GIA GIA80MTSRecalls & reported problems

Covidien LP · Medium/Thick Stapler with Tri-Staple Technology · Class 2

All models of GIA (26)

Data updated Nov 27, 2026 from official FDA records · Editorial policy

Quick summary

This Covidien GIA medium-thick surgical stapler has 263 adverse event reports, including 208 malfunctions, 54 injuries and 1 death. Reports mostly describe the stapler not firing or not forming staples, misfiring, and parts coming loose during surgery, and some mention tissue breakdown, bleeding, and failed connections between tissues. There are no recalls that name this device.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
261
Deaths reported
1
Injuries reported
53
Recalls naming it
0

Summary

The FDA's MAUDE database contains 261 adverse event reports that list this exact device (DI 10884521723450), received between Mar 30, 2021 and Sep 17, 2026 — classified as 1 death, 53 injuries, 207 malfunctions.

The most frequently reported specific device problem is “Failure to Fire” (67 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.

Reports received per month

Dec 2023: 2 reportsJan 2024: 9 reportsFeb 2024: 4 reportsMar 2024: 1 reportsApr 2024: 5 reportsMay 2024: 5 reportsJun 2024: 3 reportsJul 2024: 5 reportsAug 2024: 7 reportsSep 2024: 8 reportsOct 2024: 2 reportsNov 2024: 6 reportsDec 2024: 4 reportsJan 2025: 5 reportsFeb 2025: 6 reportsMar 2025: 7 reportsApr 2025: 4 reportsMay 2025: 1 reportsJun 2025: 5 reportsJul 2025: 10 reportsAug 2025: 3 reportsSep 2025: 5 reportsOct 2025: 4 reportsNov 2025: 2 reportsDec 2025: 13 reportsJan 2026: 4 reportsFeb 2026: 4 reportsMar 2026: 10 reportsApr 2026: 1 reportsMay 2026: 2 reportsJun 2026: 1 reportsJul 2026: 1 reportsAug 2026: 4 reportsSep 2026: 4 reportsOct 2026: 0 reportsNov 2026: 0 reports20242025202613
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Failure to Fire67
  • Adverse Event Without Identified Device or Use Problem50
  • Failure to Form Staple41
  • Misfire38
  • Detachment of Device or Device Component34
  • Break20
  • Failure to Cut10
  • Mechanics Altered10

Patient problems

  • No Clinical Signs, Symptoms or Conditions137
  • Tissue Breakdown44
  • Failure to Anastomose40
  • Insufficient Information38
  • Unspecified Tissue Injury31
  • Hemorrhage/Blood Loss/Bleeding17
  • Bowel Perforation9
  • Unspecified Infection8

Recent adverse event reports

The 25 most recent of 261 reports, in the reporter's own words as published by the FDA.

  • MalfunctionSep 17, 2026Misfire

    According to the reporter, during a transverse colectomy, when performing functional end-to-end (FEEA) anastomosis, when it was used for closure of the insertion hole on the fourth firing, the firing knob became too hard midway and could no longer be advanced. Upon confirmation, forceps and the like were not caught inside. Hence, it could not be advanced; the firing knob was returned. To resolve the issue, a new stapler and cartridge were taken out and used, and resection was performed.

    Patient problems: Failure to Anastomose

    MAUDE report 50805313 ↗
  • MalfunctionSep 16, 2026Mechanics Altered

    Patient problems: Tissue Breakdown

    MAUDE report 50793267 ↗
  • MalfunctionSep 15, 2026Detachment of Device or Device Component

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50778225 ↗
  • InjurySep 12, 2026Adverse Event Without Identified Device or Use Problem

    According to the reporter, on (b)(6) 2026, post-operative of a right colectomy under coelio performed on (b)(6) 2026, an emergency CT (computed tomography) scan was done to the patient which also had Hb drop (Hemoglobin drop) and generalized defense. The patient had a low-noise anastomotic bleeding. The patient underwent re-operation on the same day to resolve the issue, and was then put on antibiotics. The patient was discharged on (b)(6) 2026.

