Chameleon CH10-40-75USRecalls & reported problems

Mozarc Medical US LLC · PTA Balloon Catheter with Injection Port · Class 2

All models of Chameleon (7)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
3
Deaths reported
0
Injuries reported
1
Recalls naming it
1

Summary

The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 10884521812314), received between Jan 12, 2024 and May 13, 2024 — classified as 1 injury, 2 malfunctions.

The most frequently reported specific device problem is “Detachment of Device or Device Component” (3 reports).

1 FDA recall record name this device's identifier, 1 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 1 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 2 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520262
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Sep 19, 2024Open, Classified

    Specific lots of the Chameleon PTA Balloon Catheter failed to meet in-house balloon pressure testing.

    Z-0257-2025 · Root cause: Under Investigation by firm · Recall details · FDA record ↗

Most reported problems

Device problems

  • Detachment of Device or Device Component3
  • Difficult to Remove1

Patient problems

  • No Clinical Signs, Symptoms or Conditions2
  • Arrhythmia1
  • Discomfort1
  • Foreign Body In Patient1
  • Insufficient Information1
  • Pain1
  • Unintended Radiation Exposure1

Recent adverse event reports

The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.

  • InjuryMay 13, 2024Detachment of Device or Device Component; Detachment of Device or Device Component

    ACCORDING TO THE REPORTER, THE PATIENT CAME IN NEEDING A CATHETER EXCHANGE IN THE RIJ (RIGHT INTERNAL JUGULAR VEIN). STENOSIS OF THE VENOTOMY SITE AND THE RIJ TUNNEL WAS SEEN. THE PREVIOUS CATHETER WAS REMOVED, AND A GUIDEWIRE WAS PLACED INTO THE RIJ (RIGHT INTERNAL JUGULAR). A 10X4 PTA (PERCUTANEOUS TRANSLUMINAL ANGIOPLASTY) BALLOON WAS THEN PLACED OVER THE WIRE AND INTO THE RIJ, WHERE ANGIOPLASTY WAS DONE AT BOTH THE VENOTOMY SITE AND DOWN THE RIJ. AFTER ANGIOPLASTY, THE PTA BALLOON WAS REMOVED, AND A 23-CM (CENTIMETER) COMPETITOR'S HEMODIALYSIS CATHETER WAS ATTEMPTED TO BE INSERTED OVER THE WIRE AND INTO THE PATIENT. THE DIALYSIS CATHETER GOT PAST THE VENOTOMY SITE BUT MET RESISTANCE (STENOSIS) AS IT CONTINUED DOWN RIJ. AT THAT TIME, THE DECISION WAS MADE TO DO MORE ANGIOPLASTY, SO THE DIALYSIS CATHETER WAS REMOVED, AND THE PTA BALLOON WAS PLACED BACK OVER THE WIRE AND DOWN THE RIJ, WHERE ADDITIONAL ANGIOPLASTY OF THE SVA (SUPERFICIAL VENOUS ARTERIALIZATION) AND RIJ WAS PERFORMED USING A 10 MM (MILLIMETER) BALLOON. UPON THE FINAL ANGIOPLASTY, THE BALLOON WAS STILL PARTIALLY INFLATED (THE BALLOON WAS PARTIALLY INFLATED AND PULLED BACK TOWARD THE VENOTOMY SITE INTENTIONALLY BY THE

    Patient problems: Arrhythmia; Discomfort; Foreign Body In Patient; Pain; Unintended Radiation Exposure

    MAUDE report 19305184 ↗
  • MalfunctionMay 7, 2024Detachment of Device or Device Component; Detachment of Device or Device Component

    ACCORDING TO THE REPORTER, DURING PROCEDURE, THE NURSE WENT TO OPEN THE BALLOON AND IT WAS SNAPPED IN HALF. THE BALLOON NEVER TOUCHED THE PATIENT. THE CATHETER WAS NOT REPAIRED. THERE WAS NO LEAK. TEGO WAS NOT UTILIZED. THERE WAS NO LUER ADAPTER ISSUE. NOTHING ELSE UNUSUAL OBSERVED ON THE DEVICE. NO OTHER DEFECTS/DAMAGES FOUND ON THE PRODUCT. THERE WAS NO DAMAGE TO THE DEVICE'S PACKAGING AND TO THE DEVICE'S BOX. THE PRODUCT WAS REPLACED WITH DIFFERENT LOT THREE DAYS AFTER THE INCIDENT AND THIS RESOLVED THE ISSUE BUT THEY STILL HAVE SOME OF THESE LOTS NUMBERS ON THEIR SHELF. THERE WERE NO CONFIRMED ISSUES OBSERVED WITH THESE REMAINING LOT NUMBERS ON THEIR SHELF. THERE WAS NO REPORTED PATIENT INJURY.

    Patient problems: Insufficient Information; No Clinical Signs, Symptoms or Conditions

    MAUDE report 19264486 ↗
  • MalfunctionJan 12, 2024Difficult to Remove; Detachment of Device or Device Component; Difficult to Remove; Detachment of Device or Device Component

    ACCORDING TO THE REPORTER, THE PATIENT WAS PRESENTED WITH A NARROWING IN UPPER PORTION OF BRACHIAL VEIN MEDIAL TO A COVERED STENT. THE DOCTOR GAINED INITIAL ACCESS VIA A MICRO PUNCTURE SET, WHERE THE DOCTOR THEN PLACED AN ANGIOCATH FOR INITIAL IMAGING. PROPER FLUSHING PERFORMED OF SHEATH, GUIDEWIRE, AND CATHETER PRIOR TO CASE WITH NORMAL RESULT. ONCE THE LESION WAS IMAGED, A 7FR (FRENCH) COMPETITOR¿S SHEATH WAS INTRODUCED OVER THE WIRE FOLLOWED BY A 10X40 CHAMELEON BALLOON. THE BALLOON WAS ADVANCED TO THE AREA OF STENOSIS AND INFLATED TO RBP (RATED BURST PRESSURE) OF 14. BALLOON WAS THEN RETRACTED TO THE AREA AT THE MEDIAL ENDPOINT OF THE STENT AND ANGIOPLASTY WAS PERFORMED IN A SECOND AREA USING INFLATION DEVICE TO THE RBP. BALLOON WAS DEFLATED AFTER 60 SECONDS, AND IMAGING WAS PERFORMED TO DETERMINE THAT THERE WAS NO RESIDUAL STENOSIS POST ANGIOPLASTY. THE DOCTOR WENT TO REMOVE THE BALLOON AND AS THE DOCTOR PULLED THROUGH THE SHEATH, THE DOCTOR RECEIVED RESISTANCE. THE DOCTOR WAS ABLE TO PULL AGAIN BUT THE RESISTANCE INCREASED AND AT THIS POINT, THE DOCTOR KNEW THAT AT LEAST A PORTION OF THE BALLOON WAS INSIDE THE SHEATH. THE DOCTOR PULLED AGAIN, AND THE BALLOON SEPARATED FROM TH

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18503897 ↗

Other Catheter, angioplasty, peripheral, transluminal devices

All Catheter, angioplasty, peripheral, transluminal devices →