ILLUMISITE ILS-1000-CS (UDI 10884521723634)Recalls & reported problems

Covidien LP · Console · Class 2

All models of ILLUMISITE (7)

Data updated Nov 27, 2026 from official FDA records · Editorial policy

Quick summary

This Covidien Illumisite console is an x-ray-based computed tomography system. The 103 reports mostly describe software and connection problems, such as blank or error screens, disconnected location boards, and aborted procedures; 1 death and 75 injuries are described, including pneumothorax, hemorrhage and cardiac arrest. No recalls name this device. Reports fell from 20 in the prior 12 months to 4 in the last 12 months.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
104
Deaths reported
1
Injuries reported
76
Recalls naming it
0

Summary

The FDA's MAUDE database contains 104 adverse event reports that list this exact device (DI 10884521723634), received between Jan 30, 2020 and Sep 15, 2026 — classified as 1 death, 76 injuries, 27 malfunctions.

The most frequently reported specific device problem is “Application Program Problem” (51 reports).

1 other recall cover devices cleared under the same FDA submission (K191394); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.

Reports received per month

Dec 2023: 4 reportsJan 2024: 1 reportsFeb 2024: 2 reportsMar 2024: 1 reportsApr 2024: 0 reportsMay 2024: 2 reportsJun 2024: 6 reportsJul 2024: 1 reportsAug 2024: 0 reportsSep 2024: 1 reportsOct 2024: 2 reportsNov 2024: 2 reportsDec 2024: 1 reportsJan 2025: 3 reportsFeb 2025: 1 reportsMar 2025: 2 reportsApr 2025: 3 reportsMay 2025: 0 reportsJun 2025: 1 reportsJul 2025: 3 reportsAug 2025: 2 reportsSep 2025: 1 reportsOct 2025: 1 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 2 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 2 reportsSep 2026: 1 reportsOct 2026: 0 reportsNov 2026: 0 reports2024202520266
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Application Program Problem51
  • Device Displays Incorrect Message37
  • Adverse Event Without Identified Device or Use Problem29
  • Patient Data Problem19
  • Connection Problem11
  • Application Network Problem9
  • Device Sensing Problem8
  • Communication or Transmission Problem7

Patient problems

  • Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available45
  • No Clinical Signs, Symptoms or Conditions31
  • Pneumothorax19
  • Insufficient Information7
  • Hemorrhage/Blood Loss/Bleeding6
  • Cardiac Arrest3
  • No Code Available2
  • Unintended Radiation Exposure2

Recent adverse event reports

The 25 most recent of 104 reports, in the reporter's own words as published by the FDA.

  • MalfunctionSep 15, 2026Patient Data Problem

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50782311 ↗
  • InjuryAug 19, 2026Application Program Problem; Device Sensing Problem

    It was reported that during the procedure, the Patient Sensor Triplet disappeared (PST) however there is no error code indicating as such. A Continues Guidance System (CGS) only restart was performed without resolving the issue, followed by a full system restart with no resolution as well, and all cables were checked with the issue persisting. The patient was under anesthesia at the time, and the physician was unable to complete the superDimension portion of the case, so the case was cancelled as a result.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50329882 ↗
  • MalfunctionAug 14, 2026Patient Data Problem

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50261793 ↗
  • InjuryFeb 26, 2026Application Program Problem; Communication or Transmission Problem

    IT WAS REPORTED THAT DURING A PROCEDURE, THE LOCATION BOARD (LB) WAS DISCONNECTED EVEN THOUGH IT WAS PLUGGED IN, WHICH RESULTED IN THE INABILITY TO COMPLETE THE SUPERDIMENSION PORTION OF THE CASE. THE CASE WAS NOT COMPLETED BY OTHER MEANS AND WAS ABORTED. REPLACEMENT OF THE CONTINUOUS GUIDANCE SYSTEM (CGS), LB, AND PATIENT SENSOR TRIPLET (PST) COMPONENTS WAS SUGGESTED, AND A NEW PIGTAIL HAS RECENTLY BEEN SENT OUT AND REPLACED BY THE SITE.

