GIA GIA6038SRecalls & reported problems
Covidien LP · Stapler with DST Series Technology · Class 2
All models of GIA (26)
Data updated Nov 27, 2026 from official FDA records · Editorial policy
Quick summary
This Covidien GIA stapler is a surgical stapler that cuts and staples tissue during operations. Of its 318 reports, 253 are malfunctions and 62 are injuries, mostly describing the stapler failing to fire or firing incorrectly, staples not forming, and parts breaking or detaching. The reports mention 3 deaths. No recalls name this device. Reports fell to 4 in the last 12 months from 22 in the previous 12.
AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.
- Adverse event reports
- 317
- Deaths reported
- 3
- Injuries reported
- 62
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 317 adverse event reports that list this exact device (DI 10884523007237), received between Sep 8, 2016 and Sep 15, 2026 — classified as 3 deaths, 62 injuries, 252 malfunctions.
The most frequently reported specific device problem is “Failure to Fire” (99 reports).
73 other recalls cover devices cleared under the same FDA submission (K111825, K221013); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.
Reports received per month
Most reported problems
Device problems
- Failure to Fire99
- Misfire40
- Adverse Event Without Identified Device or Use Problem33
- Failure to Form Staple33
- Detachment of Device or Device Component28
- Mechanics Altered26
- Break20
- Retraction Problem19
Patient problems
- No Clinical Signs, Symptoms or Conditions112
- No Known Impact Or Consequence To Patient61
- No Consequences Or Impact To Patient55
- No Code Available42
- Tissue Damage23
- Failure to Anastomose21
- Insufficient Information21
- Unspecified Tissue Injury16
Recent adverse event reports
The 25 most recent of 317 reports, in the reporter's own words as published by the FDA.
MalfunctionSep 15, 2026Failure to Form Staple
According to the reporter, during a procedure using a open stapler, the staple line was incomplete. The specific location of the incomplete staple line could not be identified; however, tissue closure was not performed smoothly. To resolve the issue, a new device of the same product type was opened and used.
Patient problems: Tissue Breakdown
MAUDE report 50766668 ↗MalfunctionJan 6, 2026Failure to Fire
ACCORDING TO THE REPORTER, DURING THE PROCEDURE, THE STAPLER DID NOT FIRE. ANOTHER STAPLER WAS USED TO RESOLVE THE ISSUE. THERE WAS NO PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23983891 ↗MalfunctionDec 23, 2025Detachment of Device or Device Component; Detachment of Device or Device Component
IT WAS REPORTED THAT DURING A LAPAROSCOPIC RIGHT COLON RESECTION, THE FIRING KNOB OF THE DEVICE BROKE OFF DURING TRANSECTION. THE BROKEN PART DISENGAGED BUT DID NOT FALL INTO THE PATIENT CAVITY. THE RESECTION WAS COMPLETED USING THE OTHER BLACK KNOB, AND A SECOND STAPLER WAS USED FOR THE SECOND FIRE. THE LAPAROSCOPIC PROCEDURE WAS CONVERTED TO OPEN, WHICH WAS NOT RELATED TO THE DEVICE PROBLEM. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23882380 ↗MalfunctionSep 19, 2025Mechanics Altered; Mechanics Altered
IT WAS REPORTED THAT DURING AN OPEN RIGHT COLECTOMY, THE OPEN STAPLER WAS USED FOR RESECTION OF A SEGMENT OF THE COLON OR POSSIBLY FOR MAKING THE ANASTOMOSIS. THE CASE WAS EXTENDED BY APPROXIMATELY HALF AN HOUR, THE STAPLES DID NOT DEPLOY AND THE KNIFE WAS DEPLOYED WITHOUT STAPLES DURING TISSUE RESECTION. IT REQUIRED SO MUCH MANIPULATION OF THE WHITE FLAG TO GET THE TWO SIDES TOGETHER AND THE SURGEON WOULD NOT HAVE BEEN ABLE TO DO THIS DURING THE CREATION OF THE ANASTOMOSIS. THE TISSUE WAS RESECTED AND RE-STAPLED DUE TO THE DEVICE ISSUE.
