Vantage recallZ-0024-2022 · Exactech, Inc. · initiated Aug 30, 2021
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Exactech, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Inserts were packaged in vacuum bags that lacked an additional oxygen barrier layer.
FDA root cause category: Process control
What the company is doing
Exactech notified distributors and sales representatives on about 08/30/2021 via letter titled "URGENT MEDICAL DEVICE RECALL." Actions included removing all Knee and Ankle UHMWPE products labeled with an 8-year shelf life and not packaged in EVOH/Nylon bags, in a phased approach over 12 months. Phase 1: immediately return all knee and ankle UHMWPE devices labeled with an 8-year shelf life that will be 5 years old or older by 08/31/2022 not packaged in EVOH/Nylon bags. Phase 2: between 05/31/2022 to 08/31/2022, returning all remaining knee and ankle UHMWPE devices labeled with an 8-year shelf life not packaged in EVOH/Nylon bags. A communication to healthcare professionals should follow. Instructions to recipients included to immediately identify and quarantine any affected product in inventory, notify customers down the line if further distributed, and complete and return the Recall Inventory Response Form to Exactech via email at [email protected]. Exactech expanded their recall and sent an updated notification to US agents and distributors on 02/07/2022. On about 02/08/2022, a letter to surgeons and hospitals was sent. The notifications indicated that all non-conforming devices were included in the recall, to cease use of those devices, and work with sales agents to return any inventory. The letters to surgeons and hospitals also advised physicians regarding follow up with implanted patients.
Recall details
- Recall number
- Z-0024-2022
- Status
- Open, Classified
- Initiated by the company
- Aug 30, 2021
- Posted by the FDA
- Oct 4, 2021
- Recalling firm
- Exactech, Inc.
- Quantity
- 2,217 devices
- Product code
- HSN
- FDA submissions
- K152217
- Distribution
- Worldwide distribution - US Nationwide distribution and the countries of ALGERIA, ARGENTINA, AUSTRALIA, AUSTRIA, BELGIUM, BRAZIL, CANADA, CHILE, CHINA, COLOMBIA, CYPRUS, ECUADOR, FRANCE, FRENCH POLYNESIA, GERMANY, GHANA, GREAT BRITAIN, GREECE, GUATEMALA, INDIA, ITALY, JAPAN, JORDAN, KOREA, REPUBLIC OF, LEBANON, LUXEMBOURG, MALAYSIA, MEXICO, NETHERLANDS, NEW ZEALAND, POLAND, PORTUGAL, ROMANIA, SINGAPORE, SPAIN, SWITZERLAND, SYRIAN ARAB REPUBLIC, THAILAND, TUNISIA, TURKEY, VENEZUELA .
Product description
Vantage Total Ankle System including a. Vantage Fixed-Bearing Polyethylene Liner Component (Product Line: 350-21-XX, 350-22-XX); b. Vantage Mobile-Bearing Polyethylene Liner Component (Product Line: 350-41-XX, 350-42-XX). NOTE: Information reflects Exactech recall expansion initiated on 02/07/2022.
Code information (models, lots, serial numbers)
Show the full code information (23k characters)
a. Item Number (Device Identifier; Serial number(s)): 350-21-01 (10885862276216; 4833710, 4833714, 4833715, 4833717, 4833723, 5188005, 5188007, 5188008, 5188009, 5188011, 5188013, 5188015, 5188017, 5188021, 5188022, 5188023, 5939820, 5939826, 5939831, 5939832, 5939833, 5939835, 5939836, 5939837, 5939838, 5939839, 5939840, 5939841, 5939842, 5939844, 5939845, 5939846, 6015553, 6015554, 6015556, 6015557, 6015560, 6015563, 6015566, 6015567, 6015569, 6396486, 6396489, 6396493, 6396495, 6396500, 6396504, 6396505, 6489379, 6489381, 6489382, 6489384, 6489386, 6606659, 6606667, 6606678, 6860126, 6860127, 6900582, 6900585, 6900588, Z060698, Z060703, Z060715, Z064072, Z064105, Z064107); 350-21-02 (10885862276223; 4536664, 4536665, 