Skip to content

Vantage 350-21-11Recalls & reported problems

Exactech, Inc. · Class 2 · Implant

All models of Vantage (66)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
1
Deaths reported
0
Injuries reported
1
Recalls naming it
1

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 10885862276254), received between Jun 7, 2023 and Jun 7, 2023 — classified as 1 injury.

The most frequently reported specific device problem is “Naturally Worn” (1 report).

1 FDA recall record name this device's identifier, 1 still open.

5 other recalls cover devices cleared under the same FDA submission (K152217); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Recalls naming this device

  1. Aug 30, 2021Open, Classified

    Inserts were packaged in vacuum bags that lacked an additional oxygen barrier layer.

    Z-0024-2022 · Root cause: Process control · Recall details · FDA record ↗

Most reported problems

Device problems

  • Naturally Worn1

Patient problems

  • Cyst(s)1
  • Failure of Implant1
  • Implant Pain1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • InjuryJun 7, 2023Naturally Worn; Naturally Worn

    AS REPORTED VIA LEGAL DOCUMENTATION A 54 Y/O FEMALE PATIENT HAD A LEFT ANKLE REPLACEMENT ON (B)(6) 2019. APPROXIMATELY 3 YEARS AND 2 MONTHS AFTER THE INITIAL PROCEDURE THE PATIENT HAD A LEFT ANKLE REVISION ON (B)(6) 2023. CLINICAL EXAM AND RADIOGRAPHS DEMONSTRATED FAILURE OF TIBIAL COMPONENT WITH SUBSIDENCE POSTERIORLY. THERE IS NO OTHER PATIENT DEMOGRAPHIC OR MEDICAL HISTORY AVAILABLE. THERE IS NO DEVICE RETURN. THERE ARE NO PHOTOS OR OTHER IMAGES OF THE DEVICE PROVIDED. NO ADDITIONAL INFORMATION IS AVAILABLE. REVISION OP REPORT ON (B)(6) 2023. DIAGNOSIS: FAILED LEFT TOTAL ANKLE REPLACEMENT. CAPSULE FROM THE ANTERIOR ASPECT OF THE ANKLE WAS SENT TO PATHOLOGY AS WELL AS MICROBIOLOGY. THERE WERE NO COMPLICATIONS. THE PATIENT WAS STABLE UPON TRANSFER TO THE PACU.

    Patient problems: Cyst(s); Failure of Implant; Implant Pain

    MAUDE report 17083607 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K152217. One submission often covers many models, so check the product and lot details in each record.

  • Z-0587-2024 · Nov 21, 2023 · Open, Classified
    Exactech Vantage Ankle System, Catalog Numbers: a) 350-21-04, b) 350-22-03
  • Z-2926-2020 · Aug 10, 2020 · Open, Classified
    The Vantage Total Ankle Tibial Insert, Fixed Bearing, Size 4, Right 10mm.
  • Z-2966-2018 · Jul 13, 2018 · Terminated
    Tibial Alignment Guide, REF 351-10-00 Vantage Fixed Ankle system components for ankle replacements.
  • Z-0200-2018 · May 16, 2017 · Terminated
    Vantage Total Ankle Tibial Insert, Fixed Bearing, 7 mm, Left, Size 3. Exactech Inc., Intended for patients with ankle joints damage by severe rheumatoid, post-traumatic, or dege
  • Z-0201-2018 · May 16, 2017 · Terminated
    Vantage Total Ankle Tibial Insert, Fixed Bearing, 10 mm, Left, Size 3. Exactech Inc. Intended for patients with ankle joints damage by severe rheumatoid, post-traumatic, or degene

Other Prosthesis, Ankle, Semi-Constrained, Cemented, Metal/Polymer devices

All Prosthesis, Ankle, Semi-Constrained, Cemented, Metal/Polymer devices →