Multifire Endo GIA 030807LRecalls & reported problems
Covidien LP · Straight Loading Unit · Class 2 · Implant
All models of Multifire Endo GIA (5)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 4
- Deaths reported
- 0
- Injuries reported
- 2
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 10884523003772), received between Jun 7, 2017 and Jun 19, 2020 — classified as 2 injuries, 2 malfunctions.
The most frequently reported specific device problem is “Failure to Form Staple” (2 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Most reported problems
Device problems
- Failure to Form Staple2
- Failure to Cut1
- Mechanics Altered1
Patient problems
- No Code Available3
- Tissue Damage2
- No Consequences Or Impact To Patient1
- No Known Impact Or Consequence To Patient1
Recent adverse event reports
The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.
InjuryJun 19, 2020Mechanics Altered; Mechanics Altered
ACCORDING TO THE REPORTER, DURING AN ENDOSCOPIC PHARYNGEAL POUCH STAPLING PROCEDURE, WHEN THE SURGEON TRIED TO FIRE THE DEVICE, IT CUTS THE POUCH BUT NONE OF THE STAPLES WERE DEPLOYED. THE SURGICAL TIME WAS EXTENDED FOR MORE THAN 30 MINUTES AND THERE WAS TISSUE DAMAGE AS A RESULT OF THE DEVICE ISSUE. THE SURGEON TRIED TO FIRE GUN RELOAD (NOT ON THE PATIENT) STAPLES DEPLOYED. THE PATIENT HAD TO UNDERGO AN EMERGENCY OPERATION A WEEK LATER, TO REPAIR THE IATROGENIC ESOPHAGEAL PERFORATION OF THE PATIENT WELL AS THE ASSOCIATED INFECTION, AND THE PATIENT HOSPITALIZATION WAS EXTENDED. THE PATIENT IS STILL CURRENTLY IN THE HOSPITAL, KEPT NBM WITH A NASOGASTRIC FEEDING TUBE, LONG TERM ANTIBIOTIC AND A CORRUGATED DRAIN IN THE NECK TO REMOVE ANY ON-GOING SALIVA & INFECTIVE MATERIAL IN THE NECK/SUPERIOR MEDIASTINUM.
Patient problems: No Code Available; Tissue Damage
MAUDE report 10174043 ↗MalfunctionMar 21, 2019Failure to Cut; Failure to Cut
ACCORDING TO THE REPORTER, DURING A VIDEO-ASSISTED THORACIC SURGERY (VATS) WEDGE RESECTION, THE SECOND UNIT WAS USED AND THE TISSUE WAS CLAMPED BUT NOT COMPLETELY SEVERED BY THE SCALPEL. IT WAS REPORTED THAT THE SCALPEL WAS BENT. ANOTHER STAPLER WAS USED TO COMPLETE THE PROCEDURE. THERE WAS NO PATIENT INJURY.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 8441916 ↗MalfunctionAug 17, 2017Failure to Form Staple; Failure to Form Staple
ACCORDING TO THE REPORTER: DURING LAPAROSCOPIC APPENDECTOMY PROCEDURE, THE DEVICE WAS STAPLED THE LINE INCOMPLETE. TO FINISH THE PROCEDURE, THE SURGEON HAND SEWN THE OPENING WHERE THE APPENDIX WAS TAKEN OFF. NO PATIENT INJURY HAS BEEN NOTED.
Patient problems: No Code Available; No Known Impact Or Consequence To Patient
MAUDE report 6801496 ↗InjuryJun 7, 2017Failure to Form Staple; Failure to Form Staple
ACCORDING TO THE REPORTER, DURING THE LAPAROSCOPIC APPENDECTOMY PROCEDURE, THE STAPLER ONLY RELEASED STAPLES ON ONE SIDE SO INTESTINE WAS OPEN AND THEY NEEDED TO STAPLE ONE MORE TIME. IN ORDER TO COMPLETE THE CASE, THEY TOOK A NEW RELOAD AND STAPLED ONE MORE TIME. IT WAS NOTED THERE WAS UNANTICIPATED TISSUE LOSS MORE THAN AN ORDINARY SURGERY AND TISSUE DAMAGE AS A RESULT OF THE PROBLEM. THERE WAS NO BLOOD LOSS OF 500CC OR MORE AND THE SURGICAL TIME WAS NOT EXTENDED BY 30 MIN OR MORE. THERE WERE NO TEMPORARY OR PERMANENT COMPLICATIONS. THE PATIENT WAS FINE POST-OP.
Patient problems: No Code Available; Tissue Damage
MAUDE report 6618457 ↗
Other STAPLE, IMPLANTABLE devices
- Endo GIA EGIA60AMTCovidien LP · STAPLE, IMPLANTABLE19 deaths10,680 reports
- Endo GIA Ultra EGIAUSTNDCovidien LP · STAPLE, IMPLANTABLE5 deaths9,730 reports
- Signia SIGPHANDLECovidien LP · STAPLE, IMPLANTABLE1 recall3 deaths5,356 reports
- ECHELON FLEX PSEE60AEthicon Endo-Surgery, LLC · STAPLE, IMPLANTABLE1 recall13 deaths3,294 reports
- Endo GIA Ultra EGIAUSHORTCovidien LP · STAPLE, IMPLANTABLE2 deaths3,011 reports
- Endo GIA EGIA45AMTCovidien LP · STAPLE, IMPLANTABLE10 deaths2,965 reports
- Endo GIA Ultra EGIAUXLCovidien LP · STAPLE, IMPLANTABLE2,909 reports
- ENDOPATH ATS45Ethicon Endo-Surgery, LLC · STAPLE, IMPLANTABLE2 deaths2,046 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page