EEA EEA28Recalls & reported problems

Covidien LP · Circular Stapler with DST Series Technology · Class 2

All models of EEA (39)

Data updated Nov 27, 2026 from official FDA records · Editorial policy

Quick summary

This Covidien EEA28 circular stapler is a surgical stapling tool used to reconnect sections of bowel during surgery. The 995 reports often describe staples not forming or firing, the device breaking or detaching, and being hard to remove. Reports include 6 deaths and 457 injuries. No recalls name this device.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
989
Deaths reported
6
Injuries reported
455
Recalls naming it
0

Summary

The FDA's MAUDE database contains 989 adverse event reports that list this exact device (DI 10884523005530), received between Jul 6, 2016 and Aug 13, 2026 — classified as 6 deaths, 455 injuries, 528 malfunctions.

The most frequently reported specific device problem is “Failure to Form Staple” (262 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.

Reports received per month

Dec 2023: 7 reportsJan 2024: 7 reportsFeb 2024: 4 reportsMar 2024: 2 reportsApr 2024: 9 reportsMay 2024: 6 reportsJun 2024: 7 reportsJul 2024: 11 reportsAug 2024: 4 reportsSep 2024: 9 reportsOct 2024: 4 reportsNov 2024: 8 reportsDec 2024: 0 reportsJan 2025: 12 reportsFeb 2025: 1 reportsMar 2025: 6 reportsApr 2025: 5 reportsMay 2025: 9 reportsJun 2025: 7 reportsJul 2025: 6 reportsAug 2025: 7 reportsSep 2025: 9 reportsOct 2025: 4 reportsNov 2025: 12 reportsDec 2025: 9 reportsJan 2026: 3 reportsFeb 2026: 8 reportsMar 2026: 6 reportsApr 2026: 2 reportsMay 2026: 1 reportsJun 2026: 1 reportsJul 2026: 2 reportsAug 2026: 3 reportsSep 2026: 0 reportsOct 2026: 0 reportsNov 2026: 0 reports20242025202612
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Failure to Form Staple262
  • Adverse Event Without Identified Device or Use Problem196
  • Difficult to Remove155
  • Detachment of Device or Device Component130
  • Failure to Fire111
  • Misfire72
  • Loss of or Failure to Bond66
  • Break58

Patient problems

  • Failure to Anastomose283
  • No Code Available230
  • No Clinical Signs, Symptoms or Conditions201
  • Unspecified Tissue Injury144
  • No Known Impact Or Consequence To Patient105
  • Tissue Damage99
  • No Consequences Or Impact To Patient83
  • Tissue Breakdown61

Recent adverse event reports

The 25 most recent of 989 reports, in the reporter's own words as published by the FDA.

  • MalfunctionAug 13, 2026Failure to Form Staple

    Patient problems: Failure to Anastomose

    MAUDE report 50236580 ↗
  • MalfunctionAug 12, 2026Detachment of Device or Device Component

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50220833 ↗
  • MalfunctionAug 6, 2026Device Slipped

    It was reported that during an open colectomy, while being used for colon anastomosis, the circular stapler was unable to close. Unusual effort was reportedly required to turn the twist knob in order to visualize the green bar. The stapler was replaced with another circular stapler to complete the case. No patient complications have been reported as a result of this event.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50160982 ↗
  • MalfunctionJul 14, 2026Detachment of Device or Device Component; Material Twisted/Bent

    ACCORDING TO THE REPORTER, DURING A PARTIAL COLECTOMY, WHEN REMOVING THE DEVICE AFTER FIRING, THE ANVIL BECAME DISCONNECTED FROM THE HANDLE, AND THE SURGEON HAD TO REMOVE IT MANUALLY BY USING A CLAMP. THE ANVIL WAS BENT. THERE WAS NO PATIENT INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25809457 ↗
  • MalfunctionJul 9, 2026Failure to Form Staple

    ACCORDING TO THE REPORTER, DURING A LAPAROSCOPIC RADICAL RESECTION FOR RECTAL CANCER, THE INNER RING OF STAPLES DID NOT FULLY CLOSE DURING DIGESTIVE TRACT RECONSTRUCTION AND COLORECTAL ANASTOMOSIS, RESULTING IN POOR STAPLE FORMATION. MANUAL REINFORCEMENT SUTURING, RESECTION, AND RE-STAPLING WERE PERFORMED TO COMPLETE THE PROCEDURE.

    Patient problems: Tissue Breakdown

    MAUDE report 25769206 ↗
  • InjuryJun 11, 2026Detachment of Device or Device Component

    ACCORDING TO THE REPORTER, DURING A LAPAROSCOPIC VIDEO-ASSISTED RECTOSIGMOIDECTOMY, DURING ANORECTAL ANASTOMOSIS, THE ANVIL DETACHED PRIOR TO STAPLING. THE SURGICAL PROCEDURE WAS CONVERTED TO OPEN SURGERY FOR REMOVAL OF THE ANVIL. THE ANVIL WAS REMOVED THROUGH THE OPENING OF THE CAVITY. THE SURGICAL TIME WAS EXTENDED BY ONE HOUR AND A NEW CIRCULAR STAPLER WAS USED TO COMPLETE THE CASE.

    Patient problems: Laceration(s)

    MAUDE report 25474130 ↗
  • MalfunctionMay 22, 2026Failure to Form Staple

    ACCORDING TO THE REPORTER, DURING THE LAPAROSCOPIC RIGHT HEMICOLECTOMY, AT THE TIME OF ANASTOMOSIS, THERE WAS POOR STAPLE FORMATION. ADDITIONAL RESECTION AND RE-STAPLING WAS PERFORMED USING A NEW STAPLER.

    Patient problems: Failure to Anastomose

    MAUDE report 25263081 ↗
  • MalfunctionApr 27, 2026Misfire

    IT WAS REPORTED THAT DURING A GASTROTOMY, THE TISSUE DONUT WAS INCOMPLETE AFTER FIRING. A NEW DEVICE WAS USED TO RESOLVE THE ISSUE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: Failure to Anastomose; No Clinical Signs, Symptoms or Conditions

    MAUDE report 25000629 ↗
  • InjuryApr 20, 2026Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT POST OPERATIVELY, THE PATIENT EXPERIENCED RECTAL BLEEDING AFTER USING THE TWO CIRCULAR STAPLERS WITH BLEEDING REPORTED AT THE STAPLE LINE AND AT THE ANASTOMOSIS. THE BLEEDING THEN STOPPED SPONTANEOUSLY.

    Patient problems: Hemorrhage/Blood Loss/Bleeding; Tissue Breakdown

    MAUDE report 24932061 ↗
  • InjuryMar 31, 2026Detachment of Device or Device Component

    IT WAS REPORTED THAT DURING A LAPAROSCOPIC PROCEDURE, AFTER FIRING, THE STAPLER¿S ANVIL REMAINED INSIDE THE LEFT COLON. THE PROCEDURE WAS CONVERTED TO AN OPEN SURGERY TO RETRIEVE THE ANVIL FROM INSIDE THE COLON, FOLLOWED BY DISMANTLING AND REPAIR UPSTREAM AND DOWNSTREAM OF THE ANASTOMOSIS.

    Patient problems: Laceration(s)

    MAUDE report 24746478 ↗
  • MalfunctionMar 19, 2026Failure to Fire

    IT WAS REPORTED THAT DURING A LOW ANTERIOR RESECTION, THAT THE INSTRUMENT DID NOT FIRE. THE HANDLE WAS FULLY SQUEEZED UNTIL IT CAME IN CONTACT WITH THE INSTRUMENT BODY. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24637635 ↗
  • MalfunctionMar 17, 2026Failure to Form Staple

    IT WAS REPORTED THAT DURING THE PROCEDURE, AN INCOMPLETE STAPLE LINE (PARTIALLY FIRED) OCCURRED. THE DEVICE WAS REPLACED WITH ANOTHER CIRCULAR STAPLER TO RESOLVE THE ISSUE.

    Patient problems: Failure to Anastomose; Tissue Breakdown

    MAUDE report 24623905 ↗
  • MalfunctionMar 11, 2026Failure to Fire

    IT WAS REPORTED THAT DURING AN OPEN LEFT HEMICOLECTOMY PROCEDURE, AT THE TIME OF DIGESTIVE TRACT RECONSTRUCTION, THE SPACE BETWEEN THE CARTRIDGE AND ANVIL WAS CLOSED, AND THE GREEN BAR WAS VISIBLE IN THE INDICATOR WINDOW, BUT THE STAPLER DID FIRE. THE ISSUE WAS RESOLVED BY REPLACING THE STAPLER WITH ANOTHER DEVICE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24568036 ↗
  • MalfunctionMar 4, 2026Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT DURING USE, AFTER FIRING, THE STAPLE LINE WAS NOT HOLDING STEADY. THE DEVICE WAS REPLACED BY A NEW CIRCULAR STAPLER TO RESOLVE THE ISSUE.

