SmartSiteRecalls & reported problems across all models
BD Switzerland Sàrl · 5 models in the FDA device registry
Data updated Oct 16, 2026 from official FDA records · Editorial policy
2 open recalls name at least one SmartSite model. Check whether yours is affected in the recall details, and contact the manufacturer or your healthcare provider.
- Adverse event reports
- 51
- Deaths reported
- 0
- Injuries reported
- 2
- Recalls
- 3
We found 5 models of SmartSite by BD Switzerland Sàrl in the FDA's device registry (GUDID). Together they are named in 51 adverse event reports in the FDA's MAUDE database, received between Jun 11, 2018 and May 5, 2025 — classified as 2 injuries, 49 malfunctions.
The most frequently reported specific device problem is “Fluid/Blood Leak” (33 reports).
3 FDA recall records name at least one of these models by its device identifier, 2 still open.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event. Widely used products collect more reports, so totals can't be used to compare products or estimate failure rates. Only reports that name a model's exact device identifier are counted. What reports can and can't tell you · How this data works
Reports received per month, all models
SmartSite recalls
FDA recall records that name at least one of these models by its device identifier, newest first. Each recall usually covers specific models or lots, so check the details.
- May 12, 2025Open, Classified
Sterile, single use closed system drug transfer devices, standalone or on infusion sets or bonded syringes, may leak at the interface between the closed male luer component and the needle-free connector, which may cause procedure delay, treatment interruptions, exposure to hazardous drugs, and/or underdosing of medication.
Z-1970-2025 · Root cause: Process control · Recall details · FDA record ↗
- Oct 11, 2023Open, Classified
Affected infusion sets labeled as free of DEHP actually do contain di(2-ethylhexyl) phthalate (DEHP).
Z-0351-2024 · Root cause: Error in labeling · Recall details · FDA record ↗
- May 6, 2019Terminated
An incomplete occlusion can be caused by a variation in the wall thickness of the pumping segment of the affected infusion sets used with the. The issue has the potential to lead to unintended delivery of medication when the pump module is not in running status or can result in faster than expected drug delivery flow when the pump is infusing.
Z-1770-2019 · Root cause: Nonconforming Material/Component · Recall details · FDA record ↗
Most reported SmartSite problems
Device problems
- Fluid/Blood Leak33
- Leak/Splash8
- Material Separation7
- Material Puncture/Hole2
- Component Missing1
- Contamination1
- Crack1
- Defective Component1
Patient problems
- No Consequences Or Impact To Patient27
- No Clinical Signs, Symptoms or Conditions11
- No Information5
- No Known Impact Or Consequence To Patient4
- No Patient Involvement4
- Insufficient Information2
SmartSite models
5 models, most reported first. Each model has its own page with its reports, recalls and device details.
- SmartSite 20350ETSet, administration, intravascular2 recalls12 reports
- SmartSite 2477-0000Set, administration, intravascular1 recall35 reports
- SmartSite 10012143Set, administration, intravascular2 reports
- SmartSite 10237425Stopcock, i.V. Set1 report
- SmartSite 10016240Set, administration, intravascular1 report
Recent SmartSite adverse event reports
The 15 most recent reports across all models, in the reporter's own words as published by the FDA.
MalfunctionMay 5, 2025Leak/Splash; Leak/Splash
IT WAS REPORTED THAT BD ALARIS TEXIUM SMARTSITE LOW SORBING EXTENSION SET WAS LEAKING. THE FOLLOWING INFORMATION WAS RECEIVED BY THE INITIAL REPORTER WITH THE FOLLOWING VERBATIM. IT WAS REPORTED BY CUSTOMER THAT TEXIUM BONDED SET LEAKED UPON EITHER CONNECTION OR DISCONNECTION TO SMARTSITE. DRUG LEAKED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21963391 ↗MalfunctionApr 17, 2025Leak/Splash; Leak/Splash
IT WAS REPORTED THAT BD THE FOLLOWING INFORMATION WAS RECEIVED BY THE INITIAL REPORTER WITH THE FOLLOWING. I HAVE REPRESENTATIVE SAMPLES, IN PACKAGE, NOT THE ONES INVOLVED WITH THE PATIENT. I MET WITH XXXX AND HER STAFF-- WHILE BEING ON SITE AND LOOKING AT TEXIUM, HOW THINGS OCCURRED WITH RECENT PIR'S (2 WEEKS AGO) THE NURSE NOTICED HER GLOVED BEING WET--LOOKED LIKE THE MED (NOT HAZARDOUS) WAS "SEEPING" WHERE THE TEXIUM CONNECTION (BINDED TO THE 203050ET FILTER SET ) TO THE 2000E SMARTSITE---STAND ALONE SMART SITE ATTACHED TO A FEMALE LUER OF A PORT ACCESS NEEDLE. 6-7 INFUSIONS OCCURRING ON OTHER PATIENTS SOME WITH MULTIPLE TEXIUM IN USE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21862419 ↗MalfunctionMar 18, 2025Leak/Splash; Leak/Splash
IT WAS REPORTED THAT BD ALARIS TEXIUM SMARTSITE LOW SORBING EXTENSION SET WAS LEAKING THE FOLLOWING INFORMATION WAS RECEIVED BY THE INITIAL REPORTER WITH THE FOLLOWING VERBATIM. VERBATIM: THE TUBING WITH REF #(B)(4) BD LOW SORBING EXTENSION SET 0.2 MICRON LOW PROTEIN BINDING FILTER IS LEAKING AT THE POINT OF CONNECTION. IT IS A TRACE AMOUNT, BUT IT IS LEAKING. WE HAVE BEEN ATTACHING THE FILTER TUBING TO THE END OF THE CHEMO TUBING SO IT IS ATTACHED CLOSER TO THE PATIENT, BUT I HAD A PATIENT WHO HAD A SMALL WET SPOT ON HIS SHIRT DURING THE INFUSION. I WAS ABLE TO GET HIM CLEANED UP AND WRAPPED THE TEXIUM LUER WITH GAUZE AND TAPE TO PREVENT ANY MORE CHEMO GETTING ONTO THE PATIENT AND WE WERE ABLE TO COMPLETE THE INFUSION. IT HAPPENED AGAIN TODAY. I ATTACHED THE FILTER TUBING TO THE BOTTOM OF THE SECONDARY LINE TO KEEP IT AWAY FROM THE PATIENT AND WHEN I WENT TO CHECK THE LINES, THERE WAS AGAIN A SMALL AMOUNT OF WETNESS AT THE CONNECTION SITE. I HAVE ATTACHED A PHOTO OF THE HOOK UP TODAY WITH AN ARROW SHOWING WHERE THE LEAKING IS HAPPENING. I AGAIN WRAPPED THE TEXIUM LUER WITH GAUZE AND TAPE AND WAS ABLE TO FINISH THE INFUSION. THERE WAS NO LIQUID NOTED ON THE FLOOR OR ACTIVELY DRIPPI
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21639188 ↗MalfunctionMar 18, 2025Leak/Splash; Leak/Splash
IT WAS REPORTED THAT BD ALARIS TEXIUM SMARTSITE LOW SORBING EXTENSION SET WAS LEAKING THE FOLLOWING INFORMATION WAS RECEIVED BY THE INITIAL REPORTER WITH THE FOLLOWING VERBATIM VERBATIM: THE TUBING WITH REF #20350ET BD LOW SORBING EXTENSION SET 0.2 MICRON LOW PROTEIN BINDING FILTER IS LEAKING AT THE POINT OF CONNECTION. IT IS A TRACE AMOUNT, BUT IT IS LEAKING. WE HAVE BEEN ATTACHING THE FILTER TUBING TO THE END OF THE CHEMO TUBING SO IT IS ATTACHED CLOSER TO THE PATIENT, BUT I HAD A PATIENT WHO HAD A SMALL WET SPOT ON HIS SHIRT DURING THE INFUSION. I WAS ABLE TO GET HIM CLEANED UP AND WRAPPED THE TEXIUM LUER WITH GAUZE AND TAPE TO PREVENT ANY MORE CHEMO GETTING ONTO THE PATIENT AND WE WERE ABLE TO COMPLETE THE INFUSION. IT HAPPENED AGAIN TODAY. I ATTACHED THE FILTER TUBING TO THE BOTTOM OF THE SECONDARY LINE TO KEEP IT AWAY FROM THE PATIENT AND WHEN I WENT TO CHECK THE LINES, THERE WAS AGAIN A SMALL AMOUNT OF WETNESS AT THE CONNECTION SITE. I HAVE ATTACHED A PHOTO OF THE HOOK UP TODAY WITH AN ARROW SHOWING WHERE THE LEAKING IS HAPPENING. I AGAIN WRAPPED THE TEXIUM LUER WITH GAUZE AND TAPE AND WAS ABLE TO FINISH THE INFUSION. THERE WAS NO LIQUID NOTED ON THE FLOOR OR ACTIVELY DRIPPI
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21639187 ↗MalfunctionJan 30, 2024Leak/Splash; Leak/Splash
IT WAS REPORTED THAT BD ALARIS PUMP MODULE SMARTSITE BLOOD SET WAS LEAKING THE FOLLOWING INFORMATION WAS RECEIVED BY THE INITIAL REPORTER WITH THE VERBATIM: ON (B)(6) 2023. A PATIENT IN OUR MEDICAL DAY CLINIC WAS RECEIVING A UNIT OF RED BLOOD CELLS AND WHEN THE NURSE WAS DOING THEIR 15 MINUTES POST START OF TRANSFUSION ASSESSMENT NOTED THAT THE TUBING JUST ABOVE THE NONVENTED DRIP CHAMBER WAS LEAKING BLOOD AT 1105H. THE TRANSFUSION WAS STOPPED AND THE UNIT OF RED BLOOD CELLS WITH TUBING WAS SENT BACK TO OUR TRANSFUSION MEDICINE DEPARTMENT. I HAVE REACHED OUT TO THE OUR TRANSFUSION MEDICINE DEPARTMENT TO SEE IF THE TUBING IS STILL AVAILABLE TO SEND BACK TO BD FOR INVESTIGATION AND WILL LET YOU IF IT IS.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18607395 ↗MalfunctionJan 2, 2024Leak/Splash; Leak/Splash
IT WAS REPORTED THAT BD ALARIS TEXIUM SMARTSITE LOW SORBING EXTENSION SET WAS DAMAGED THE FOLLOWING INFORMATION WAS RECEIVED BY THE INITIAL REPORTER WITH THE VERBATIM:WE HAVE EXPERIENCED ANOTHER INCIDENT OF PRECIPITATION IN THE LOW SORBING TEXIUM TUBING(24601-B007T) AS WELL AS THE 0.2 MICRON FILTER WITH TEXIUM(20350ET) DURING A 24 HOUR PACLITAXEL INFUSION. I DO NOT HAVE THE LOTS AVAILABLE TO ME, BUT ATTACHED ARE PHOTOS OF THIS
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18423544 ↗MalfunctionSep 25, 2023Defective Component; Defective Component
IT WAS REPORTED THAT BD SMART SITE EXTENSION SET WAS DAMAGED. THE FOLLOWING INFORMATION WAS RECEIVED BY THE INITIAL REPORTER WITH THE VERBATIM: DESCRIBE THE EVENT OR PROBLEM: ADULT FEMALE GESTITIONAL AGE 40W6D RECEIVING PITOCIN DURING L&D; RN NOTICED BEDDING WAS SOAKING AND NOTED PITOCIN DRAINING FROM HUB ON Y CONNECTOR AS IT WAS CRACKED. PT HAD PROLONGED SECOND STAGE, VAC ASSIST W/3 POP-OFFS AND POTENTIAL C/S BUT BABY DELIVERED VAGINALLY FROM ROA POSITION. MOM & BABY TOLERATED DELIVERY WELL, NO COMPLICATIONS.
Patient problems: Insufficient Information; No Clinical Signs, Symptoms or Conditions
MAUDE report 17808048 ↗MalfunctionJun 12, 2023Contamination; Contamination
IT WAS REPORTED BY CUSTOMER THAT THE BD 20MM SMARTSITE VAD-ZYMOGENETICS HAD SOME TYPE OF DEBRIS IN PACKAGING. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: "SOME TYPE OF DEBRIS IN PACKAGING".
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17107571 ↗MalfunctionJun 9, 2023Component Missing; Component Missing
IT WAS REPORTED THERE WERE 2 OCCURRENCES OF OPENING BD SMARTSITE¿ EXTENSION SET, LOW SORBING, 0.2 MICRON FILTER THAT THE END CAP WAS MISSING. THERE WAS NO REPORT OF PATIENT IMPACT. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: 2 OF THE IN-LINE FILTER OPENED TODAY HAVE NO WHITE CAP ATTACHED AT THE END OF THE LINE. PICTURE SHOWN IS ONE OF THEM, WE HAVE A TOTAL OF 2. NAME OF THE IN-LINE FILTER IS BD LOW SORBING EXTENSION SET (0.2 MICRON LOW PROTEIN BINDING FILTER), REF (B)(4).
Patient problems: Insufficient Information
MAUDE report 17095862 ↗MalfunctionMay 19, 2023Material Separation; Material Separation
IT WAS REPORTED WHILE USING BD SMARTSITE¿ EXTENSION SET THE COMPONENTS SEPARATED. THERE WAS NO REPORT OF PATIENT IMPACT. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: WE OPENED A NEW IN-LINE FILTER YESTERDAY 4/25/2023, AND IT CAME BROKEN AROUND THE DISCS AREA.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16962170 ↗MalfunctionMay 11, 2023Leak/Splash; Leak/Splash
IT WAS REPORTED, WHILE USING BD ALARIS SECONDARY SET LEAKAGE OCCURRED. THERE WAS NO REPORT OF PATIENT IMPACT. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: CALLED TO REPORT, THAT ON 04/18/23. THEY MADE A CHEMO TAXOL BAG, FOR A PATIENT AT THE (B)(6) CLINIC AND THE FILTER IS LEAKING. SHE IS NOT SURE, IF THERE IS A CRACK AROUND THE FILTER AREA. PRODUCT 20350ET, EXTENSION SET W/.2MF, SMARTSITE & TEXIUM. PHONE (B)(6).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16916736 ↗MalfunctionApr 16, 2023Disconnection; Disconnection
IT WAS REPORTED THAT THE BD ALARIS¿ TEXIUM¿ SMARTSITE¿ LOW SORBING EXTENSION SET LINE DETACHED FROM THE FILTER. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: "ON THE EXTENSION SET THE BLUE LINE WAS DETACH FROM THE WHITE FILTER."
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16748693 ↗MalfunctionSep 27, 2020Leak/Splash; Leak/Splash
IT WAS REPORTED THAT THE EXTENSION SET W/.2MF, SMARTSITE & TEXIUM EXPERIENCED LEAKAGE. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: THE CUSTOMER HAS THE DEFECTIVE SET HOWEVER IT HAS CHEMO IN IT. IT LEAKED WHERE THE TUBING MEETS/IS FUSED WITH THE FILTER.
Patient problems: No Patient Involvement
MAUDE report 10591200 ↗MalfunctionMar 23, 2020Fluid/Blood Leak; Fluid/Blood Leak
IT WAS REPORTED THAT THE TUBING SET LEAKED FROM THE ROUND DISK ON THE FILTER DURING PRIMING OF NORMAL SALINE, BEFORE IT WAS CONNECTED TO THE PATIENT. EVENT OCCURRED IN OUTPATIENT INFUSION CENTER. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Patient Involvement
MAUDE report 9868232 ↗MalfunctionMar 23, 2020Fluid/Blood Leak; Fluid/Blood Leak
IT WAS REPORTED THAT THE TUBING SET LEAKED FROM THE ROUND DISK ON THE FILTER DURING PRIMING OF NORMAL SALINE, BEFORE IT WAS CONNECTED TO THE PATIENT. EVENT OCCURRED IN OUTPATIENT INFUSION CENTER. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Patient Involvement
MAUDE report 9866023 ↗
Other BD Switzerland Sàrl products
- Alaris, SmartSite160 models · 10,227 reports · 8 recalls
- Alaris85 models · 2,797 reports · 10 recalls
- MaxZero47 models · 2,377 reports
- MaxGuard49 models · 780 reports · 11 recalls
- BD PhaSeal23 models · 992 reports · 2 recalls
- MaxPlus18 models · 762 reports · 2 recalls
- BD SmartSite29 models · 568 reports · 4 recalls
- Alaris, SmartSite, Texium12 models · 291 reports · 5 recalls
Sources and updates
- Models: FDA GUDID, file dated Oct 13, 2026. Models are grouped by the brand name the manufacturer registered.
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page