4536666, 4536668, 4536669, 4536670, 4536672, 4842955, 4842956, 4842957, 4842958, 4842959, 5184645, 5184647, 5184648, 5184649, 5184651, 5184652, 5184653, 5184654, 5184656, 5184657, 5184658, 5855814, 5855815, 5855816, 5855819, 5855820, 5855823, 5855824, 5855825, 5855826, 5855827, 5855828, 5855829, 5855830, 6240459, 6240461, 6240463, 6240464, 6240467, 6259881, 6259882, 6259884, 6259885, 6259888, 6356551, 6356552, 6356561, 6356562, 6356564, 6356566, 6356568, 6356569, 6356570, 6395847, 6395848, 6395852, 6395854, 6395855, 6395856, 6395858, 6395860, 6489055, 6489056, 6489059, 6489063, 6489065, 6489066, 6489069, 6489070, 6489071, 6489072, 6489073, 6583838, 6583839, 6583842, 6583843, 6583844, 6583845, 6583846, 6583849, 6583852, 6583853, 6583855, 6583857, 6730685, 6730688, 6730690, 6730696, 6730697, 6899729, 6899744, 6899751, Z068705, Z075029, Z075042); 350-21-03 (10885862276230; 4536738, 4536739, 4536740, 4536741, 4536742, 4536743, 4536745, 4536746, 4841023, 4841024, 4841025, 5184662, 5184663, 5184665, 5184666, 5184668, 5184669, 5184671, 5184672, 5184673, 5184674, 5184675, 5184678, 5184679, 5184680, 5221264, 5221265, 5221266, 5221267, 5221268, 5221269, 5221270, 5221272, 5221273, 5221274, 5221275, 5221276, 5940499, 5940500, 5940502, 5940507, 5940511, 5940516, 5940517, 5940518, 5940519, 5940520, 6039207, 6039208, 6039210, 6039212, 6039213, 6039214, 6039215, 6039218, 6039220, 6039221, 6039222, 6039223, 6039225, 6173447, 6173449, 6173450, 6173451, 6173452, 6173455, 6173456, 6173458, 6173459, 6173460, 6173462, 6173463, 6173464, 6173465, 6372717, 6372718, 6372719, 6372720, 6372721, 6372726, 6372727, 6372729, 6372730, 6372735, 6489128, 6489129, 6489131, 6489133, 6489134, 6489135, 6489136, 6489137, 6489145, 6489146, 6535721, 6535723, 6535724, 6535729, 6535730, 6535731, 6535732, 6535734, 6742242, 6742248, 6742249, 6742250, 6742251, 6742253, 6742255, 6742259, 6811600, 6811611, 6811612, 6900322, Z074962, Z074972, Z074984, Z075184); 350-21-04 (10885862276247; 4537204, 4537205, 4537206, 4537207, 4537212, 4750991, 5184539, 5184541, 5184543, 5184548, 5184550, 5184553, 5184554, 5184557, 5240718, 5240721, 5240727, 5240728, 5909332, 5909335, 5909337, 5909338, 5909344, 5909350, 5954808, 5954810, 5954815, 5954820, 6356463, 6356464, 6356466, 6356471, 6356472, 6356473, 6356477, 6489076, 6489081, 6489083, 6489093, 6758223, 6758226, 6758227, 6856945, 6856949, 6995782, Z068655); 350-21-11 (10885862276254; 4536710, 4536711, 4536714, 4536715, 4536716, 4536717, 4536718, 4704930, 4704938, 4704939, 4704940, 4704944, 4704945, 4704949, 4704951, 4704956, 4704957, 4787630, 4787631, 5184636, 5184639, 5184640, 5184643, 5231854, 5231855, 5231858, 5231860, 5969127, 5969128, 5969129, 5969132, 6356520, 6356523, 6489037, 6489045, 6489052); 350-21-12 (10885862276261; 4704789, 4771382, 4771383, 4771384, 4771387, 4771388, 5184684, 5184689, 5184690, 5184691, 5184697, 5184698, 5237420, 5237421, 5237422, 5237423, 5237424, 5237426, 5237430, 5237431, 5237432, 5939798, 5939801, 5939802, 5939803, 5939807, 5939810, 6356669, 6356670, 6356671, 6356672, 6356677, 6356680, 6356681, 6356683, 6356684, 6489360, 6489363, 6489365, 6489366, 6489369, 6489370, 6489372, 6598979, 6598980, 6598983, 6598985, 6843245, 6843247, 6843261, 6883796, Z058384, Z074986, Z074999); 350-21-13 (10885862276278; 4704790, 4771176, 4977232, 4977233, 4977235, 5181919, 5181921, 5181923, 5181926, 5181927, 5181931, 5181932, 5181934, 5222348, 5222349, 5222352, 5222355, 5222356, 5222357, 5222358, 5908535, 5908536, 5908537, 5908540, 5908541, 5908542, 5908544, 5908548, 5908549, 6061322, 6061327, 6323250, 6323261, 6395423, 6395424, 6395426, 6395428, 6488292, 6488297, 6488301, Z075109, Z075175); 