Patient problems: Tissue Breakdown
MAUDE report 23092787 ↗MalfunctionSep 10, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT DURING A PROCEDURE, AFTER FIRING THE RELOADABLE STAPLER, THE STAPLE LINE DID NOT HOLD AND SUBSEQUENTLY OPENED UP. THE DEVICE WAS REPLACED AND EXPLANTED.
Patient problems: Tissue Breakdown
MAUDE report 23022864 ↗MalfunctionAug 6, 2025Failure to Fire; Failure to Fire
ACCORDING TO THE REPORTER, PRIOR TO USE, THE DEVICE DID NOT FIRE. A NEW DEVICE WAS USED TO RESOLVE THE ISSUE. THERE WAS NO PATIENT INVOLVED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22728804 ↗MalfunctionJul 29, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
ACCORDING TO THE REPORTER, POSTOPERATIVELY OF SEGMENTAL SMALL BOWEL RESECTION, THERE WAS ANASTOMOTIC LEAK.
Patient problems: Failure to Anastomose
MAUDE report 22640549 ↗MalfunctionJul 25, 2025Misfire; Misfire
ACCORDING TO THE REPORTER, DURING A LAPAROSCOPIC RADICAL COLON CANCER SURGERY, THE DEVICE PARTIALLY FIRED. A NEW DEVICE WAS USED TO RESOLVE THE ISSUE. THERE WAS NO PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22608359 ↗MalfunctionJun 10, 2025Failure to Fire; Component or Accessory Incompatibility; Human-Device Interface Problem; Failure to Fire; Component or Accessory Incompatibility; Human-Device Interface Problem
IT WAS REPORTED THAT DURING A PROCEDURE, THE DEVICE WAS DIFFICULT TO LOAD ON THE SECOND LOADING UNIT. WHEN IT WAS TIME TO STAPLE, THE DEVICE WAS UNABLE TO FIRE AND WAS DIFFICULT TO REMOVE THE RELOAD IN PLACE. THE DEVICE WAS REPLACED TO RESOLVE THE ISSUE. THERE WAS NO PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22175887 ↗MalfunctionApr 30, 2025Retraction Problem; Human-Device Interface Problem; Retraction Problem; Human-Device Interface Problem
IT WAS REPORTED THAT DURING AN OPEN SIGMOID RESECTION, THERE WERE ISSUES WITH CHANGING THE RELOADS ON THREE STAPLERS, LEADING TO THE USE OF PRELOADED RELOADS. AFTER FIRING, THERE WERE DIFFICULTIES RETRACTING THE KNIFE BLADE. THE FIRST DEVICE WAS REPLACED BY TWO OTHER DEVICES, BUT ALL EXHIBITED THE SAME PROBLEM. THE DEVICES WERE REPLACED, AND IT WAS NOTED THAT THERE WERE NO CONSEQUENCES OR IMPACT TO THE PATIENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21937290 ↗MalfunctionApr 30, 2025Retraction Problem; Human-Device Interface Problem; Retraction Problem; Human-Device Interface Problem
IT WAS REPORTED THAT DURING AN OPEN SIGMOID RESECTION, THERE WERE ISSUES WITH CHANGING THE RELOADS ON THREE STAPLERS, LEADING TO THE USE OF PRELOADED RELOADS. AFTER FIRING, THERE WERE DIFFICULTIES RETRACTING THE KNIFE BLADE. THE FIRST DEVICE WAS REPLACED BY TWO OTHER DEVICES, BUT ALL EXHIBITED THE SAME PROBLEM. THE DEVICES WERE REPLACED, AND IT WAS NOTED THAT THERE WERE NO CONSEQUENCES OR IMPACT TO THE PATIENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21937260 ↗MalfunctionApr 30, 2025Retraction Problem; Human-Device Interface Problem; Retraction Problem; Human-Device Interface Problem
IT WAS REPORTED THAT DURING AN OPEN SIGMOID RESECTION, THERE WERE ISSUES WITH CHANGING THE RELOADS ON THREE STAPLERS, LEADING TO THE USE OF PRELOADED RELOADS. AFTER FIRING, THERE WERE DIFFICULTIES RETRACTING THE KNIFE BLADE. THE FIRST DEVICE WAS REPLACED BY TWO OTHER DEVICES, BUT ALL EXHIBITED THE SAME PROBLEM. THE DEVICES WERE REPLACED, AND IT WAS NOTED THAT THERE WERE NO CONSEQUENCES OR IMPACT TO THE PATIENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21937058 ↗MalfunctionApr 18, 2025Failure to Form Staple; Failure to Form Staple
ACCORDING TO THE REPORTER, INTRA-OPERATIVELY, THE STAPLE LINE WAS INCOMPLETE. ANOTHER DEVICE WAS USED TO COMPLETE THE CASE.
