACHED TROCAR TIP WAS LOCATED, AND REMOVED. THERE WAS NO PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24946262 ↗InjuryApr 21, 2026Device Difficult to Setup or Prepare
IT WAS REPORTED THAT DURING THE LAPAROSCOPIC ROUX-EN-Y REVISION, DURING PLACEMENT OF THE ANVIL THROUGH THE STOMACH, THE TROCAR TIP FELL OFF INTO THE BODY CAVITY WHILE TRYING TO PIERCE THE TISSUE. THE CASE WAS DELAYED AS X-RAY WAS DONE TO LOCATE AND REMOVE THE DEVICE WITH SPECIMEN BAG. THE CURRENT STAPLER WAS THEN USED TO COMPLETE THE CASE.
Patient problems: Unintended Radiation Exposure
MAUDE report 24945926 ↗MalfunctionApr 13, 2026Device Difficult to Setup or Prepare
IT WAS REPORTED THAT, DURING A LAPAROSCOPIC ROUX-EN-Y PROCEDURE IN A PATIENT WITH A HISTORY OF OBESITY, THERE WAS A PROCEDURAL IRREGULARITY WHEN REMOVING THE TROCAR TIP FROM THE ANVIL, AS THE SURGEON COMMENTED THAT THE WHITE TROCAR TIP WAS REMOVED MUCH EASIER. NO INTERVENTION WAS DONE OR REQUIRED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24853760 ↗MalfunctionApr 7, 2026Mechanics Altered
ACCORDING TO THE REPORTER, DURING A LAPAROSCOPIC LOW ANTERIOR RESECTION, DURING CREATION OF THE ANASTOMOSIS, THE CIRCULAR STAPLER CUT BUT THE STAPLES DID NOT DEPLOY. THE SURGEON MANUALLY SUTURED THE ANASTOMOSIS, AND THE SURGICAL TIME WAS EXTENDED BY ONE HOUR.
Patient problems: Failure to Anastomose; Insufficient Information
MAUDE report 24798987 ↗
Other Staple, implantable devices
- Endo GIA EGIA60AMTCovidien LP · STAPLE, IMPLANTABLE19 deaths10,680 reports
- Endo GIA Ultra EGIAUSTNDCovidien LP · STAPLE, IMPLANTABLE5 deaths9,730 reports
- Signia SIGPHANDLECovidien LP · STAPLE, IMPLANTABLE1 recall3 deaths5,356 reports
- ECHELON FLEX PSEE60AEthicon Endo-Surgery, LLC · STAPLE, IMPLANTABLE1 recall13 deaths3,294 reports
- Endo GIA Ultra EGIAUSHORTCovidien LP · STAPLE, IMPLANTABLE2 deaths3,011 reports
- Endo GIA EGIA45AMTCovidien LP · STAPLE, IMPLANTABLE10 deaths2,965 reports
- Endo GIA Ultra EGIAUXLCovidien LP · STAPLE, IMPLANTABLE2,909 reports
- ENDOPATH ATS45Ethicon Endo-Surgery, LLC · STAPLE, IMPLANTABLE2 deaths2,046 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.