Skip to content

Endo GIARecalls & reported problems across all models

Covidien LP · 48 models in the FDA device registry

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
26,277
Deaths reported
71
Injuries reported
3,377
Recalls
8

We found 48 models of Endo GIA by Covidien LP in the FDA's device registry (GUDID). Together they are named in 26,277 adverse event reports in the FDA's MAUDE database, received between Jan 26, 2016 and Sep 30, 2026 — classified as 71 deaths, 3,377 injuries, 22,828 malfunctions.

The most frequently reported specific device problem is “Failure to Fire” (12,118 reports).

8 FDA recall records name at least one of these models by its device identifier.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event. Widely used products collect more reports, so totals can't be used to compare products or estimate failure rates. Only reports that name a model's exact device identifier are counted. What reports can and can't tell you · How this data works

Reports received per month, all models

Nov 2023: 212 reportsDec 2023: 261 reportsJan 2024: 315 reportsFeb 2024: 260 reportsMar 2024: 225 reportsApr 2024: 293 reportsMay 2024: 217 reportsJun 2024: 271 reportsJul 2024: 223 reportsAug 2024: 226 reportsSep 2024: 258 reportsOct 2024: 219 reportsNov 2024: 227 reportsDec 2024: 96 reportsJan 2025: 239 reportsFeb 2025: 117 reportsMar 2025: 159 reportsApr 2025: 117 reportsMay 2025: 146 reportsJun 2025: 154 reportsJul 2025: 261 reportsAug 2025: 151 reportsSep 2025: 135 reportsOct 2025: 121 reportsNov 2025: 136 reportsDec 2025: 182 reportsJan 2026: 113 reportsFeb 2026: 143 reportsMar 2026: 174 reportsApr 2026: 85 reportsMay 2026: 34 reportsJun 2026: 41 reportsJul 2026: 18 reportsAug 2026: 68 reportsSep 2026: 60 reportsOct 2026: 0 reports202420252026315
Last 36 months. The most recent month may be incomplete.

Endo GIA recalls

FDA recall records that name at least one of these models by its device identifier, newest first. Each recall usually covers specific models or lots, so check the details.

  1. Apr 26, 2018Terminated

    Device may be missing a sled component. The sled component is responsible for staple deployment. This may result in failure to form a staple line when tissue is divided, leading to bleeding or leakage of luminal contents.

    Z-1822-2018 · Root cause: Process control · Recall details · FDA record ↗

  2. Apr 26, 2018Terminated

    Device may be missing a sled component. The sled component is responsible for staple deployment. This may result in failure to form a staple line when tissue is divided, leading to bleeding or leakage of luminal contents.

    Z-1820-2018 · Root cause: Process control · Recall details · FDA record ↗

  3. Apr 26, 2018Terminated

    Device may be missing a sled component. The sled component is responsible for staple deployment. This may result in failure to form a staple line when tissue is divided, leading to bleeding or leakage of luminal contents.

    Z-1819-2018 · Root cause: Process control · Recall details · FDA record ↗

  4. Apr 26, 2018Terminated

    Device may be missing a sled component. The sled component is responsible for staple deployment. This may result in failure to form a staple line when tissue is divided, leading to bleeding or leakage of luminal contents.

    Z-1818-2018 · Root cause: Process control · Recall details · FDA record ↗

  5. Apr 26, 2018Terminated

    Device may be missing a sled component. The sled component is responsible for staple deployment. This may result in failure to form a staple line when tissue is divided, leading to bleeding or leakage of luminal contents.

    Z-1817-2018 · Root cause: Process control · Recall details · FDA record ↗

  6. Apr 26, 2018Terminated

    Device may be missing a sled component. The sled component is responsible for staple deployment. This may result in failure to form a staple line when tissue is divided, leading to bleeding or leakage of luminal contents.

    Z-1816-2018 · Root cause: Process control · Recall details · FDA record ↗

  7. Apr 26, 2018Terminated

    Device may be missing a sled component. The sled component is responsible for staple deployment. This may result in failure to form a staple line when tissue is divided, leading to bleeding or leakage of luminal contents.

    Z-1815-2018 · Root cause: Process control · Recall details · FDA record ↗

  8. Apr 26, 2018Terminated

    Device may be missing a sled component. The sled component is responsible for staple deployment. This may result in failure to form a staple line when tissue is divided, leading to bleeding or leakage of luminal contents.

    Z-1814-2018 · Root cause: Process control · Recall details · FDA record ↗

Most reported Endo GIA problems

Device problems

  • Failure to Fire12,118
  • Failure to Form Staple4,783
  • Misfire3,936
  • Difficult to Open or Close2,779
  • Adverse Event Without Identified Device or Use Problem2,243
  • Entrapment of Device1,383
  • Break1,320
  • Human-Device Interface Problem1,044

Patient problems

  • No Clinical Signs, Symptoms or Conditions9,671
  • No Known Impact Or Consequence To Patient5,336
  • No Consequences Or Impact To Patient4,960
  • No Code Available1,697
  • Tissue Breakdown1,614
  • Insufficient Information1,390
  • Unspecified Tissue Injury1,389
  • Hemorrhage/Blood Loss/Bleeding1,107

Endo GIA models

48 models, most reported first. Each model has its own page with its reports, recalls and device details.

Recent Endo GIA adverse event reports

The 15 most recent reports across all models, in the reporter's own words as published by the FDA.

Other Covidien LP products

All Covidien LP devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page