    Patient problems: Failure to Anastomose; Hemorrhage/Blood Loss/Bleeding; Pain; Tissue Breakdown; Unintended Radiation Exposure; Unspecified Blood or Lymphatic problem

    MAUDE report 50745197 ↗
  • MalfunctionAug 18, 2026Misfire

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50307145 ↗
  • MalfunctionAug 18, 2026Failure to Form Staple

    Patient problems: Tissue Breakdown

    MAUDE report 50303893 ↗
  • MalfunctionAug 14, 2026Misfire

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50256395 ↗
  • MalfunctionAug 7, 2026Misfire; Mechanical Jam

    Patient problems: Insufficient Information

    MAUDE report 50168713 ↗
  • MalfunctionJul 29, 2026Mechanics Altered

    Patient problems: Tissue Breakdown

    MAUDE report 50065718 ↗
  • MalfunctionJun 22, 2026Failure to Form Staple; Mechanics Altered

    IT WAS REPORTED THAT, DURING A LIVER SURGERY, THE STAPLER WAS USED FOR AN ANASTOMOSIS, THE STAPLES DID NOT DEPLOYED TO FORM A PROPER B-SHAPE AND THE BLADE CUT TISSUE. BLEEDING WAS REPORTED BECAUSE THE TISSUE WAS CUT WITHOUT STAPLING. THE DEVICE WAS REPLACED DURING THE PROCEDURE BY USING ANOTHER STAPLER. THE PROCEDURE WAS PROLONGED.

    Patient problems: Hemorrhage/Blood Loss/Bleeding; Tissue Breakdown

    MAUDE report 25587718 ↗
  • MalfunctionMay 4, 2026Detachment of Device or Device Component

    IT WAS REPORTED THAT DURING A PROCEDURE, WHEN USING THE DEVICE, THE PUSHER THUMB PIECE POPPED OFF OF THE SLED, AND IT DISENGAGED FROM THE STAPLER. THE DEVICE WAS REPLACED TO RESOLVE THE ISSUE. THERE WAS NO PATIENT INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25072844 ↗
  • MalfunctionMay 4, 2026Mechanics Altered

    IT WAS REPORTED THAT DURING A WHIPPLE PROCEDURE INVOLVING RESECTION OF THE STOMACH, THE STAPLES DID NOT DEPLOY (CUT WITHOUT STAPLING). THE DEVICE WAS REPLACED WITH ANOTHER MANUFACTURER¿S STAPLER TO RESOLVE THE ISSUE.

    Patient problems: Tissue Breakdown

    MAUDE report 25070839 ↗
  • MalfunctionApr 20, 2026Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT DURING A GASTRIC RESECTION PROCEDURE, AN OPEN STAPLER WAS USED AND, AFTER FIRING, THE STAPLE LINE BURST OPEN AND A STAPLE LINE LEAK WAS OBSERVED WITH LEAKAGE OF THE CONTENTS BEING HELD BY THE TISSUE, WHICH WAS DISCOVERED DURING THE PROCEDURE. A LEAK TEST RESULT WAS RECORDED AS NOT AVAILABLE. THE REPORTED PATIENT IMPACT WAS UNKNOWN, AND NO PATIENT SYMPTOMS WERE REPORTED.

    Patient problems: Insufficient Information

    MAUDE report 24923182 ↗
  • MalfunctionMar 23, 2026Failure to Form Staple; Insufficient Device Problem Information

    ACCORDING TO THE REPORTER, DURING A COLON RESECTION, THE STAPLE LINE FROM THE STAPLER AND TWO ADDITIONAL RELOADS WAS INCOMPLETE. TO RESOLVE THE ISSUE, A PURPLE RELOAD, WHICH WAS NOT INDICATED FOR USE IN THE COLON, WAS UTILIZED TO COMPLETE THE COLON RESECTION.

    Patient problems: Tissue Breakdown

    MAUDE report 24664910 ↗
  • MalfunctionMar 23, 2026Detachment of Device or Device Component

    IT WAS REPORTED THAT, DURING A PROCEDURE, PRIOR TO FIRING, THE TIP PART OF THE STAPLER WAS DAMAGED DURING DISASSEMBLY OF THE BASE MOUNTED CARTRIDGE AND BROKE OFF. ANOTHER SAME TYPE OF INSTRUMENT WAS USED TO RESOLVE THE ISSUE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24663563 ↗
  • MalfunctionMar 20, 2026Protective Measures Problem

    IT WAS REPORTED THAT PREOPERATIVELY, WHEN THE DEVICE WAS PASSED TO THE SURGEON, THE BLADE WAS ALREADY ON WITHOUT ACTIVATING THE DEVICE. THE CARTRIDGE WAS REPLACED TO RESOLVE THE ISSUE. THERE WAS NO PATIENT INVOLVED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24656515 ↗
  • MalfunctionMar 13, 2026Failure to Fire

    IT WAS REPORTED THAT DURING A DIGESTIVE SURGERY, WHEN THE SURGEON ATTEMPTED TO STAPLE, THE RELOAD WAS STUCK AND COULD NOT BE USED. ANOTHER DEVICE WAS REPLACED TO RESOLVE THE ISSUE. THERE WAS NO PATIENT INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24592559 ↗
  • MalfunctionMar 11, 2026Break

    IT WAS REPORTED THAT DURING A PROCEDURE, WHILE THE TECHNICIAN WAS ASSEMBLING THE DEVICE, IT WAS OBSERVED THAT THE AREA WHERE THE RELOAD MECHANISM WAS INSERTED WAS BROKEN. THE DEVICE WAS NOT THEN USED. TO RESOLVE THE ISSUE, A NEW STAPLER WAS USED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24570555 ↗
  • MalfunctionMar 7, 2026Failure to Fire

    IT WAS REPORTED THAT DURING A SURGICAL PROCEDURE INVOLVING THE OPEN STAPLER, THE INSTRUMENT DID NOT FIRE, WHICH RESULTED IN AN EXTENDED SURGICAL TIME OF 30 MINUTES. SUTURING WAS PERFORMED AS A STAPLE LINE REPLACEMENT TO ADDRESS THE ISSUE.

    Patient problems: Tissue Breakdown

    MAUDE report 24540731 ↗
  • MalfunctionMar 7, 2026Failure to Fire

    IT WAS REPORTED THAT DURING A SURGICAL PROCEDURE INVOLVING THE OPEN STAPLER, THE INSTRUMENT DID NOT FIRE, WHICH RESULTED IN AN EXTENDED SURGICAL TIME OF 30 MINUTES. SUTURING WAS PERFORMED AS A STAPLE LINE REPLACEMENT TO ADDRESS THE ISSUE.

    Patient problems: Tissue Breakdown

    MAUDE report 24540518 ↗
  • MalfunctionMar 7, 2026Failure to Fire

    IT WAS REPORTED THAT DURING A SURGICAL PROCEDURE INVOLVING THE OPEN STAPLER, THE INSTRUMENT DID NOT FIRE, WHICH RESULTED IN AN EXTENDED SURGICAL TIME OF 30 MINUTES. SUTURING WAS PERFORMED AS A STAPLE LINE REPLACEMENT TO ADDRESS THE ISSUE.

    Patient problems: Tissue Breakdown

    MAUDE report 24539738 ↗
  • MalfunctionMar 4, 2026Failure to Form Staple; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT DURING HEMICOLECTOMY AND ILEOCOLIC ANASTOMOSIS, AN INCOMPLETE STAPLE LINE WAS NOTED. THERE WAS NO CONTINUITY OF STAPLES IN TWO SEGMENTS, RESULTING IN INTESTINAL CONTENT LEAKAGE. THE AFFECTED INTESTINAL SEGMENT REQUIRED MANUAL RESECTION AND ANASTOMOSIS.

    Patient problems: Failure to Anastomose

    MAUDE report 24512953 ↗
  • MalfunctionMar 4, 2026Failure to Form Staple; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT DURING A SURGICAL PROCEDURE, THE STAPLE LINE DID NOT CLOSE. THIS PROCEDURAL ISSUE RESULTED IN AN EXTENDED OPERATING TIME, AS THE SURGEON ELECTED TO SUTURE THE STAPLE LINE WITH THREAD RATHER THAN APPLY NEW STAPLES.

    Patient problems: Tissue Breakdown

    MAUDE report 24511537 ↗
  • MalfunctionFeb 18, 2026Failure to Form Staple

    IT WAS REPORTED THAT DURING A RIGHT HEMICOLECTOMY, DURING ILEAL DETACHMENT, THE STAPLE HAD NOT FULLY CLOSED WHEN THE BLADE FIRED. MANUAL SUTURING WAS REQUIRED TO COMPLETE THE CASE.

    Patient problems: Tissue Breakdown

    MAUDE report 24381718 ↗
  • MalfunctionFeb 17, 2026Misfire; Failure to Fire

    IT WAS REPORTED THAT DURING AN OPEN HEMICOLECTOMY SURGERY, THE AFTER FIRING ONE RELOAD SUCCESSFULLY, THE NEXT RELOAD DID NOT FIRE AT ALL. THE SURGEON REPLACED THE HANDLE AND RELOAD. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24365337 ↗

FDA clearance and classification

Cleared through Special 510(k) K221006 on Apr 29, 2022 (decision: Substantially Equivalent). Submitted by Covidien as “GIA Stapler with Tri-Staple Technology”. The FDA review took 24 days. FDA record ↗

Cleared through 510(k) K192720 on Mar 13, 2020 (decision: Substantially Equivalent). Submitted by Covidien as “GIA Stapler with Tri-Staple Technology”. The FDA review took 168 days. FDA record ↗

The FDA regulates product code GDW (staple, implantable) under 21 CFR 878.4750 as a Class II device. 21 CFR 878.4750 ↗

Common questions about the GIA GIA80MTS

Has the GIA GIA80MTS been recalled?

No FDA recall record in our data names this exact device identifier (DI 10884521723450), based on the FDA recall file dated Nov 27, 2026. Recalls are sometimes announced by product family or lot, so the manufacturer’s own notices are worth checking too.

What problems have been reported with the GIA GIA80MTS?

The FDA's MAUDE database lists 261 adverse event reports for this device, received between Mar 30, 2021 and Sep 17, 2026. In those reports, the most common specific device problems are “Failure to Fire” (67), “Failure to Form Staple” (41) and “Misfire” (38); the most common patient problems are “Tissue Breakdown” (44), “Failure to Anastomose” (40) and “Unspecified Tissue Injury” (31). Reports are not verified by the FDA and do not prove the device caused the problem.

Have deaths or injuries been reported with the GIA GIA80MTS?

1 report is classified as a death, 53 as injuries and 207 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.

Is the GIA GIA80MTS FDA approved?

It was cleared through 510(k) K192720 on Mar 13, 2020; its record lists 2 FDA submissions in total. Strictly, the FDA "clears" 510(k) devices as substantially equivalent to devices already on the market rather than "approving" them; approval is the PMA route.

Is the GIA GIA80MTS MRI safe?

According to the device's FDA GUDID record, its labeling lists it as MR Conditional: it can be scanned only under the specific conditions given in the labeling. The radiology team should check those conditions before an MRI.

Who makes the GIA GIA80MTS, and is it still sold?

The FDA's device database lists Covidien LP as the company responsible for it (brand GIA, model GIA80MTS). Its record shows it as in commercial distribution.