Skip to content

Equinoxe 308-10-05Recalls & reported problems

Exactech, Inc. · Class 2 · Implant

All models of Equinoxe (305)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
1
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 10885862268129), received between Jul 3, 2019 and Jul 3, 2019 — classified as 1 injury.

The most frequently reported specific device problem is “Device Dislodged or Dislocated” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Most reported problems

Device problems

  • Adverse Event Without Identified Device or Use Problem1
  • Device Dislodged or Dislocated1

Patient problems

  • Joint Dislocation1
  • Unspecified Infection1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • InjuryJul 3, 2019Adverse Event Without Identified Device or Use Problem; Device Dislodged or Dislocated; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT A MALE PATIENT RECEIVED AN HRP PROSTHESIS DURING REVISION OF A COMPETITOR¿S DEVICE DUE TO OSTEONECROSIS OF THE PROXIMAL HUMERUS. DURING AN INITIAL REVISION ATTEMPT EARLIER THE SAME MONTH, AN INFECTION WAS FOUND, AND A SPACER IMPLANTED. THE TUBERCULII WERE FULLY NECROTIZED AND THE ROTATOR CUFF NOT RE-FIXABLE. DUE TO THIS, A LARGE PROXIMAL HUMERAL BODY IN 0°RETROVERSION AND A RTSA WAS IMPLANTED. AT THE TIME OF HIS LAST FOLLOW-UP, THE PATIENT WAS ABLE TO ELEVATE HIS ARM UP TO 120°, WITHOUT ANY SIGNIFICANT ROTATION POSSIBLE. ALTHOUGH SATISFIED WITH THE FUNCTION, THE JOINT TENDS TO DISLOCATE ANTERIORLY WHILE SLEEPING AND UNDER STRESS CAUSING ANTERIOR MOVEMENT OF THE PROXIMAL HUMORUS. THE SURGEON BELIEVES THIS IS DUE TO THE 0° RETROVERSION IN COMBINATION WITH A TOO LARGE PROXIMAL HUMERUS AND IS PLANNING TO IMPLANT A SMALLER SIZE OF THE PROXIMAL HUMERUS IN CONNECTION WITH MORE RETROVERSION AND INCREASE STABILITY WITH THE HELP OF THE REVERSE IMPLANTS IN THE FUTURE.

    Patient problems: Joint Dislocation; Unspecified Infection

    MAUDE report 8758403 ↗

Other Prosthesis, Shoulder, Hemi-, Humeral, Metallic Uncemented devices

All Prosthesis, Shoulder, Hemi-, Humeral, Metallic Uncemented devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page