350-21-14 (10885862276285; 4746009, 5184848, 5184851, 5184852, 5184853, 5184858, 5184860, 5184864, 5184865, 6143314, 6143323, 6143326, 6143329, 6143331, 6395394, 6395395, 6719894, Z076042); 350-21-21 (10885862276292; 4676330, 4676331, 4676332, 4676333, 4676334, 4676335, 4676336, 4676338, 4676345, 4676347, 4676351, 4676355, 5185107, 5185108, 5185110, 5185111, 5185113, 5185114, 5185117, 5185118, 5185119, 5245298, 5906409, 5906414, 5938419, 5938422, 5938424, 5938428, 6356532, 6356533, 6356537, 6356542, 6356549, 6486889, 6486890, 6486896, 6804245, 6900926); 350-21-22 (10885862276308; 4750965, 4771177, 5184866, 5184867, 5184870, 5184872, 5184873, 5184874, 5184878, 5184879, 5184880, 5184881, 5184883, 5876458, 5876461, 5876467, 6142807, 6142808, 6142811, 6142819, 6142825, 6411781, 6411784, 6428914, 6428917, 6428918, 6435219, 6435225, 6621519, 6621521, 6621524, 6621536, 6883664, 6883667, 6883671, 6959803, Z059779); 350-21-23 (10885862276315; 4771244, 4866126, 4866127, 4866129, 5172559, 5172560, 5172565, 5172566, 5172567, 5172569, 5172570, 5172572, 5172574, 5172577, 5172580, 5172581, 5172582, 5172583, 5172584, 5172588, 5172589, 5989847, 5989848, 5989849, 5989850, 5989853, 5989859, 5989861, 6408952, 6408954, 6408958, 6408959, 6408960, 6487891, 6487893, 6487895, 6487896, 6487897, 6487900, 6487902, 6487908, 6487909, 6644822, 6644825, 6644826, 6644831, 6644834, 6644835, 6644838, Z058458, Z058462, Z060686); 350-21-24 (10885862276322; 4536720, 4536721, 4536723, 4536724, 5185120, 5185127, 5185129, 5185132, 5185133, 5185134, 5185135, 5245294, 5245296, 5882353, 5882355, 5882358, 5882364, 5882365, 6142592, 6142601, 6142602, 6142605, 6142607, 6142608, 6396151, 6396156, 6851740, Z058435); 350-21-41 (10885862293213; 5181205, 5181206, 5181209, 5181210, 5181211, 5181212, 5181214, 5181215, 5181217, 5181218, 5181219, 5181222, 5954729, 5954730, 5954736, 5954745, 6409727, 6747749, 6747752, 6897935); 350-21-42 (10885862293220; 5184830, 5184831, 5184833, 5184837, 5184838, 5184839, 5184840, 5184844, 5184845, 5201342, 5201345, 5201349, 5909253, 5940479, 5940481, 5940482, 5940484, 5940487, 5940493, 5940495, 6143274, 6143276, 6143277, 6143282, 6143285, 6143286, 6143289, 6143292, 6396172, 6540417, 6540421, 6911569, Z059753); 350-21-43 (10885862293237; 5102165, 5102166, 5102171, 5102173, 5102174, 5102178, 5102179, 5102180, 5102183, 5230186, 5230190, 5230192, 5230196, 5230197, 5230199, 5230200, 5230203, 5912834, 5912837, 5912838, 5912840, 5912844, 5912846, 5912848, 6015723, 6015725, 6015728, 6395893, 6395900, 6571764, 6571775, 6860987, 6913035, Z077359); 350-21-44 (10885862293244; 5181165, 5181177, 5181178, 5967776, 5967781, 6325734, 6398023, 6424762, 6709425); 350-21-61 (10885862293374; 5185137, 5185147, 5909303); 350-21-62 (10885862293381; 6143359, 6143360, 6143368, 6143370, Z060582); 350-21-63 (10885862293398; 4733268, 5185157, 6487566); 350-21-64 (10885862293404; 5185168, 6143340, 6143348, 6488098); 350-22-01 (10885862276339; 4681920, 4681922, 4681923, 4681925, 4681927, 4681928, 4681930, 4681931, 4681932, 4681933, 4681934, 4681935, 4681937, 4681938, 4681940, 4681943, 4681945, 4681946, 4681948, 4681949, 5139951, 5139952, 5139953, 5139955, 5139956, 5139962, 5139963, 5139968, 5139970, 5139971, 5876391, 5876392, 5876393, 5876396, 5876398, 5876399, 5876403, 5876407, 5876412, 6127984, 6127985, 6127987, 6127993, 6127995, 6127997, 6128000, 6128003, 6395402, 6395406, 6395407, 6395410, 6395417, 6487092, 6487093, 6487094, 6487095, 6487097, 6487100, 6487102, 6487103, 6487106, 6487107, 6487108, 6487110, 6720137, 6720139, 6720142, 6720145, 6720146, 6720150, 6720152, 6720153, 6907877, 6907886, 6907887, 6907888, 6907892, 6907896, 6907899, 