The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 10884521807860), received between Aug 24, 2023 and Aug 24, 2023 — classified as 1 malfunction.

The most frequently reported specific device problem is “Detachment of Device or Device Component” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Most reported problems

Device problems

  • Detachment of Device or Device Component1
  • Intermittent Energy Output1

Patient problems

  • No Clinical Signs, Symptoms or Conditions1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • MalfunctionAug 24, 2023Detachment of Device or Device Component; Intermittent Energy Output; Detachment of Device or Device Component; Intermittent Energy Output

    ACCORDING TO THE REPORTER, THE DEVICE DID NOT ACTIVATE. THERE WAS NO PATIENT INVOLVEMENT. MEDTRONIC'S INITIAL EVALUATION OF THE INCIDENT DEVICE FOUND THAT THE DISPOSABLE DEVICE REVEALED THAT THE TIP OF THE WAVEGUIDE HAD FRACTURED AND DISENGAGED.

    Patient problems: No Clinical Signs, Symptoms or Conditions