SmartSite 20350ETRecalls & reported problems
BD Switzerland Sàrl · Low Sorbing Extension Set, 0.2 Micron Low Protein Binding Filter Needle-Free Valve Closed Male Luer Priming Cap (Orange Colored) 51 cm (20 in) · Class 2
All models of SmartSite (5)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
2 open recalls name this device. See the recall details and contact the manufacturer or your healthcare provider.
- Adverse event reports
- 12
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 2
Summary
The FDA's MAUDE database contains 12 adverse event reports that list this exact device (DI 10885403229756), received between Mar 23, 2020 and May 5, 2025 — classified as 12 malfunctions.
The most frequently reported specific device problem is “Leak/Splash” (7 reports).
2 FDA recall records name this device's identifier, 2 still open.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Reports received per month
Recalls naming this device
- May 12, 2025Open, Classified
Sterile, single use closed system drug transfer devices, standalone or on infusion sets or bonded syringes, may leak at the interface between the closed male luer component and the needle-free connector, which may cause procedure delay, treatment interruptions, exposure to hazardous drugs, and/or underdosing of medication.
Z-1970-2025 · Root cause: Process control · Recall details · FDA record ↗
- Oct 11, 2023Open, Classified
Affected infusion sets labeled as free of DEHP actually do contain di(2-ethylhexyl) phthalate (DEHP).
Z-0351-2024 · Root cause: Error in labeling · Recall details · FDA record ↗
Most reported problems
Device problems
- Leak/Splash7
- Fluid/Blood Leak2
- Component Missing1
- Disconnection1
- Material Separation1
Patient problems
- No Clinical Signs, Symptoms or Conditions8
- No Patient Involvement3
- Insufficient Information1
Recent adverse event reports
The 12 most recent of 12 reports, in the reporter's own words as published by the FDA.
MalfunctionMay 5, 2025Leak/Splash; Leak/Splash
IT WAS REPORTED THAT BD ALARIS TEXIUM SMARTSITE LOW SORBING EXTENSION SET WAS LEAKING. THE FOLLOWING INFORMATION WAS RECEIVED BY THE INITIAL REPORTER WITH THE FOLLOWING VERBATIM. IT WAS REPORTED BY CUSTOMER THAT TEXIUM BONDED SET LEAKED UPON EITHER CONNECTION OR DISCONNECTION TO SMARTSITE. DRUG LEAKED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21963391 ↗MalfunctionApr 17, 2025Leak/Splash; Leak/Splash
IT WAS REPORTED THAT BD THE FOLLOWING INFORMATION WAS RECEIVED BY THE INITIAL REPORTER WITH THE FOLLOWING. I HAVE REPRESENTATIVE SAMPLES, IN PACKAGE, NOT THE ONES INVOLVED WITH THE PATIENT. I MET WITH XXXX AND HER STAFF-- WHILE BEING ON SITE AND LOOKING AT TEXIUM, HOW THINGS OCCURRED WITH RECENT PIR'S (2 WEEKS AGO) THE NURSE NOTICED HER GLOVED BEING WET--LOOKED LIKE THE MED (NOT HAZARDOUS) WAS "SEEPING" WHERE THE TEXIUM CONNECTION (BINDED TO THE 203050ET FILTER SET ) TO THE 2000E SMARTSITE---STAND ALONE SMART SITE ATTACHED TO A FEMALE LUER OF A PORT ACCESS NEEDLE. 6-7 INFUSIONS OCCURRING ON OTHER PATIENTS SOME WITH MULTIPLE TEXIUM IN USE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21862419 ↗MalfunctionMar 18, 2025Leak/Splash; Leak/Splash