6907901, 6911587, 6911594, Z073322, Z073329); 350-22-02 (10885862276346, 4771234, 4771235, 4771236, 4771237, 4771238, 4771239, 4771240, 5181224, 5181225, 5181228, 5181229, 5181234, 5181235, 5181236, 5181237, 5181238, 5181239, 5181240, 5181241, 5181242, 5181244, 5245289, 5245290, 5245291, 5245292, 5245293, 5985462, 5985463, 5985464, 5985466, 5985468, 5985470, 5985471, 5985472, 5985473, 5985474, 5985475, 5985476, 5985477, 5985478, 5985480, 5985481, 6048026, 6048027, 6048028, 6048030, 6048031, 6067781, 6067782, 6067783, 6067785, 6067786, 6067787, 6067788, 6067790, 6067792, 6259896, 6259897, 6308445, 6308446, 6308447, 6308448, 6308451, 6308452, 6308454, 6308458, 6411977, 6411978, 6411979, 6411980, 6411981, 6411983, 6411984, 6411985, 6411989, 6434808, 6434811, 6434812, 6434813, 6434814, 6434816, 6434817, 6434818, 6434820, 6434821, 6434822, 6434823, 6434824, 6532890, 6532891, 6532895, 6532898, 6532899, 6532900, 6532902, 6532903, 6532904, 6532907, 6535639, 6535641, 6535643, 6535644, 6535645, 6535646, 6535649, 6535650, 6535651, 6535654, 6535657, 6758072, 6758073, 6758074, 6899905, 6899911, 6899914, 6899919, Z071104, Z071116, Z071136, Z076051); 350-22-03 (10885862276353; 4884296, 4884297, 4884298, 4884299, 5185183, 5185184, 5185187, 5185189, 5185190, 5185191, 5185192, 5185194, 5185195, 5185196, 5256071, 5256072, 5256073, 5256074, 5256075, 5256077, 5256078, 5256079, 5256080, 5938438, 5938439, 5938440, 5938441, 5938442, 5938447, 5938448, 5938451, 5938452, 5938453, 5938454, 5938456, 5938457, 5988014, 5988015, 5988016, 5988018, 5988019, 5988020, 5988022, 5988023, 5988025, 5988026, 5988027, 5988029, 5988030, 6252633, 6252634, 6252636, 6252638, 6252640, 6252641, 6285206, 6285207, 6285208, 6285210, 6285211, 6285212, 6285213, 6285215, 6340621, 6340630, 6340631, 6340632, 6340635, 6395438, 6395439, 6395440, 6395441, 6395444, 6395446, 6395451, 6395456, 6395457, 6482649, 6482650, 6482656, 6482662, 6482664, 6482665, 6509210, 6509214, 6509215, 6509216, 6509218, 6509219, 6509220, 6509223, 6509224, 6509227, 6683471, 6683474, 6683475, 6683477, 6683482, 6683485, 6683488, 6683492, 6718699, 6718707, 6718715, 6718716, 6852366, 6852367, 6852368, 6852375, 6852377, 6852379, 6853420, 6853427, 7010354, Z072696, Z073295); 350-22-04 (10885862276360; 5181310, 5181313, 5181314, 5181315, 5181317, 5181323, 5181327, 5262727, 5262729, 5262730, 5262731, 5262733, 5262735, 5262736, 5262737, 5907662, 5907666, 5907670, 5907671, 5907672, 5907673, 5907674, 5907676, 5967477, 5967482, 5967487, 5967488, 5967489, 5967491, 5967493, 5967494, 5967495, 6337400, 6337402, 6337406, 6337407, 6482499, 6502574, 6502576, 6502580, 6502586, 6502589, 6846082, 6846088, Z058442, Z058443, Z075357); 350-22-11 (10885862276377; 4703168, 4703171, 4703174, 4703177, 4703178, 4703179, 4703180, 4703183, 4703184, 4703187, 4703188, 4703190, 4703191, 4703192, 4703193, 4703194, 4703195, 4703196, 4703197, 4857844, 4857845, 4857846, 4857847, 5181328, 5181331, 5181336, 5181337, 5181338, 5181339, 5181341, 5181343, 5181345, 5954756, 5954758, 5954759, 5954761, 5954765, 6137105, 6137106, 6137115, 6137119, 6137121, 6372459, 6372468, 6372471, 6372474, 6372475, 6487131, 6487137, 6898086, Z075079); 350-22-12 (10885862276384; 4704788, 4771243, 5181347, 5181348, 5181349, 5181351, 5181352, 5181353, 5181354, 5181355, 5181356, 5181359, 5181360, 5181363, 5181364, 5193990, 5193991, 5193993, 5193995, 5193996, 5193997, 5193998, 5193999, 5194001, 5194002, 5194003, 5194004, 5882202, 5882208, 5882209, 5882210, 5882211, 5882212, 5882214, 5882215, 5882216, 5882218, 5882219, 5882282, 5882283, 5882284, 5882285, 5882286, 5882292, 5882295, 6356685, 6356691, 6356694, 6356695, 6356703, 6428946, 