    Patient problems: Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available

    MAUDE report 24461874 ↗
  • MalfunctionFeb 5, 2026Poor Quality Image; Device Displays Incorrect Message

    IT WAS REPORTED THAT DURING THE MIDDLE OF A CASE, ALL SCREENS WERE MISSING EXCEPT FOR THE LOCAL REGISTRATION SCREEN, AND AN OUT OF SENSING VOLUME ERROR WAS OBSERVED WHILE THE PATIENT WAS UNDER ANESTHESIA. A RESTART OF THE INDUSTRIAL PERSONAL COMPUTER (IPC) WAS SUGGESTED AND PERFORMED, AFTER WHICH THE PHYSICIAN WAS ABLE TO BEGIN RE-REGISTERING. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24272181 ↗
  • InjuryOct 20, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    ACCORDING TO THE REPORTER, POST OPERATIVELY, PNEUMOTHORAX OCCURRED ON THE PATIENT AND WAS DISCOVERED ON ROUTINE POST-PROCEDURAL X-RAY. CHEST TUBE PLACEMENT WAS DONE AS PART OF THE INTERVENTION.

    Patient problems: Pneumothorax

    MAUDE report 23344365 ↗
  • InjurySep 11, 2025Break; Application Program Problem; Communication or Transmission Problem; Break; Application Program Problem; Communication or Transmission Problem

    IT WAS REPORTED THAT DURING THE PROCEDURE, THE PORT THAT HAS THE CABLE THAT CONNECTS TO THE LOCATION BOARD WAS DAMAGED. THE CASE WAS CANCELLED DUE TO INABILITY TO CONNECT THE LOCATION BOARD (LB). THE ELECTROMAGNETIC NAVIGATIONAL BRONCHOSCOPY (ENB) PORTION WAS NOT COMPLETED AND NOT COMPLETED BY OTHER MEANS.

    Patient problems: Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available

    MAUDE report 23025487 ↗
  • InjuryAug 29, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    ACCORDING TO THE LITERATURE SOURCE OF STUDY PERFORMED BETWEEN MARCH 2019 TO FEBRUARY 2024, A RETROSPECTIVE STUDY ANALYZED THE HYBRID TREATMENT OF MULTIFOCAL LUNG MALIGNANCY BY CONCOMITANT TRANSBRONCHIAL MICROWAVE ABLATION WITH SAME-SESSION LUNG RESECTION AND POST-LUNG RESECTION ABLATION WITH CONCOMITANT VIDEO-ASSISTED THORACOSCOPIC LUNG RESECTION. A TOTAL OF 92 PATIENTS WERE INCLUDED IN THE STUDY. THERE WERE 103 PROCEDURES PERFORMED AND 162 LUNG NODULES TREATED 142 LUNG NODULES WERE ABLATED AND 20 LUNG NODULES WERE RESECTED. 6 PATIENTS HAD COMPLICATIONS 1) PERSISTED FEVER, (2) AIRWAY BLEEDING, (2) PNEUMOTHORAX AND (1) BRONCHOPLEURAL FISTULA. ORAL ANTIBIOTICS WERE GIVEN, CHEST TUBE DRAINAGE WAS DONE AND ENDOBRONCHIAL VALVE PLACEMENT ON. RE-ADMISSION WAS REQUIRED. HYBRID TREATMENT OF MULTIFOCAL LUNG MALIGNANCY BY CONCOMITANT TRANSBRONCHIAL MICROWAVE ABLATION WITH SAME-SESSION LUNG RESECTION AND POST-LUNG RESECTION ABLATION.  DR. CALVIN NG, 2025, INTERDISCIPLINARY CARDIOVASCULAR AND THORACIC SURGERY.

    Patient problems: Fever; Fistula; Pneumothorax

    MAUDE report 22926527 ↗
  • MalfunctionAug 25, 2025Unintended Application Program Shut Down; Unintended Application Program Shut Down

    IT WAS REPORTED THAT DURING THE PROCEDURE, THE MONITOR WENT DARK AND THE SYSTEM SHUT OFF, BUT SUBSEQUENTLY CAME BACK ON AND REQUIRED RE-REGISTRATION. THE PATIENT WAS UNDER ANESTHESIA AT THE TIME. THE PHYSICIAN WAS ABLE TO COMPLETE THE ELECTROMAGNETIC NAVIGATIONAL BRONCHOSCOPY (ENB) PORTION OF THE CASE. THERE WAS NO PATIENT INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22891091 ↗
  • MalfunctionJul 25, 2025Unintended Application Program Shut Down; Unintended Application Program Shut Down

    IT WAS REPORTED THAT DURING THE PROCEDURE, THE ILLUMISITE SYSTEM WENT BLACK DURING THE CASE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22616746 ↗
  • MalfunctionJul 21, 2025Failure to Advance; Device Displays Incorrect Message; Failure to Advance; Device Displays Incorrect Message