Patient problems: Tissue Breakdown
MAUDE report 21866256 ↗MalfunctionMar 25, 2025Failure to Fire; Failure to Fire
IT WAS REPORTED THAT DURING A PROCEDURE, THE STAPLER INSTRUMENT DID NOT FIRE.
Patient problems: Insufficient Information
MAUDE report 21691816 ↗MalfunctionMar 10, 2025Failure to Fire; Failure to Fire
ACCORDING TO THE REPORTER, DURING A RIGHT HEMICOLECTOMY, THE FIRING KNOB COULD NOT BE PUSHED AND THE DEVICE COULD NOT BE FIRED. ANOTHER RELOAD WAS USED BUT DID NOT WORK EITHER. ANOTHER HANDLE WAS REPLACED TO RESOLVE THE ISSUE. THERE WAS NO PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21561441 ↗DeathFeb 20, 2025Mechanics Altered; Adverse Event Without Identified Device or Use Problem; Mechanics Altered; Adverse Event Without Identified Device or Use Problem
THE PRODUCT WAS USED FOR A SMALL BOWEL RESECTION WITH ANASTOMOSIS. DURING SUBSEQUENT SURGERY IT WAS DISCOVERED THAT THE FREE END OF THE SMALL BOWEL STAPLE LINE HAD LEAKAGE ALONG THE STAPLE LINE INDICATIVE OF A FAILED STAPLE LINE AND THE ANASTOMOSIS DID NOT SEAL. IT WAS REPORTED THAT AFTER USE, THE PATIENT EXPERIENCED SURGICAL STAPLER FAILURE, FAILURE TO THE STAPLER DEVICE, BOWEL LEAKAGE, DEFECTIVE DEVICE, PAIN, SHOCK, MENTAL ANGUISH, UNABLE TO PERFORM NORMAL ACTIVITIES, LEUKOCYTOSIS, ASCITES, SEROSITIS, AND DEATH. POST-OPERATIVE PATIENT TREATMENT INCLUDED EXPLORATORY LAPAROTOMY WITH LYSIS OF ADHESIONS, SMALL BOWEL RESECTION WITH PRIMARY ANASTOMOSIS, DIVERTING LOOP ILEOSTOMY, & CT SCAN. INFORMATION RECEIVED INDICATES THE PATIENT IS DECEASED. NO INFORMATION WAS PROVIDED REGARDING THE CIRCUMSTANCES OF EXPIRATION.
Patient problems: Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available; Ascites; Emotional Changes; Failure to Anastomose; Inflammation; Pain; Shock; Unspecified Tissue Injury
MAUDE report 21425379 ↗InjuryFeb 11, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
ACCORDING TO THE REPORTER, POST-OPERATIVELY TO THE RIGHT HEMICOLECTOMY IT WAS FOUND THAT THERE WAS A STAPLE FORMATION ISSUE. THE SEGMENT THAT WAS REMOVED IN THE SECOND SURGERY WAS SENT TO PATHOLOGY AND THE OPENING OF THE STAPLE LINE WAS DETERMINED. A STAPLE LINE REINFORCEMENT WAS DONE TO RESOLVE THE ISSUE. IT WAS NOTED THAT THERE WAS AN EXTENDED HOSPITALIZATION OF TWO WEEKS AND THE PATIENT WAS SENT TO THE INTENSIVE CARE UNIT AS THE PATIENT WAS EXHIBITING SYMPTOMS OF ANASTOMOTIC LEAKAGE. THE PATIENT EXPERIENCED A LIFE THREATENING EVENT DUE TO SEPSIS IN THE ABDOMINAL CAVITY. AS A RESOLUTION ADDITIONAL SURGICAL PROCEDURE WAS PERFORMED.