6489020, 6489021, 6489025, 6489026, 6489029, 6489031, 6502624, 6502629, 6502633, 6502634, 6683224, 6683231, 6898077, 6970223, 6970236, Z075900); 350-22-13 (10885862276391; 4771241, 4771242, 5120877, 5120882, 5120885, 5120886, 5120887, 5120889, 5204239, 5204240, 5204242, 5204244, 5204245, 5204247, 5204248, 5204250, 5204251, 5204252, 5939735, 5939741, 5939742, 5939744, 5939745, 5939746, 5939748, 5939749, 5939750, 5939751, 5939752, 5939753, 5939754, 6136641, 6136646, 6136649, 6136650, 6136651, 6428891, 6428893, 6489094, 6489097, 6489109, 6489124, 6489125, 6856921, 6856930, 6856931, Z058368, Z058370, Z074916); 350-22-14 (10885862276407; 5181368, 5181373, 5181374, 5181376, 5181380, 5181381, 5181383, 5906479, 5967548, 5967552, 5967553, 5967560, 5967564, 5967567, 6174579, 6174584, 6174589, 6569281, 6569289); 350-22-21 (10885862276414; 4703140, 4703141, 4703142, 4703143, 4703144, 4703147, 4703148, 4703149, 4703152, 4703154, 4703155, 4703156, 4703158, 4703160, 4703162, 4703163, 4703165, 4703167, 5185215, 5185216, 5185219, 5185221, 5185223, 5185225, 5185228, 5214261, 5214264, 5214265, 5939780, 5939788, 5939789, 6259901, 6395459, 6395470, 6591152, 6591161, 6591163); 350-22-22 (10885862276421; 4748658, 4771389, 5120893, 5120894, 5120895, 5120897, 5120898, 5120899, 5120900, 5120901, 5120902, 5120906, 5120907, 5120909, 5120912, 5120913, 5120914, 5120915, 5241242, 5241243, 5241244, 5241245, 5241246, 5241247, 5241248, 5241249, 5241250, 5241251, 5241252, 6015770, 6015777, 6015788, 6395920, 6487170, 6487175, 6487177, 6487179, 6487180, 6487181, 6742061, 6742065, 6742069, 6911514, Z075188, Z075281); 350-22-23 (10885862276438; 4771390, 4771392, 4771393, 4771394, 5181291, 5181294, 5181295, 5181296, 5181297, 5181299, 5181300, 5181301, 5181302, 5181303, 5181304, 5181305, 5181308, 5880510, 5880511, 5880514, 5880525, 5938089, 5938090, 5938091, 5938093, 5938096, 5938097, 5938099, 5938103, 5938105, 5938106, 5938107, 6395828, 6395831, 6395833, 6395837, 6395843, 6411724, 6411725, 6411726, 6411729, 6411730, 6411731, 6411735, 6411738, 6411741, 6506227, 6506228, 6506231, 6506234, 6506235, 6506239, 6506242, 6506243, 6734441, 6734442, 6734444, 6734447, 6734454, 6734455, 6870162, 6870183, 6870188, Z075253); 350-22-24 (10885862276445; 5181451, 5181453, 5181455, 5181458, 5181459, 5181464, 5232860, 5232861, 5232864, 5232867, 5232869, 5232870, 5232871, 5232874, 5232876, 5232877, 5907747, 5907751, 5939721, 5939727, 5939732, 6395493, 6395506, 6487126, 6487130, 6822884, 6822896, Z060416, Z069324); 350-22-41 (10885862293251; 5187946, 5187954, 5187956, 5907658, 5954768, 5954774, 5954779, 6396518, 6396521, 6489006); 350-22-42 (10885862293268; 5185783, 5185787, 5185788, 5185789, 5185790, 5185793, 5185794, 5185795, 5185799, 5185801, 5201264, 5201265, 5201266, 5201272, 5201273, 5201275, 5201276, 5201278, 5938401, 5938402, 5938404, 5938409, 5938411, 5938416, 6391264, 6483678, 6483684, 6483687, 6520721, 6520724, 6520725, 6520732, 6520733, 6520737, 6738668, 6738674, 6898015, 6898023, Z070796, Z070802); 350-22-43 (10885862293275; 5122629, 5122632, 5122635, 5122637, 5122638, 5122643, 5122645, 5122646, 5122648, 5260636, 5260638, 5260642, 5260643, 5260645, 5883315, 5883316, 5883324, 5940026, 5940027, 5940029, 5940033, 5940037, 5940038, 5940039, 5940041, 5940042, 6273558, 6273559, 6273562, 6273568, 6273570, 6412019, 6579440, 6579442, 6579443, 6579448, 6886404, Z060757); 350-22-44 (10885862293282; 5185203, 5185206, 5185209, 5967508, 5967510, 6417987, 6683339, 6737449); 350-22-62 (10885862293428; 5181465, 5181473, 5181484, 5948972, 5948990); 350-22-63 (10885862293435; 5181395, 5906439, 5906440, 6409703, 6409711, 6409712, Z077335); 350-22-64 (10885862293442; 