    IT WAS REPORTED THAT DURING A PROCEDURE, WHEN A PHYSICIAN WENT TO PERFORM AUTOMATIC REGISTRATION, THE FIRST CATHETER WENT BETWEEN THE ERROR MESSAGES OF THE NAVIGATION CATHETER OUT OF SENSING VOLUME AND NOT COUPLED.  THE SURGEON REPLACED THE CABLES AND THE PATIENT'S POSITION WAS CHECKED, BUT THE SAME ISSUE OCCURRED. ANOTHER CATHETER WAS THEN USED. LATER IN THE PROCEDURE, AFTER NAVIGATION AND BIOPSIES, THE CATHETER WAS NOT COUPLING, EVENTUALLY, AN ERROR OF REPLACED LOCATABLE GUIDE (LG) APPEARED. ADDITIONALLY, THE CATHETER WAS BROKEN. THE CATHETER WAS REPLACED AGAIN, RESOLVING THE ISSUE. THE PHYSICIAN WAS ABLE TO COMPLETE THE SUPERDIMENSION PORTION OF THE CASE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22561557 ↗
  • MalfunctionJul 3, 2025Computer Software Problem; Erratic or Intermittent Display; Device Displays Incorrect Message; Computer Software Problem; Erratic or Intermittent Display; Device Displays Incorrect Message

    IT WAS REPORTED THAT DURING A PROCEDURE, OUTSIDE OF MAGNETIC VOLUME ERRORS WERE OBSERVED DURING REGISTRATION, ALTHOUGH REGISTRATION APPEARED NORMAL AND THE PROCEDURE WAS ABLE TO PROCEED MOSTLY AS USUAL. DURING BRUSH PASSES, THE SCREEN COMPLETELY FROZE. THE PHYSICIAN WAS ABLE TO COMPLETE THE ELECTROMAGNETIC NAVIGATIONAL BRONCHOSCOPY (ENB) PORTION OF THE CASE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22401124 ↗
  • InjuryJun 17, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT FOLLOWING A PROCEDURE USING THE CONSOLE, A PATIENT WITH A MEDICAL HISTORY OF A LUNG NODULE EXPERIENCED A PNEUMOTHORAX, WHICH WAS DISCOVERED ON A ROUTINE POST-OPERATIVE X-RAY. THE PNEUMOTHORAX RESULTED IN REQUIRING A CHEST TUBE PLACEMENT AS TREATMENT. THE PATIENT¿S HOSPITALIZATION WAS EXTENDED.

    Patient problems: Pneumothorax; Unintended Radiation Exposure

    MAUDE report 22248066 ↗
  • InjuryApr 29, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT DURING THE PROCEDURE, THE SURGEON WENT AFTER THE RIGHT UPPER LOBE (RUL) IN THE VERY PROXIMAL POSTERIOR SEGMENT. THE USER DID THE NAVIGATION, THE NEEDLES PASSED AND THE FORCEPS BITE WITH NO ISSUES. A BIOPSY SYSTEM FROM ANOTHER MANUFACTURER WAS USED AT THE END OF THE CASE. THE SURGEON AND FELLOW DID THE PASSES USING THAT SYSTEM. THE SURGEON GRABBED PART OF THE PULMONARY ARTERY, AND A MASSIVE AIRWAY BLEED WAS DISCOVERED WHEN PERFORMING A BRONCHOALVEOLAR LAVAGE (BAL). THE PHYSICIAN WAS ABLE TO COMPLETE THE ELECTROMAGNETIC NAVIGATIONAL BRONCHOSCOPY (ENB) PORTION OF THE CASE. THE PATIENT WAS STABILIZED AND TRANSFERRED TO THE INTENSIVE CARE UNIT (ICU). THE EVENT LED TO AN EXTENSION OF THE PATIENT'S HOSPITALIZATION AND RESULTED IN A TEMPORARY INJURY.

    Patient problems: Hemorrhage/Blood Loss/Bleeding

    MAUDE report 21926270 ↗
  • InjuryApr 3, 2025Application Program Problem; Patient Data Problem; Application Program Problem; Patient Data Problem

    IT WAS REPORTED THAT DURING A PROCEDURE, THE AUTOMATIC REGISTRATION WOULD NOT BE COMPLETED. THE PHYSICIAN ATTEMPTED TO COMPLETE THE AUTOMATIC REGISTRATION WITHOUT SUCCESS, TRYING DIFFERENT CATHETERS AND CABLES, REBOOTING THE SYSTEM, AND REPLANNING THE PROCEDURE, BUT NONE OF THESE TROUBLESHOOTING STEPS WERE SUCCESSFUL. GENERAL ANESTHESIA WAS USED, AND THE PROCEDURE WAS ABORTED, BEING UNABLE TO DIAGNOSE THE PATIENT. IT WAS NOTED THAT UPON REPLACEMENT OF ALL CABLES, THE SYSTEM WAS FUNCTIONING PROPERLY.