Patient problems: Tissue Breakdown
MAUDE report 21356219 ↗MalfunctionJan 29, 2025Mechanics Altered; Mechanics Altered
IT WAS REPORTED THAT DURING A SURGICAL PROCEDURE INVOLVING THE GIA6038S STAPLER DEVICE, THE STAPLES DID NOT DEPLOY, RESULTING IN CUTTING WITHOUT STAPLING. THIS LED TO AN EXTENDED SURGICAL TIME BY 30 MINUTES. FECES WERE FOUND IN THE ABDOMEN, NECESSITATING AN INTERNAL WASH AND THE ADMINISTRATION OF EXTRA ANTIBIOTICS. THE ISSUE WAS RESOLVED BY REPLACING THE DEVICE AND RESECTING AND RE-STAPLING THE AFFECTED AREA.
Patient problems: Tissue Breakdown
MAUDE report 21263286 ↗MalfunctionJan 16, 2025Detachment of Device or Device Component; Detachment of Device or Device Component
ACCORDING TO THE REPORTER, DURING A RADICAL RESECTION OF COLON CANCER IN THE OPERATING ROOM, THE HEPATOBILIARY AND PANCREATIC LAPAROSCOPIC SURGERY DEPARTMENT PLANNED TO ANASTOMOSE THE COLON STUMPS. THE MIDDLE BUCKLE OF THE STAPLER POPPED UP BY ITSELF AND COULD NOT BE CLOSED NORMALLY, SO THE STAPLER HAD TO BE REPLACED. THERE WAS NO PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21170518 ↗MalfunctionDec 13, 2024Failure to Fire; Failure to Fire
ACCORDING TO THE REPORTER, DURING A WHIPPLE'S PROCEDURE ON THE THIRD FIRING THE BLACK KNOB WOULD NOT ADVANCE AND THE DEVICE DID NOT FIRE. ANOTHER DEVICE WAS USED TO RESOLVE THE ISSUE. THERE WAS NO PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20924867 ↗MalfunctionNov 28, 2024Mechanics Altered; Mechanics Altered
ACCORDING TO THE REPORTER, DURING A PROCEDURE, THE DEVICE WAS ABLE TO BE FIRE HOWEVER THE STAPLES DID NOT DEPLOY AND THE DEVICE WAS ABLE TO CUT THE TISSUE. ANOTHER DEVICE WAS USED TO RESOLVE THE ISSUE.
Patient problems: Tissue Breakdown
MAUDE report 20806032 ↗MalfunctionNov 22, 2024Failure to Fire; Failure to Fire
ACCORDING TO THE REPORTER, PREOPERATIVELY, THE STAPLER DID NOT FIRE. THE DEVICE WAS REPLACED TO RESOLVE THE ISSUE. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20759816 ↗MalfunctionNov 12, 2024Retraction Problem; Component or Accessory Incompatibility; Retraction Problem; Component or Accessory Incompatibility
ACCORDING TO THE REPORTER, DURING AN OPEN WHIPPLE, DURING RESECTING THE BOWEL, AFTER FIRING THE SECOND TIME, THE KNIFE BLADE DID NOT GO DOWN. A NEW OPEN STAPLER WAS USED TO RESOLVE THE ISSUE. THERE WAS NO PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20670192 ↗MalfunctionNov 12, 2024Entrapment of Device; Entrapment of Device
ACCORDING TO THE REPORTER, DURING A COLON/INTESTINAL RESECTION, AFTER THE RESECTION, THE TISSUE CAME OFF BEFORE OPENING THE HANDLE. NO INTERVENTION WAS PERFORMED. THERE WAS NO PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20670156 ↗InjuryNov 5, 2024Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
ACCORDING TO THE REPORTER, FOLLOWING AN OPEN COLECTOMY, ON THE ANASTOMOSIS, THE STAPLE LINE HAS LEAKS. NO INTERVENTION WAS DONE. HOS PITALIZATION WAS EXTENDED.