5181406, 5181407, 5181414, 5181421, 5975072, 5975080); b. Item Number (Device Identifier; Serial number(s)): 350-41-01 (10885862276452; 4851797, 4851798, 4851799, 4851800, 4964108, 4986270, 5194348, 5194350, 5194353, 5194356, 5194357, 5194358, 6409670, 6506001, 6506009, 6857644, 6857646, 6857652); 350-41-02 (10885862276469; 4850102, 4850103, 4850104, 4850105, 4963948, 4963951, 5194275, 5194276, 5194277, 5194279, 5194280, 5194281, 5194282, 5194283, 5194285, 5194286, 5194287, 5194288, 5194289, 5194290, 5194291, 5194292, 5865068, 5865072, 5865073, 5865076, 5865081, 5865086, 5865087, 5865089, 6330721, 6330728, 6330732, 6507180, 6507186, 6507198, 6592234, 6592235); 350-41-03 (10885862276476; 4851814, 4851816, 4963859, 4963860, 4963862, 5181425, 5181426, 5181428, 5181430, 5181432, 5181433, 5181434, 5181435, 5181436, 5181438, 5181440, 5181441, 5181443, 5181444, 6138428, 6409639, 6409640, 6409644, 6409647, 6409650, 6409651, 6409652, 6409658, 6535259, 6535265, 6535268, 6900013); 350-41-04 (10885862276483; 4851782, 4851783, 4851784, 4913716, 5193807, 5193811, 5193813, 5193819, 6143103, 6143110, 6506247, 6506251, 6506260, 6506262, 6506263); 350-41-05 (10885862302038; 5192099); 350-41-11 (10885862276490; 4851670, 4851671, 4851672, 4963858, 5194262, 5194265, 5194268, 5194274, 6506369, 6506380); 350-41-12 (10885862276506; 4850106, 4850107, 4850108, 4850109, 4913690, 4913691, 4913693, 5194293, 5194294, 5194297, 5194304, 5194307, 6505962, 6505963, 6505968, 6566991, 6566994, 6566999, 6567000); 350-41-13 (10885862276513; 4850078, 4850079, 4850080, 4850081, 4963842, 5194329, 5194331, 5194333, 5194334, 5194336, 5194338, 5194339, 5194342, 5194345, 6141325, 6141326, 6141329, 6439029, 6439031, 6439041, 6825226); 350-41-14 (10885862276520; 4851603, 4913723, 4913724, 4913725, 5193794, 5193795, 5193798, 5193806, 6143269, 6776618, 6776626); 350-41-15 (10885862302045; 5194254); 350-41-21 (10885862276537; 4850098, 4850099, 4850100, 4963849, 4963850, 4963852, 5205336, 5205342, 5205346, 5205347, 6506333, 6506347); 350-41-22 (10885862276544; 4851637, 4851638, 4851639, 4851640, 4930455, 4930456, 4930458, 4930460, 5250981, 5250983, 5250985, 5250986, 5250987, 5250988, 6143140, 6507141, 6922888, Z081039); 350-41-23 (10885862276551; 4850110, 4850111, 4850112, 4913660, 4913661, 4913662, 4913663, 5194236, 5194237, 5194239, 5194240, 5194246, 5194247, 5194248, 5194250, 5194251, 5194252, 6141292, 6141293, 6466622, 6504528, 6504529, 6775088, 6775098, 6775102); 350-41-24 (10885862276568; 4850094, 4850095, 4850096, 4850097, 4963769, 4963772, 5184563, 5184566, 5184568, 5184571, 5184576, 6141203, 6505950, 6505958); 350-41-25 (10885862302052; 5194176, 5194177); 350-41-41 (10885862293534; 4851727, 4963764, 4963765, 5194226, 5194231); 350-41-42 (10885862293541; 4850090, 4850091, 4850092, 4850093, 4963759); 350-41-43 (10885862293558; 4851761, 4851763, 4963777, 4963778, 4963780, 5194201, 5194203, 5194206, 5194212, 5194215, 5194216, 6143085, 6506501); 350-41-44 (10885862293565; 4851693, 4963661, 4963662, 5193851, 5193852, 6139778, 6139786); 350-41-45 (10885862302076; 5194180); 350-41-61 (10885862293695; 4851597, 4851598); 350-41-62 (10885862293701; 4851588, 4963649, 5194100); 350-41-63 (10885862293718; 4851593, 4851594); 350-41-64 (10885862293725; 4851599, 4851600, 5194096); 350-41-65 (10885862302090, 5194116, 5194118); 350-42-01 (10885862276575; 4851765, 4851766, 4851767, 4851768, 4962062, 4962065, 4962067, 5181941, 5181942, 5181945, 5181948, 5181949, 5181951, 5181955, 5181956, 5181957, 6504541, 6504544, 6504545, 6504547, 6504548, 6504550, 6813572, 6813577, 7009630); 350-42-02 (10885862276582; 4847130, 4847131, 4847132, 4962961, 4962964, 