    Patient problems: Failure to Anastomose; Tissue Breakdown

    MAUDE report 24504252 ↗
  • MalfunctionMar 2, 2026Mechanics Altered

    IT WAS REPORTED THAT DURING A LAPAROSCOPIC ANASTOMOSIS INTESTINAL PROCEDURE, WHEN ATTEMPTING TO PERFORM THE ANASTOMOSIS, THE INTESTINE ONLY TORE AND DID NOT STAPLE. SUTURING WAS DONE AS A STAPLE LINE REPLACEMENT AND TO RESOLVE THE ISSUE. IT WAS NOTED THAT THE SURGICAL TIME EXTENDED FOR 30 MINUTES OR MORE.

    Patient problems: Failure to Anastomose

    MAUDE report 24484316 ↗
  • MalfunctionFeb 27, 2026Mechanics Altered

    IT WAS REPORTED THAT DURING A PROCEDURE, THE DEVICE CUT BUT THE STAPLES DID NOT DEPLOY. SUTURING WAS PERFORMED AS A STAPLE LINE REPLACEMENT TO ADDRESS THE ISSUE.

    Patient problems: Failure to Anastomose

    MAUDE report 24469213 ↗
  • MalfunctionFeb 24, 2026Failure to Fire

    ACCORDING TO THE REPORTER, DURING AN OPEN PROCEDURE FOR RECTAL CANCER INVOLVING AN END-TO-END RECTAL ANASTOMOSIS, THE DEVICE DID NOT FIRE. THE ISSUE WAS RESOLVED BY REPLACING THE DEVICE WITH A NEW INSTRUMENT TO COMPLETE THE CASE. THERE WAS NO PATIENT INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24430830 ↗
  • MalfunctionFeb 23, 2026Failure to Form Staple; Noise, Audible

    IT WAS REPORTED THAT DURING A ROBOTIC LAPAROSCOPIC HARTMANN'S PROCEDURE REVERSAL, SOME STAPLES WERE FORMED WHILE OTHERS REMAINED UNFORMED IN THE PATIENT. IT WAS ALSO NOTED THAT THE STAPLER DID NOT SOUND NORMAL WHEN FIRED, DESCRIBING IT AS A LOUD CRACK. THE PROCEDURE WAS EXTENDED BY APPROXIMATELY TWO HOURS DUE TO THE ANASTOMOSIS FALLING APART, WHICH WAS ATTRIBUTED TO A FAILURE OF THE STAPLE LINE. SUTURING WAS PERFORMED AS A REPLACEMENT FOR THE STAPLE LINE TO RESOLVE THE ISSUE.

    Patient problems: Failure to Anastomose; Insufficient Information

    MAUDE report 24424785 ↗
  • MalfunctionFeb 18, 2026Misfire

    IT WAS REPORTED THAT DURING A SURGERY, THE CIRCULAR STAPLER SPONTANEOUSLY FIRED A PROJECTILE, WHICH RENDERED THE DEVICE UNUSABLE. THE DEVICE WAS REPLACED WITH NEW MATERIAL. THERE WAS NO PATIENT INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24381059 ↗
  • MalfunctionFeb 12, 2026Failure to Fire; Human-Device Interface Problem

    IT WAS REPORTED THAT INTRAOPERATIVELY, THE INSTRUMENT DID NOT FIRE, AND THERE WAS DIFFICULTY WITH THE SAFETY LEVER. THE DEVICE WAS REPLACED WITH A SIMILAR PRODUCT. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24328794 ↗
  • MalfunctionFeb 11, 2026Failure to Form Staple; Firing Problem

    IT WAS REPORTED THAT DURING THE LAPAROSCOPIC RIGHT HEMICOLECTOMY, THE CIRCULAR STAPLER WAS APPLIED AT THE RIGHT HEMI COLON ANASTOMOSIS SITE. POOR STAPLE FORMATION WAS OBSERVED, WITH SOME STAPLES NOT PERFECTLY FORMED INTO A "B" SHAPE, INDICATING INCOMPLETE STAPLE FORMATION. SLIGHT RESISTANCE WAS FELT AT THE END OF THE HANDLE DURING USE, AND THE ANVIL COULD BE TILTED AFTER FIRING. THE STAPLE LINE WAS RESECTED AND RE-STAPLED AS A CORRECTIVE ACTION, AND THE STAPLER WAS REPLACED WITH A NEW ONE TO COMPLETE THE PROCEDURE.