IT WAS REPORTED THAT BD ALARIS TEXIUM SMARTSITE LOW SORBING EXTENSION SET WAS LEAKING THE FOLLOWING INFORMATION WAS RECEIVED BY THE INITIAL REPORTER WITH THE FOLLOWING VERBATIM. VERBATIM: THE TUBING WITH REF #(B)(4) BD LOW SORBING EXTENSION SET 0.2 MICRON LOW PROTEIN BINDING FILTER IS LEAKING AT THE POINT OF CONNECTION. IT IS A TRACE AMOUNT, BUT IT IS LEAKING. WE HAVE BEEN ATTACHING THE FILTER TUBING TO THE END OF THE CHEMO TUBING SO IT IS ATTACHED CLOSER TO THE PATIENT, BUT I HAD A PATIENT WHO HAD A SMALL WET SPOT ON HIS SHIRT DURING THE INFUSION. I WAS ABLE TO GET HIM CLEANED UP AND WRAPPED THE TEXIUM LUER WITH GAUZE AND TAPE TO PREVENT ANY MORE CHEMO GETTING ONTO THE PATIENT AND WE WERE ABLE TO COMPLETE THE INFUSION. IT HAPPENED AGAIN TODAY. I ATTACHED THE FILTER TUBING TO THE BOTTOM OF THE SECONDARY LINE TO KEEP IT AWAY FROM THE PATIENT AND WHEN I WENT TO CHECK THE LINES, THERE WAS AGAIN A SMALL AMOUNT OF WETNESS AT THE CONNECTION SITE. I HAVE ATTACHED A PHOTO OF THE HOOK UP TODAY WITH AN ARROW SHOWING WHERE THE LEAKING IS HAPPENING. I AGAIN WRAPPED THE TEXIUM LUER WITH GAUZE AND TAPE AND WAS ABLE TO FINISH THE INFUSION. THERE WAS NO LIQUID NOTED ON THE FLOOR OR ACTIVELY DRIPPI
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21639188 ↗MalfunctionMar 18, 2025Leak/Splash; Leak/Splash
IT WAS REPORTED THAT BD ALARIS TEXIUM SMARTSITE LOW SORBING EXTENSION SET WAS LEAKING THE FOLLOWING INFORMATION WAS RECEIVED BY THE INITIAL REPORTER WITH THE FOLLOWING VERBATIM VERBATIM: THE TUBING WITH REF #20350ET BD LOW SORBING EXTENSION SET 0.2 MICRON LOW PROTEIN BINDING FILTER IS LEAKING AT THE POINT OF CONNECTION. IT IS A TRACE AMOUNT, BUT IT IS LEAKING. WE HAVE BEEN ATTACHING THE FILTER TUBING TO THE END OF THE CHEMO TUBING SO IT IS ATTACHED CLOSER TO THE PATIENT, BUT I HAD A PATIENT WHO HAD A SMALL WET SPOT ON HIS SHIRT DURING THE INFUSION. I WAS ABLE TO GET HIM CLEANED UP AND WRAPPED THE TEXIUM LUER WITH GAUZE AND TAPE TO PREVENT ANY MORE CHEMO GETTING ONTO THE PATIENT AND WE WERE ABLE TO COMPLETE THE INFUSION. IT HAPPENED AGAIN TODAY. I ATTACHED THE FILTER TUBING TO THE BOTTOM OF THE SECONDARY LINE TO KEEP IT AWAY FROM THE PATIENT AND WHEN I WENT TO CHECK THE LINES, THERE WAS AGAIN A SMALL AMOUNT OF WETNESS AT THE CONNECTION SITE. I HAVE ATTACHED A PHOTO OF THE HOOK UP TODAY WITH AN ARROW SHOWING WHERE THE LEAKING IS HAPPENING. I AGAIN WRAPPED THE TEXIUM LUER WITH GAUZE AND TAPE AND WAS ABLE TO FINISH THE INFUSION. THERE WAS NO LIQUID NOTED ON THE FLOOR OR ACTIVELY DRIPPI
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21639187 ↗MalfunctionJan 2, 2024Leak/Splash; Leak/Splash
IT WAS REPORTED THAT BD ALARIS TEXIUM SMARTSITE LOW SORBING EXTENSION SET WAS DAMAGED THE FOLLOWING INFORMATION WAS RECEIVED BY THE INITIAL REPORTER WITH THE VERBATIM:WE HAVE EXPERIENCED ANOTHER INCIDENT OF PRECIPITATION IN THE LOW SORBING TEXIUM TUBING(24601-B007T) AS WELL AS THE 0.2 MICRON FILTER WITH TEXIUM(20350ET) DURING A 24 HOUR PACLITAXEL INFUSION. I DO NOT HAVE THE LOTS AVAILABLE TO ME, BUT ATTACHED ARE PHOTOS OF THIS
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18423544 ↗MalfunctionJun 9, 2023Component Missing; Component Missing
IT WAS REPORTED THERE WERE 2 OCCURRENCES OF OPENING BD SMARTSITE¿ EXTENSION SET, LOW SORBING, 0.2 MICRON FILTER THAT THE END CAP WAS MISSING. THERE WAS NO REPORT OF PATIENT IMPACT. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: 2 OF THE IN-LINE FILTER OPENED TODAY HAVE NO WHITE CAP ATTACHED AT THE END OF THE LINE. PICTURE SHOWN IS ONE OF THEM, WE HAVE A TOTAL OF 2. NAME OF THE IN-LINE FILTER IS BD LOW SORBING EXTENSION SET (0.2 MICRON LOW PROTEIN BINDING FILTER), REF (B)(4).
Patient problems: Insufficient Information
MAUDE report 17095862 ↗MalfunctionMay 19, 2023Material Separation; Material Separation
IT WAS REPORTED WHILE USING BD SMARTSITE¿ EXTENSION SET THE COMPONENTS SEPARATED. THERE WAS NO REPORT OF PATIENT IMPACT. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: WE OPENED A NEW IN-LINE FILTER YESTERDAY 4/25/2023, AND IT CAME BROKEN AROUND THE DISCS AREA.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16962170 ↗MalfunctionMay 11, 2023Leak/Splash; Leak/Splash
IT WAS REPORTED, WHILE USING BD ALARIS SECONDARY SET LEAKAGE OCCURRED. THERE WAS NO REPORT OF PATIENT IMPACT. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: CALLED TO REPORT, THAT ON 04/18/23. THEY MADE A CHEMO TAXOL BAG, FOR A PATIENT AT THE (B)(6) CLINIC AND THE FILTER IS LEAKING. SHE IS NOT SURE, IF THERE IS A CRACK AROUND THE FILTER AREA. PRODUCT 20350ET, EXTENSION SET W/.2MF, SMARTSITE & TEXIUM. PHONE (B)(6).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16916736 ↗MalfunctionApr 16, 2023Disconnection; Disconnection
IT WAS REPORTED THAT THE BD ALARIS¿ TEXIUM¿ SMARTSITE¿ LOW SORBING EXTENSION SET LINE DETACHED FROM THE FILTER. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: "ON THE EXTENSION SET THE BLUE LINE WAS DETACH FROM THE WHITE FILTER."
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16748693 ↗MalfunctionSep 27, 2020Leak/Splash; Leak/Splash
IT WAS REPORTED THAT THE EXTENSION SET W/.2MF, SMARTSITE & TEXIUM EXPERIENCED LEAKAGE. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: THE CUSTOMER HAS THE DEFECTIVE SET HOWEVER IT HAS CHEMO IN IT. IT LEAKED WHERE THE TUBING MEETS/IS FUSED WITH THE FILTER.
Patient problems: No Patient Involvement
MAUDE report 10591200 ↗MalfunctionMar 23, 2020Fluid/Blood Leak; Fluid/Blood Leak
IT WAS REPORTED THAT THE TUBING SET LEAKED FROM THE ROUND DISK ON THE FILTER DURING PRIMING OF NORMAL SALINE, BEFORE IT WAS CONNECTED TO THE PATIENT. EVENT OCCURRED IN OUTPATIENT INFUSION CENTER. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Patient Involvement
MAUDE report 9868232 ↗MalfunctionMar 23, 2020Fluid/Blood Leak; Fluid/Blood Leak
IT WAS REPORTED THAT THE TUBING SET LEAKED FROM THE ROUND DISK ON THE FILTER DURING PRIMING OF NORMAL SALINE, BEFORE IT WAS CONNECTED TO THE PATIENT. EVENT OCCURRED IN OUTPATIENT INFUSION CENTER. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Patient Involvement
MAUDE report 9866023 ↗
Other Set, administration, intravascular devices
- Autosoft XC 1001680Unomedical A/S · Set, Administration, Intravascular1 death16,105 reports
- Autosoft90™ 1002817Unomedical A/S · Set, Administration, Intravascular1 recall11,907 reports
- Minimed™ Quick-set™ MMT-399AUnomedical A/S · Set, Administration, Intravascular9 deaths8,669 reports
- Medtronic Extended MMT-431AUnomedical A/S · Set, Administration, Intravascular3 deaths8,574 reports
- MiniMed™ Mio™ Advance MMT-242AUnomedical A/S · Set, Administration, Intravascular15 deaths8,164 reports
- CADD 21-7302-24Icu Medical, Inc. · Set, administration, intravascular4 recalls1 death4,657 reports
- Autosoft XC 1001681Unomedical A/S · Set, Administration, Intravascular1 death4,216 reports
- Minimed™ Quick-set™ MMT-397AUnomedical A/S · Set, Administration, Intravascular19 deaths4,134 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026 · each recall links to the FDA recall notice
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page