4962965, 5194138, 5194139, 5194140, 5194141, 5194143, 5194145, 5194147, 5194149, 5194150, 5194152, 5194153, 5194155, 5910497, 5910507, 6442342, 6442348, 6442351, 6506023, 6506024, 6506029, 7011500); 350-42-03 (10885862276599; 4847207, 4847208, 4847209, 4847210, 4863354, 4863355, 4863356, 4863357, 4863358, 4863359, 4863360, 5229805, 5229806, 5229807, 5229809, 5229812, 5229813, 5229815, 5229817, 5229819, 5229820, 5229821, 5229822, 6070811, 6070812, 6356277, 6356281, 6356282, 6356284, 6356286, 6356288, 6356289, 6356292, 6356293, 6356295, 6570281, 6570284, 6570291, 6570292, 6570297, 6862327, 6862328, 6862329, 6862332, 6862333); 350-42-04 (10885862276605; 4847172, 4847173, 4847174, 4914356, 4914359, 4959353, 4959356, 4959357, 5245932, 5245934, 5245935, 5245936, 5245938, 5245939, 5245942, 5245944, 6095443, 6095450, 6095453, 6095455, 6505375, 6505376, 6505383, 6505384, 6505385, 6505390, 6505392); 350-42-05 (10885862302106; 5192871); 350-42-11 (10885862276612; 4851829, 4851830, 4851831, 4851832, 4962030, 5192065, 5192067, 5192070, 5192071, 5192075, 5192077, 6143134, 6506128, 6506130, 6981225); 350-42-12 (10885862276629; 4847136, 4847137, 4914322, 4914323, 4914325, 4959360, 4959361, 4959363, 4959364, 5234643, 5234651, 5234652, 5234654, 6138480, 6138483, 6138489, 6505702, 6505703, 6505708, 6505709, 6505711, 6861040, 6861055); 350-42-13 (10885862276636; 4851580, 4851581, 4851582, 4959342, 4959346, 5181960, 5181961, 5181963, 5181964, 5181965, 5181966, 5181967, 5181970, 5181971, 5181972, 5181974, 5181975, 5181976, 5181977, 6143249, 6143253, 6456685, 6456692, 6456695, 6456703, 6591960, 6591961, 6591962, 6982233, 6982235, 6982236, Z081206, Z081219, Z081222); 350-42-14 (10885862276643; 4851584, 4851586, 4914362, 4914363, 4914364, 4914365, 4959348, 4959350, 5247865, 5247867, 5247870, 5247876, 6095825, 6095826, 6095835, 6458504, 6504558, 6504563, 6795422, 6795424); 350-42-15 (10885862302113; 5192877, 5192879); 350-42-21 (10885862276650; 4851469, 4851470, 4851471, 4851472, 4959335, 5192816, 5192820, 5192822, 6143099, 6143102, 6505394, 6505399, 6505404, 6505407, 6505409, 6505410); 350-42-22 (10885862276667; 4851477, 4851478, 4851479, 4851480, 4963995, 5192850, 5192851, 5192852, 5192853, 5192854, 5192855, 5192856, 5192857, 5192859, 5192862, 5192864, 5192865, 5192866, 6013085, 6013087, 6013088, 6504632, 6504637, 6504640, 6504641, 6900864); 350-42-23 (10885862276674; 4850086, 4850087, 4850088, 4850089, 4915608, 4915609, 4915610, 4959366, 4959367, 4959368, 4959369, 4959370, 5247851, 5247852, 5247853, 5247855, 5247857, 5247861, 6140620, 6140624, 6140626, 6505022, 6505024, 6505035, 6505037); 350-42-24 (10885862276681; 4851474, 4851476, 4964102, 4964105, 4964106, 5192835, 5192837, 5192841, 5192842, 6505078, 6505084, 6505086, 6505090, 6505093, 6505094, 6505095); 350-42-25 (10885862302120; 5192873); 350-42-41 (10885862293572; 4851484, 4964051, 4964054, 4964055, 6140502, 6505414, 6505426, 6505429); 350-42-42 (10885862293589; 4851486, 4851487, 4851488, 4964095, 4964096, 4964097, 5192764, 5192765, 5192771, 5192777, 6143120, 6566946, 6566948); 350-42-43 (10885862293596; 4851501, 4851502, 4851503, 4851504, 5192747); 350-42-44 (10885862293602; 4850481, 4850482, 4850484, 4964061, 5192734); 350-42-45 (10885862302144; 5192492); 350-42-61 (10885862293732; 4850491, 4964127, 4964130); 350-42-62 (10885862293749; 4850477, 4850479, 5193940); 350-42-63 (10885862293756; 4850485, 4850486, 4850488, 5194360, 6138511); 350-42-64 (10885862293763; 4964131, 4964133, 5194389, 5194394); 350-42-65 (10885862302168, 5192508);
Devices named in this recall
39 devices in the FDA device registry whose identifier appears in the recall's code information.