Patient problems: Failure to Anastomose; Insufficient Information
MAUDE report 20613311 ↗
FDA clearance and classification
Cleared through Special 510(k) K221013 on May 5, 2022 (decision: Substantially Equivalent). Submitted by Covidien as “GIA Auto Suture Stapler with DST Series Technology”. The FDA review took 30 days. FDA record ↗
Cleared through 510(k) K111825 on Oct 25, 2011 (decision: Substantially Equivalent). Submitted by Covidien, Formerly US Surgical A Divison of Tyco H as “Endo Gia(Tm) Staplers, Dst Series(Tm) Gia(Tm) Staplers, Dst Series(Tm) Ta(Tm) Staplers”. The FDA review took 119 days. FDA record ↗
The FDA regulates product code GDW (staple, implantable) under 21 CFR 878.4750 as a Class II device. 21 CFR 878.4750 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K111825, K221013. One submission often covers many models, so check the product and lot details in each record.
- Z-0748-2021 · Nov 23, 2020 · CompletedCovidien TA Auto Suture Vascular Loading Unit with DST Series Technology 30 mm- V3. Intended for abdominal, gynecological, pediatric and thoracic surgical procedures for resection
- Z-0747-2021 · Nov 23, 2020 · CompletedCovidien TA Auto Suture Vascular Stapler with DST Series Technology 30mm-V 3. Intended for abdominal, gynecological, pediatric and thoracic surgical procedures for resection or tra
- Z-0002-2021 · Aug 12, 2020 · Open, ClassifiedEndo GIA Auto Suture Universal Articulating Loading Unit 30 mm - 2.5 mm, Item Code 030451 - Product Usage: The Endo GIA universal staplers have applications in abdominal, gynecolo
- Z-0008-2021 · Aug 12, 2020 · Open, ClassifiedEndo GIA Auto Suture Universal Articulating Loading Unit 60 mm - 3.5 mm, Item Code 030458 - Product Usage: The Endo GIA universal staplers have applications in abdominal, gynecolo
- Z-0003-2021 · Aug 12, 2020 · Open, ClassifiedEndo GIA Auto Suture Universal Articulating Loading Unit 30 mm - 3.5 mm, Item Code 030452 - Product Usage: The Endo GIA universal staplers have applications in abdominal, gynecolo
- Z-0007-2021 · Aug 12, 2020 · Open, ClassifiedEndo GIA Auto Suture Universal Loading Unit, Item Code 030457 - Product Usage: The Endo GIA universal staplers have applications in abdominal, gynecologic, pediatric and thoracic
- Z-0009-2021 · Aug 12, 2020 · Open, ClassifiedCovidien Best Practices Procedure Kits containing Auto Suture Endo GIA Universal Loading Units, Parent Codes 00Z2760 - Product Usage: The Endo GIA universal staplers have applicat
- Z-0001-2021 · Aug 12, 2020 · Open, ClassifiedEndo GIA Auto Suture Gray Universal Articulating Loading Unit 30 mm - 2.0 mm, Item Code 030450 - Product Usage: The Endo GIA universal staplers have applications in abdominal, gyn
- Z-0006-2021 · Aug 12, 2020 · Open, ClassifiedEndo GIA Auto Suture Universal Articulating Loading Unit 45 mm - 3.5 mm, Item Code 030455 - Product Usage: The Endo GIA universal staplers have applications in abdominal, gynecolo
- Z-0004-2021 · Aug 12, 2020 · Open, ClassifiedEndo GIA Auto Suture Universal Articulating Loading Unit 45 mm - 2.0 mm, Item Code 030453 - Product Usage: The Endo GIA universal staplers have applications in abdominal, gynecolo
- Z-0005-2021 · Aug 12, 2020 · Open, ClassifiedEndo GIA Auto Suture Universal Loading Unit 45 mm - 2.5mm, Item Code 030454 - Product Usage: The Endo GIA universal staplers have applications in abdominal, gynecologic, pediatric
- Z-1264-2020 · Jan 17, 2020 · TerminatedCovidien Endo GIA Ultra Universal Stapler 12 mm-intended for resection, transection in general surgery Item Code: EGIAUSTND
- Z-1831-2019 · May 17, 2019 · Open, ClassifiedEGIA45 CT AR MD THK REL, EGIA45CTAMT
- Z-1834-2019 · May 17, 2019 · Open, ClassifiedSIGNIA TRI-STAPLE 2.0 BLACK 60MM EXTRA THICK RELOA, SIG60AXT
- Z-1833-2019 · May 17, 2019 · Open, ClassifiedSIGNIA TRI-STAPLE 2.0 BLACK 45MM EXTRA THICK RELOA, SIG45AXT
…and 58 more.
