BD Alaris 8015Recalls & reported problems

Carefusion 303, Inc. · BD Alaris PC Unit (REM V12.3+) · Class 2

Adverse event reports
1
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 10885403517921), received between May 28, 2024 and May 28, 2024 — classified as 1 malfunction.

The most frequently reported specific device problem is “Computer Operating System Problem” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 1 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Computer Operating System Problem1
  • Device Alarm System1

Patient problems

  • No Clinical Signs, Symptoms or Conditions1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • MalfunctionMay 28, 2024Device Alarm System; Computer Operating System Problem; Device Alarm System; Computer Operating System Problem

    IT WAS REPORTED THAT THE PC UNIT (SOFTWARE VERSION 12.3.1.2) HAD A SYSTEM ERROR (ERROR CODE 9-1513-202) DURING SIMULATION TESTING. THE USER WAS INITIALLY ATTEMPTED TO RECREATE AN ISSUE ¿WITHOUT A MAX IN THE PCA AND THEN ADDING IT IN AFTER THE INFUSION STARTED¿ AND THE USER WAS UNABLE TO RECREATE A SYSTEM ERROR AT THAT TIME¿. A FEW HOURS LATER, THE USER STATED THAT THE DEVICE ¿SUDDENLY GOT A SYSTEM ERROR¿. THE USER WAS UNABLE TO SILENCE THE ALARM. NONE OF THE MODULES WERE INFUSING AT THE TIME. THE PROGRAMMING THAT THE USER WAS DOING WAS ¿KETAMINE GREATER THAN 45KG CONCENTRATION 500MG/50ML PCA DOSE 10 MG LOCKOUT INTERVAL 15 MINUTES AND CONTINUOUS DOSE 15MG/HR. NO MAX AND THEN START AND ONCE STARTED WENT BACK AND ADDED A MAX¿. THE USER REPORTEDLY INITIALLY DID THIS SEVERAL TIMES WITHOUT AN ERROR. AS PREVIOUSLY STATED, A FEW HOURS LATER, WHILE SITTING IN STANDBY WITH DEVICE TURNED ON WITHOUT AN INFUSION RUNNING, IT WENT INTO SYSTEM ERROR. THERE WAS NO PATIENT INVOLVEMENT AS THE EVENT OCCURRED IN A TEST ENVIRONMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19402558 ↗

Other Pump, infusion devices

All Pump, infusion devices →

Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.