- Vantage 350-22-03Exactech, Inc. · Prosthesis, Ankle, Semi-Constrained, Cemented, Metal/Polymer2 recalls2 reports
- Vantage 350-21-04Exactech, Inc. · Prosthesis, Ankle, Semi-Constrained, Cemented, Metal/Polymer2 recalls0 reports
- Vantage 350-22-02Exactech, Inc. · Prosthesis, Ankle, Semi-Constrained, Cemented, Metal/Polymer1 recall5 reports
- Vantage 350-21-02Exactech, Inc. · Prosthesis, Ankle, Semi-Constrained, Cemented, Metal/Polymer1 recall4 reports
- Vantage 350-22-01Exactech, Inc. · Prosthesis, Ankle, Semi-Constrained, Cemented, Metal/Polymer1 recall3 reports
- Vantage 350-22-11Exactech, Inc. · Prosthesis, Ankle, Semi-Constrained, Cemented, Metal/Polymer1 recall3 reports
- Vantage 350-21-01Exactech, Inc. · Prosthesis, Ankle, Semi-Constrained, Cemented, Metal/Polymer1 recall2 reports
- Vantage 350-21-03Exactech, Inc. · Prosthesis, Ankle, Semi-Constrained, Cemented, Metal/Polymer1 recall2 reports
- Vantage 350-21-13Exactech, Inc. · Prosthesis, Ankle, Semi-Constrained, Cemented, Metal/Polymer1 recall2 reports
- Vantage 350-21-22Exactech, Inc. · Prosthesis, Ankle, Semi-Constrained, Cemented, Metal/Polymer1 recall2 reports
- Vantage 350-21-23Exactech, Inc. · Prosthesis, Ankle, Semi-Constrained, Cemented, Metal/Polymer1 recall2 reports
- Vantage 350-22-12Exactech, Inc. · Prosthesis, Ankle, Semi-Constrained, Cemented, Metal/Polymer1 recall2 reports
- Vantage 350-22-23Exactech, Inc. · Prosthesis, Ankle, Semi-Constrained, Cemented, Metal/Polymer1 recall2 reports
- Vantage 350-22-43Exactech, Inc. · Prosthesis, Ankle, Semi-Constrained, Cemented, Metal/Polymer1 recall2 reports
- Vantage 350-22-44Exactech, Inc. · Prosthesis, Ankle, Semi-Constrained, Cemented, Metal/Polymer1 recall2 reports
- Vantage 350-21-11Exactech, Inc. · Prosthesis, Ankle, Semi-Constrained, Cemented, Metal/Polymer1 recall1 report
- Vantage 350-21-14Exactech, Inc. · Prosthesis, Ankle, Semi-Constrained, Cemented, Metal/Polymer1 recall1 report
- Vantage 350-22-64Exactech, Inc. · Prosthesis, Ankle, Semi-Constrained, Cemented, Metal/Polymer1 recall1 report
- Vantage 350-21-12Exactech, Inc. · Prosthesis, Ankle, Semi-Constrained, Cemented, Metal/Polymer1 recall0 reports
- Vantage 350-21-21Exactech, Inc. · Prosthesis, Ankle, Semi-Constrained, Cemented, Metal/Polymer1 recall0 reports
- Vantage 350-21-24Exactech, Inc. · Prosthesis, Ankle, Semi-Constrained, Cemented, Metal/Polymer1 recall0 reports
- Vantage 350-22-04Exactech, Inc. · Prosthesis, Ankle, Semi-Constrained, Cemented, Metal/Polymer1 recall0 reports
- Vantage 350-22-13Exactech, Inc. · Prosthesis, Ankle, Semi-Constrained, Cemented, Metal/Polymer1 recall0 reports
- Vantage 350-22-14Exactech, Inc. · Prosthesis, Ankle, Semi-Constrained, Cemented, Metal/Polymer1 recall0 reports
- Vantage 350-22-21Exactech, Inc. · Prosthesis, Ankle, Semi-Constrained, Cemented, Metal/Polymer1 recall0 reports
- Vantage 350-22-22Exactech, Inc. · Prosthesis, Ankle, Semi-Constrained, Cemented, Metal/Polymer1 recall0 reports
- Vantage 350-22-24Exactech, Inc. · Prosthesis, Ankle, Semi-Constrained, Cemented, Metal/Polymer1 recall0 reports
- Vantage 350-21-41Exactech, Inc. · Prosthesis, Ankle, Semi-Constrained, Cemented, Metal/Polymer1 recall0 reports
- Vantage 350-21-42Exactech, Inc. · Prosthesis, Ankle, Semi-Constrained, Cemented, Metal/Polymer1 recall0 reports
- Vantage 350-21-43Exactech, Inc. · Prosthesis, Ankle, Semi-Constrained, Cemented, Metal/Polymer1 recall0 reports
- Vantage 350-21-44Exactech, Inc. · Prosthesis, Ankle, Semi-Constrained, Cemented, Metal/Polymer1 recall0 reports
- Vantage 350-22-41Exactech, Inc. · Prosthesis, Ankle, Semi-Constrained, Cemented, Metal/Polymer1 recall0 reports
- Vantage 350-22-42Exactech, Inc. · Prosthesis, Ankle, