Skip to content

Equinoxe 308-01-08Recalls & reported problems

Exactech, Inc. · Class 2 · Implant

All models of Equinoxe (305)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
1
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 10885862267894), received between Dec 20, 2019 and Dec 20, 2019 — classified as 1 injury.

The most frequently reported specific device problem is “Loosening of Implant Not Related to Bone-Ingrowth” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Most reported problems

Device problems

  • Loosening of Implant Not Related to Bone-Ingrowth1

Patient problems

  • Failure of Implant1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • InjuryDec 20, 2019Loosening of Implant Not Related to Bone-Ingrowth; Loosening of Implant Not Related to Bone-Ingrowth

    THIS (B)(6) PATIENT WITH AN AVERAGE BUILD HAD AN INITIAL SURGERY "A COUPLE YEARS AGO". THE REPORTED INDICATION FOR THE INITIAL IMPLANTATION WAS DUE TO REMOVAL OF A TUMOR IN THIS RIGHT SHOULDER (ONCOLOGY). ON (B)(6) 2019, IT WAS NOTED THAT THE DISTAL STEM HAD BECOME LOOSE. THE STEM, COLLAR AND PROXIMAL BODY WERE REVISED. THE TRAY HEIGHT WAS ALSO BUILT UP. ALL AVAILABLE INFORMATION HAS BEEN RECEIVED. DEVICES ARE NOT AVAILABLE FOR RETURNED. UPON REVIEW OF ALL THE INFORMATION, THE CAUSE OF THE LOOSENING IS LIKELY RELATED TO THE PATIENT¿S PREVIOUS CANCER WHICH LIKELY RESULTED IN POOR BONE QUALITY (PRE-EXISTING CONDITION).

    Patient problems: Failure of Implant

    MAUDE report 9503593 ↗

Other Prosthesis, Shoulder, Hemi-, Humeral, Metallic Uncemented devices

All Prosthesis, Shoulder, Hemi-, Humeral, Metallic Uncemented devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page