Sonicision 7 SCD7A13Recalls & reported problems

Covidien LP · Curved Jaw Cordless Ultrasonic Dissector · Class U

All models of Sonicision 7 (5)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
5
Deaths reported
0
Injuries reported
2
Recalls naming it
0

Summary

The FDA's MAUDE database contains 5 adverse event reports that list this exact device (DI 10884521794771), received between Feb 8, 2023 and Jun 9, 2025 — classified as 2 injuries, 3 malfunctions.

The most frequently reported specific device problem is “Detachment of Device or Device Component” (3 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 1 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 1 reportsJun 2025: 1 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Detachment of Device or Device Component3
  • Adverse Event Without Identified Device or Use Problem2
  • Failure to Deliver Energy2
  • Failure to Align1
  • Intermittent Energy Output1
  • Naturally Worn1
  • Overheating of Device1
  • Peeled/Delaminated1

Patient problems

  • No Clinical Signs, Symptoms or Conditions2
  • Burn(s)1
  • Hemorrhage/Blood Loss/Bleeding1
  • Nerve Damage1

Recent adverse event reports

The 5 most recent of 5 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJun 9, 2025Failure to Deliver Energy; Failure to Align; Detachment of Device or Device Component; Failure to Deliver Energy; Failure to Align; Detachment of Device or Device Component

    IT WAS REPORTED THAT DURING AN OVARIECTOMY AT A VETERINARY CLINIC, THE DISTAL PART OF THE DISSECTOR'S JAW DID NOT CLOSE COMPLETELY, LEAVING A SMALL GAP OF APPROXIMATELY 1 MILLIMETER BETWEEN THE JAWS. ADDITIONALLY, THE DEVICE DID NOT COAGULATE EFFECTIVELY IN THE MIDDLE OF THE JAW. NO PATIENT COMPLICATIONS WERE REPORTED AS A RESULT OF THIS EVENT. MEDTRONIC'S INITIAL EVALUATION OF THE INCIDENT DEVICE FOUND THAT THE JAW LINER WAS MELTED AND PARTIALLY DISLODGED. THERE WERE ALSO MISSING PIECES.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22169252 ↗
  • MalfunctionMay 15, 2025Peeled/Delaminated; Detachment of Device or Device Component; Naturally Worn; Peeled/Delaminated; Detachment of Device or Device Component; Naturally Worn

    ACCORDING TO THE REPORTER, DURING A PYOMETRA PROCEDURE ON A VETERINARY CASE, THE SEAL IN THE JAW WAS BURNED AND THE COATING ON THE DISSECTOR BEGAN TO PEEL OFF. THERE WAS NO PIECE FELL INTO THE CAVITY. DESPITE THIS, THE BLOOD VESSELS AND TISSUE WERE SUCCESSFULLY SEALED AND COAGULATED. HOWEVER, THE VETERINARIAN LOST CONFIDENCE IN THE DEVICE AND IMMEDIATELY SWITCHED TO USING SUTURE. AT THAT POINT, BOTH OVARIES AND ONE SIDE OF THE CERVIX HAD ALREADY BEEN SEALED, WITH NO ACTIVE BLEEDING OBSERVED. THE FINAL VESSEL ON THE CERVIX WAS TIED OFF WITH SUTURE. AFTER CLOSING THE ANIMAL AND TRANSFERRING IT TO RECOVERY, HOSPITALIZATION WAS EXTENDED FOR ONE DAY AND A SUBSEQUENT CHECK REVEALED THAT THE ANIMAL HAD PASSED AWAY. PRELIMINARY FINDINGS OF THE AUTOPSY INDICATED INTERNAL BLEEDING AS THE CAUSE OF DEATH, THOUGH THE EXACT SOURCE REMAINS UNCLEAR. A PHOTO WAS SUBMITTED SHOWING THE JAW LINER WAS MELTED.

    Patient problems: Hemorrhage/Blood Loss/Bleeding

    MAUDE report 22026730 ↗
  • MalfunctionJan 28, 2025Detachment of Device or Device Component; Intermittent Energy Output; Detachment of Device or Device Component; Intermittent Energy Output

    ACCORDING TO THE REPORTER PRIOR TO USE, THE DEVICE DID NOT ACTIVATE. USED ANOTHER DISSECTOR AND IT WORKED FINE. THERE WAS NO PATIENT INVOLVEMENT. MEDTRONIC'S INITIAL EVALUATION OF THE INCIDENT DEVICE FOUND THAT THE DISSECTOR HAD A FRACTURED WAVEGUIDE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21252173 ↗
  • InjuryJun 30, 2023Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    ACCORDING TO THE REPORTER, DURING A THYROIDECTOMY, THE DEVICE GOT HOT AND BURNED THE PATIENT'S CAROTID ARTERY AND NEEDED TO BE TRANSFERRED TO ANOTHER FACILITY FOR OBSERVATION. THE PATIENT WAS IN A CRITICAL CARE/INTENSIVE CARE UNIT (ICU) SETTING.

    Patient problems: Burn(s)

    MAUDE report 17238126 ↗
  • InjuryFeb 8, 2023Failure to Deliver Energy; Overheating of Device; Adverse Event Without Identified Device or Use Problem; Failure to Deliver Energy; Overheating of Device; Adverse Event Without Identified Device or Use Problem

    ACCORDING TO THE REPORTER, DURING THYROIDECTOMY, THE DEVICE TAKES 30-40% LONGER TO OCCLUDE VESSELS IN MIN MODE AND RESULTED IN NERVE DAMAGE BECAUSE THE DEVICE GETS HOTTER.

    Patient problems: Nerve Damage

    MAUDE report 16326976 ↗

Other Instrument, ultrasonic surgical devices