    Patient problems: Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available

    MAUDE report 21767493 ↗
  • InjuryApr 3, 2025Device Displays Incorrect Message; Application Program Problem; Device Displays Incorrect Message; Application Program Problem

    IT WAS REPORTED THAT DURING THE PROCEDURE, OUT OF SENSING VOLUME ERROR OCCURRED DURING REGISTRATION, LEADING TO THE REGISTRATION POINTS BEING MISALIGNED TO THE RIGHT OF THE 3D TREE. AS A RESULT, THE PHYSICIAN WAS UNABLE TO COMPLETE THE ELECTROMAGNETIC NAVIGATIONAL BRONCHOSCOPY (ENB) PORTION OF THE CASE, AND THE PROCEDURE WAS ABORTED WHILE THE PATIENT WAS UNDER ANESTHESIA. IT WAS MENTIONED THAT THE PROCEDURE WAS RESCHEDULE.

    Patient problems: Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available

    MAUDE report 21766840 ↗
  • InjuryMar 26, 2025Application Program Problem; Adverse Event Without Identified Device or Use Problem; Patient Data Problem; Application Program Problem; Adverse Event Without Identified Device or Use Problem; Patient Data Problem

    IT WAS REPORTED THAT DURING A PROCEDURE, A MAJOR AIRWAY BLEED OF UNSPECIFIED AMOUNT OCCURRED DURING FINAL AUTOMATIC REGISTRATION. A STENT WAS USED TO RESOLVE THE ISSUE. THE PHYSICIAN WAS UNABLE TO COMPLETE THE SUPER DIMENSION PORTION OF THE CASE, AND THE ENTIRE NAVIGATION HAD TO BE ABORTED TO STABILIZE THE PATIENT. THE PHYSICIAN WAS USING A NAVIGATION CATHETER WHEN THE BLEED BEGAN. THE PROCEDURE WAS EXTENDED FOR MORE THAN 30 MINUTES AND WAS COMPLETED BY OTHER MEANS. THE PATIENT HAD TO STAY OVERNIGHT AND WAS ABLE TO LEAVE THE NEXT DAY.

    Patient problems: Bronchial Hemorrhage

    MAUDE report 21703999 ↗
  • MalfunctionMar 11, 2025Use of Device Problem; Application Network Problem; Activation, Positioning or Separation Problem; Unintended Application Program Shut Down; Use of Device Problem; Application Network Problem; Activation, Positioning or Separation Problem; Unintended Application Program Shut Down

    IT WAS REPORTED THAT PREOPERATIVELY, THERE WAS AN ISSUE WITHIN STARTUP OF THE UNIT WHEN IT SHUT DOWN UNEXPECTEDLY AND HAD TO BE REBOOTED. A SYSTEM CHECK FOR THE ILLUMISITE WAS REQUESTED DUE TO CONCERNS FROM END USERS. THERE WAS NO PATIENT INVOLVED. UPON SERVICING, THREE ISSUES WERE IDENTIFIED. FIRST ISSUE WAS A USER INTERFACE SCREEN POP-UP WINDOW DIFFERENCE THE ACCOUNT WERE NOT FAMILIAR WITH. ONE OF THE USER VOLUNTEERED TO BE THE PATIENT ON THE TABLE AND THE OTHER ONE AS THE CLINICIAN, WHILE THE FIELD SERVICE ENGINEER (FSE) OBSERVED, UPON ATTACHING LEADS IT WAS DISCOVERED THAT PLACEMENT AND THE ADHERING OF THE TAPE ON CLOTHES MADE THE TAPE NOT STICK WELL. UPON THE LEAD DETACHING INTERMITTENTLY, THE POP-UP WINDOW DIFFERENCE APPEARED. AFTER TROUBLESHOOTING AND DISCUSSING THE PREVIOUS PATIENT, IT WAS SUSPECTED THE ELASTICITY OF SAID PATIENT COULD HAVE CAUSED THIS ISSUE TO OCCUR. THE REMAINING TWO ISSUES WAS THE TRACKBALL AND THE LOCATER BOARD CABLE ASSEMBLY ALIGNMENT. THE FSE WAS ABLE TO CLEAN THE OPTICS WITHIN THE TRACKBALL ASSEMBLY WHICH WAS FULLY COVERED BY RESIDUE. ONCE CLEANED THROUGHOUT TRACKBALL WAS OPERATIONAL. THE LOCATER BOARD CABLING NEEDED TO BE RETIGHTENED AND REALIGNED