    Patient problems: Tissue Breakdown

    MAUDE report 24314591 ↗
  • MalfunctionFeb 6, 2026Failure to Form Staple; Firing Problem

    IT WAS REPORTED THAT DURING A LAPAROSCOPIC RIGHT HEMICOLECTOMY, AT THE TIME OF INTESTINAL ANASTOMOSIS, WHEN THE HANDLE WAS FULLY SQUEEZED, THERE WAS A SENSATION OF STICKING, AND IT WAS LATER OBSERVED THAT SOME STAPLES WERE NOT FULLY FORMED INTO A B-SHAPE. THE STAPLE LINE WAS INCOMPLETE. TO RESOLVE THE ISSUE, RESECTION AND RE-SUTURING WAS PERFORMED. THE ISSUE LED TO UNANTICIPATED TISSUE LOSS.

    Patient problems: Failure to Anastomose

    MAUDE report 24277375 ↗
  • MalfunctionFeb 5, 2026Detachment of Device or Device Component

    IT WAS REPORTED THAT AFTER FIRING, THE ANVIL DETACHED FROM THE STAPLER WHILE ATTEMPTING TO RETRACT IT. THE ANVIL WAS MANAGED AND RETRACTED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24265239 ↗
  • MalfunctionJan 28, 2026Detachment of Device or Device Component

    IT WAS REPORTED THAT DURING AN OPEN LOW ANTERIOR RESECTION (LAR) PROCEDURE, WHILE REMOVING THE INSTRUMENT FROM THE RECTUM, THE ANVIL COULD BE TILTED AFTER FIRING AND FELT THE CLICK FOR TILT. HOWEVER, THE ANVIL CAME OFF AFTER TWO FULL TURNS COUNTER CLOCKWISE. A GRASPER WAS USED TO REMOVE IT FROM THE RECTUM. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24202466 ↗
  • InjuryJan 16, 2026Adverse Event Without Identified Device or Use Problem

    ACCORDING TO THE REPORTER, POSTOPERATIVELY OF LAPAROSCOPIC LOW RESECTION OF THE RECTUM, THE PATIENT EXPERIENCED MASSIVE RECTAL BLEEDING AT THE ANASTOMOSIS SITE. COLONOSCOPY REVEALED THE ANASTOMOSIS SITE, AND HEMOSTASIS WAS ACHIEVED IMMEDIATELY WITH AN ENDOSCOPIC TITANIUM CLIP.

    Patient problems: Failure to Anastomose; Hemorrhage/Blood Loss/Bleeding; Tissue Breakdown

    MAUDE report 24095765 ↗

FDA clearance and classification

Cleared through 510(k) K221003 on Sep 30, 2022 (decision: Substantially Equivalent). Submitted by Covidien as “Autosuture EEA Stapler / circular stapler with DST Series Technology / EEA Autosuture Circular Stapler with DST Series Technology / DST Series EEA Stapler / DST EEA”. The FDA review took 178 days. FDA record ↗

Cleared through Special 510(k) K062850 on Oct 23, 2006 (decision: Substantially Equivalent). Submitted by United States Surgical, A Division of Tyco Healthc as “Autosuture Circular Eea Surgical Staplers”. The FDA review took 28 days. FDA record ↗

The FDA regulates product code GDW (staple, implantable) under 21 CFR 878.4750 as a Class II device. 21 CFR 878.4750 ↗

Common questions about the EEA EEA28

Has the EEA EEA28 been recalled?

No FDA recall record in our data names this exact device identifier (DI 10884523005530), based on the FDA recall file dated Nov 27, 2026. Recalls are sometimes announced by product family or lot, so the manufacturer’s own notices are worth checking too.

What problems have been reported with the EEA EEA28?

The FDA's MAUDE database lists 989 adverse event reports for this device, received between Jul 6, 2016 and Aug 13, 2026. In those reports, the most common specific device problems are “Failure to Form Staple” (262), “Difficult to Remove” (155) and “Detachment of Device or Device Component” (130); the most common patient problems are “Failure to Anastomose” (283), “Unspecified Tissue Injury” (144) and “Tissue Damage” (99). Reports are not verified by the FDA and do not prove the device caused the problem.

Have deaths or injuries been reported with the EEA EEA28?

6 reports are classified as deaths, 455 as injuries and 528 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.

Is the EEA EEA28 FDA approved?

It was cleared through Special 510(k) K062850 on Oct 23, 2006; its record lists 2 FDA submissions in total. Strictly, the FDA "clears" 510(k) devices as substantially equivalent to devices already on the market rather than "approving" them; approval is the PMA route.

Is the EEA EEA28 MRI safe?

According to the device's FDA GUDID record, its labeling lists it as MR Conditional: it can be scanned only under the specific conditions given in the labeling. The radiology team should check those conditions before an MRI.

Who makes the EEA EEA28, and is it still sold?

The FDA's device database lists Covidien LP as the company responsible for it (brand EEA, model EEA28). Its record shows it as in commercial distribution.