Skip to content

EEA TRIEEAXL31MTRecalls & reported problems

Covidien LP · Circular Stapler with Tri-Staple Technology · Class 2

Adverse event reports
3
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 10884521593145), received between Dec 30, 2021 and Jan 10, 2025 — classified as 1 injury, 2 malfunctions.

The most frequently reported specific device problem is “Break” (1 report).

1 other recall cover devices cleared under the same FDA submission (K172361, K221005); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 1 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 1 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Break1
  • Device Slipped1
  • Failure to Align1
  • Failure to Form Staple1
  • Incomplete or Inadequate Connection1
  • Positioning Problem1

Patient problems

  • No Clinical Signs, Symptoms or Conditions2
  • Failure to Anastomose1
  • Tissue Breakdown1

Recent adverse event reports

The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.

  • InjuryJan 10, 2025Break; Device Slipped; Failure to Form Staple; Break; Device Slipped; Failure to Form Staple

    IT WAS REPORTED THAT DURING A LOW LEFT COLON RESECTION INVOLVING THE CIRCULAR STAPLER, THE SAFETY LEVER OF THE STAPLER WAS BROKEN LIKELY DUE TO INCOMPLETE CLOSING CAUSED BY THE PRESENCE OF EXCESSIVE TISSUE. ALTHOUGH THE BLADE WORKED, BUT THE STITCHES WERE BEING SHOT OPEN, THE STAPLES WERE NOT FORMED B SHAPED BUT A BIT LINEAR. SUTURING WAS PERFORMED AS A REPLACEMENT FOR THE STAPLE LINE. A PROTECTIVE STOMA CREATION WAS NECESSARY. THE SURGICAL TIME WAS EXTENDED FOR ONE HOUR AS A RESULT.

    Patient problems: Failure to Anastomose; Tissue Breakdown

    MAUDE report 21128392 ↗
  • MalfunctionSep 13, 2024Positioning Problem; Incomplete or Inadequate Connection; Positioning Problem; Incomplete or Inadequate Connection

    ACCORDING TO THE REPORTER, DURING A COLORECTAL ANASTOMOSIS, WHEN THE SURGEON INSERTED THE DEVICE, THE TIP DID NOT COME OUT OF THE INSTRUMENT. IT WAS NOT POSSIBLE TO CONNECT THE ANVIL OF THE DEVICE. IT WAS NOTED THAT THERE WAS A STRUGGLE TO TURN THE BLACK WING NUT. THE SURGERY WAS EXTENDED BY 10 MINUTES. THE PRODUCT WAS CHANGED TO RESOLVE THE ISSUE. THERE WAS NO PATIENT INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20217240 ↗
  • MalfunctionDec 30, 2021Failure to Align; Failure to Align

    ACCORDING TO THE REPORTER, DURING A LAPAROSCOPIC ROBOTIC LEFT COLECTOMY, ON INSERTION OF THE ANVIL THROUGH A SMALL INCISION BY THE PHYSICIAN ASSISTANT, THE ANVIL TILTED PREMATURELY. THE SURGEON NOTICED IT ONCE THE ANVIL WAS INSIDE THE ABDOMINAL CAVITY. THE ANVIL WAS THEN REMOVED. A SECOND STAPLER WAS OPENED AND A NEW ANVIL WAS USED TO RESOLVE THE ISSUE. THERE WAS NO PATIENT INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 13119502 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K172361, K221005. One submission often covers many models, so check the product and lot details in each record.

  • Z-0070-2019 · Aug 17, 2018 · Terminated
    EEA Circular Stapler with Tri-Staple Technology, Part No. TRIEEA28MT TRIEEA28XT TRIEEA31MT TRIEEA31XT Product Usage: The EEA Circular Stapler with Tri-Staple Technology places a

Other Staple, implantable devices

All Staple, implantable devices →

Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.