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21578280 ↗
  • MalfunctionFeb 26, 2025Device Displays Incorrect Message; Application Program Problem; Device Sensing Problem; Device Displays Incorrect Message; Application Program Problem; Device Sensing Problem

    IT WAS REPORTED THAT DURING A PROCEDURE INVOLVING THE SUPERDIMENSION SYSTEM, AN INVALID NAVIGATION CATHETER MOVEMENT ERROR OCCURRED. INITIALLY, THE PATIENT SENSOR TRIPLET (PST) WAS NOT VISIBLE IN THE SOFTWARE, PROMPTING THE CABLE TO BE UNPLUGGED AND RECONNECTED. THE PST WAS THEN SEEN, BUT REMAINED OUT OF THE SENSING VOLUME, NECESSITATING MULTIPLE ADJUSTMENTS TO THE PATIENT'S POSITION. THE PHYSICIAN WAS UNABLE TO COMPLETE THE SUPERDIMENSION PORTION OF THE CASE. THE PROCEDURE WAS ULTIMATELY COMPLETED USING ENDOBRONCHIAL ULTRASOUND (EBUS) WHILE THE PATIENT WAS UNDER ANESTHESIA. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT. UPON SERVICING, PERFORMED ACCURACY CHECKOUT PROCEDURE AND FLOURO CHECKOUT PROCEDURE. ACCURACY CHECKOUT AND FLOURO PROCEDURE PASSED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21470899 ↗
  • InjuryJan 24, 2025Computer Software Problem; Display or Visual Feedback Problem; Application Program Problem; Computer Software Problem; Display or Visual Feedback Problem; Application Program Problem

    ACCORDING TO THE REPORTER, DURING AN ELECTROMAGNETIC NAVIGATION BRONCHOSCOPY (ENB), THE BRONCH VIDEO WAS NOT SENSED AND THEN THE APPLICATION WON'T START. THE PATIENT WAS UNDER SEDATION AND THE CASE WAS CANCELLED.

    Patient problems: Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available; No Clinical Signs, Symptoms or Conditions

    MAUDE report 21228911 ↗
  • MalfunctionJan 22, 2025Device Displays Incorrect Message; Device Displays Incorrect Message

    IT WAS REPORTED THAT DURING A PROCEDURE, THE DEVICE HAD AN ERROR MESSAGE THAT THE NAVIGATION CATHETER WAS OUT OF SENSING VOLUME. THE PATIENT WAS UNDER ANESTHESIA, AND THE PHYSICIAN WAS ABLE TO COMPLETE THE SUPERDIMENSION PORTION OF THE CASE. THERE WERE NO PATIENT COMPLICATIONS OR HARMS AS A RESULT OF THE ISSUE. A SITE VISIT WAS REQUESTED FOR FURTHER ASSESSMENT. UPON SERVICING, AN ACCURACY CHECK WAS REQUESTED, AND IT WAS COMPLETED AND PASSED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21211105 ↗
  • InjuryJan 9, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT DURING A PROCEDURE, A BIOPSY OF A VASCULAR NODULE IN THE UPPER LEFT LOBE WAS DONE, EXCESSIVE BLEEDING OCCURRED AFTER, LEADING TO A MYOCARDIAL INFARCTION. THE PATIENT WAS SUBSEQUENTLY BROUGHT TO THE INTENSIVE CARE UNIT (ICU) FOR RECOVERY AND OBSERVATION, EXTENDING HOSPITALIZATION BY ONE DAY.

