Equinoxe 308-01-13Recalls & reported problems

Exactech, Inc. · Class 2 · Implant

All models of Equinoxe (305)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
1
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 10885862267924), received between Jun 19, 2024 and Jun 19, 2024 — classified as 1 injury.

The most frequently reported specific device problem is “Loosening of Implant Not Related to Bone-Ingrowth” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 1 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Adverse Event Without Identified Device or Use Problem1
  • Loosening of Implant Not Related to Bone-Ingrowth1

Patient problems

  • Medical device site infection1
  • Pain1
  • Septic loosening1
  • Unspecified Infection1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • InjuryJun 19, 2024Adverse Event Without Identified Device or Use Problem; Loosening of Implant Not Related to Bone-Ingrowth; Adverse Event Without Identified Device or Use Problem; Loosening of Implant Not Related to Bone-Ingrowth

    AS REPORTED BY THE EQUINOXE SHOULDER STUDY, APPROXIMATELY 1 YEAR(S) AND 3 MONTH(S) POST IMPLANTATION OF LEFT A THIRD REVISED TSA (PREVIOUS REVISION REPORTED ON MFR# 1038671-2024-01824), THE PATIENT PRESENTED WITH ASEPTIC HUMERAL LOOSENING. PATIENT HAD REVISION TO A HRP AND HAS NEVER HAD RELIEF OF PAIN. A BONE SCAN WAS OBTAINED WHICH WAS RETURNED SUSPICIOUS FOR LOOSE PROSTHESIS AND NON-SPECIFIC FOCAL INFECTION. THIS OUTCOME OF THE EVENT WAS RESOLVED TWO MONTHS LATER BY STANDARD REVERSE WITH HUMERAL RECONSTRUCTION STEM REVISION OF THE LEFT SHOULDER AND S/P ANTIBIOTIC SPACER PLACEMENT FIVE MONTHS FOLLOWING WITH CONTINUED LIFT SHOULDER PAIN. PLAN TO REVISE BACK TO RTSA W/HRP. PAIN COULD BE RESULT OF CONTINUED GLENOID WEAR FROM THE SPACER. THE CLINICAL REPORT INDICATES THAT THE EVENT IS DEFINITELY RELATED TO THE DEVICE(S) AND UNLIKELY RELATED TO THE PROCEDURE. THIS EVENT REPORT WAS RECEIVED THROUGH CLINICAL DATA COLLECTION ACTIVITIES AND NO DEVICE RETURN IS ANTICIPATED.

    Patient problems: Medical device site infection; Pain; Septic loosening; Unspecified Infection

    MAUDE report 19568878 ↗

Other Prosthesis, Shoulder, Hemi-, Humeral, Metallic Uncemented devices