Common questions about the GIA GIA6038S
Has the GIA GIA6038S been recalled?
No FDA recall record in our data names this exact device identifier (DI 10884523007237), based on the FDA recall file dated Nov 27, 2026. However, 73 recalls involve devices cleared under the same FDA submission. They may or may not include this model, so check the product, model and lot details in those records.
What problems have been reported with the GIA GIA6038S?
The FDA's MAUDE database lists 317 adverse event reports for this device, received between Sep 8, 2016 and Sep 15, 2026. In those reports, the most common specific device problems are “Failure to Fire” (99), “Misfire” (40) and “Failure to Form Staple” (33); the most common patient problems are “Tissue Damage” (23), “Failure to Anastomose” (21) and “Unspecified Tissue Injury” (16). Reports are not verified by the FDA and do not prove the device caused the problem.
Have deaths or injuries been reported with the GIA GIA6038S?
3 reports are classified as deaths, 62 as injuries and 252 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.
Is the GIA GIA6038S FDA approved?
It was cleared through 510(k) K111825 on Oct 25, 2011; its record lists 2 FDA submissions in total. Strictly, the FDA "clears" 510(k) devices as substantially equivalent to devices already on the market rather than "approving" them; approval is the PMA route.
Is the GIA GIA6038S MRI safe?
According to the device's FDA GUDID record, its labeling lists it as MR Conditional: it can be scanned only under the specific conditions given in the labeling. The radiology team should check those conditions before an MRI.
Who makes the GIA GIA6038S, and is it still sold?
The FDA's device database lists Covidien LP as the company responsible for it (brand GIA, model GIA6038S). Its record shows it as in commercial distribution.
Answers are built automatically from the FDA records on this page. They are not medical advice.
Other STAPLE, IMPLANTABLE devices
- Endo GIA EGIA60AMTCovidien LP · STAPLE, IMPLANTABLE19 deaths10,660 reports
- Endo GIA Ultra EGIAUSTNDCovidien LP · STAPLE, IMPLANTABLE5 deaths9,727 reports
- Signia SIGPHANDLECovidien LP · STAPLE, IMPLANTABLE1 recall3 deaths5,350 reports
- ECHELON FLEX PSEE60AEthicon Endo-Surgery, LLC · STAPLE, IMPLANTABLE1 recall13 deaths3,279 reports
- Endo GIA Ultra EGIAUSHORTCovidien LP · STAPLE, IMPLANTABLE2 deaths3,006 reports
- Endo GIA EGIA45AMTCovidien LP · STAPLE, IMPLANTABLE10 deaths2,967 reports
- Endo GIA Ultra EGIAUXLCovidien LP · STAPLE, IMPLANTABLE2,910 reports
- ENDOPATH ATS45Ethicon Endo-Surgery, LLC · STAPLE, IMPLANTABLE2 deaths2,044 reports
Sources and updates
- Device record: FDA GUDID, file dated Nov 27, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Nov 27, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Nov 27, 2026
- Clearance and classification: FDA 510(k), PMA and product classification databases, files dated Nov 27, 2026
- Page last rebuilt: Nov 27, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page