    Patient problems: Hemorrhage/Blood Loss/Bleeding; Myocardial Infarction

    MAUDE report 21112536 ↗
  • MalfunctionDec 27, 2024Device Displays Incorrect Message; Device Displays Incorrect Message

    ACCORDING TO THE REPORTER, DURING A ELECTROMAGNETIC NAVIGATION BRONCHOSCOPY (ENB), THE LOCATABLE GUIDE WAS COMING OUT OF SENSING VOLUME WHILE THE EXTENDED WORKING CHANNEL WAS STAYING IN. THE LESION WAS IN THE ANTERIOR MID/LUNG. IT WAS SUGGESTED TO RAISE THE PATIENT UP HOWEVER THEY WERE UNABLE TO COMPLETE THIS STEP. RE-REGISTERING AND MOVING THE PATIENT SENSOR TRIPLET DID NOT RESOLVE THE ISSUE. THE SUPER DIMENSION PORTION OF THE CASE WAS COMPLETED. THERE WAS NO PATIENT INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21028157 ↗
  • InjuryNov 27, 2024Unintended Application Program Shut Down; Unintended Application Program Shut Down

    ACCORDING TO THE REPORTER, DURING A ELECTROMAGNETIC NAVIGATION BRONCHOSCOPY (ENB) IN THE MIDDLE OF THE PROCEDURE, THE INTRAPERSONAL COMPUTER TURNED OFF. A FULL SYSTEM REBOOT WAS PERFORMED WHICH DID NOT RESOLVE THE ISSUE. THEY THEN TOOK POWER FROM THE UNINTERRUPTABLE POWER SUPPLY FROM THE WALL OUTLET AND ATTEMPTED TO MANUALLY TURN ON THE DEVICE TO NO AVAIL. THE PHYSICIAN WAS ABLE TO COMPLETE THE SUPERDIMENSION PORTION OF THE CASE BUT THE CASE WAS ABORTED. IT WAS NOTED THAT THE PATIENT WAS UNDER ANESTHESIA.

    Patient problems: Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available

    MAUDE report 20801267 ↗
  • InjuryNov 11, 2024Device Displays Incorrect Message; Application Network Problem; Application Program Problem; Device Displays Incorrect Message; Application Network Problem; Application Program Problem

    ACCORDING TO THE REPORTER, DURING AN ELECTROMAGNETIC NAVIGATION BRONCHOSCOPY (ENB) PROCEDURE, THE SOFTWARE WAS SHOWING THAT THEY WERE ON THE LOWER LOBE, WHILE FLUOROSCOPIC IMAGING SHOWED THEM IN THE UPPER LOBE. IT WAS ALSO REPORTED THAT THE DEVICE WAS SHOWING AS O UTSIDE OF THE SENSOR SENSING VOLUME.

    Patient problems: Insufficient Information

    MAUDE report 20651947 ↗

FDA clearance and classification

Cleared through 510(k) K191394 on Aug 7, 2019 (decision: Substantially Equivalent). Submitted by Covidien, LLC as “ILLUMISITE Platform”. The FDA review took 75 days. FDA record ↗

The FDA regulates product code JAK (system, x-ray, tomography, computed) under 21 CFR 892.1750 as a Class II device. 21 CFR 892.1750 ↗

Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.

Recalls under the same FDA submission

These recalls involve devices cleared under K191394. One submission often covers many models, so check the product and lot details in each record.

  • Z-2008-2023 · May 22, 2023 · Open, Classified
    Medtronic ILLUMISITE Platform Console, Model Number ILS-1000-CS

Common questions about the ILLUMISITE ILS-1000-CS

Has the ILLUMISITE ILS-1000-CS been recalled?

No FDA recall record in our data names this exact device identifier (DI 10884521723634), based on the FDA recall file dated Nov 27, 2026. However, 1 recall involves devices cleared under the same FDA submission. It may or may not include this model, so check the product, model and lot details in that record.

What problems have been reported with the ILLUMISITE ILS-1000-CS?

The FDA's MAUDE database lists 104 adverse event reports for this device, received between Jan 30, 2020 and Sep 15, 2026. In those reports, the most common specific device problems are “Application Program Problem” (51), “Device Displays Incorrect Message” (37) and “Patient Data Problem” (19); the most common patient problems are “Pneumothorax” (19), “Hemorrhage/Blood Loss/Bleeding” (6) and “Cardiac Arrest” (3). Reports are not verified by the FDA and do not prove the device caused the problem.

Have deaths or injuries been reported with the ILLUMISITE ILS-1000-CS?

1 report is classified as a death, 76 as injuries and 27 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.

Is the ILLUMISITE ILS-1000-CS FDA approved?

It was cleared through 510(k) K191394 on Aug 7, 2019. Strictly, the FDA "clears" 510(k) devices as substantially equivalent to devices already on the market rather than "approving" them; approval is the PMA route.

Who makes the ILLUMISITE ILS-1000-CS, and is it still sold?

The FDA's device database lists Covidien LP as the company responsible for it (brand ILLUMISITE, model ILS-1000-CS). Its record shows it as in commercial distribution.

Answers are built automatically from the FDA records on this page. They are not medical advice.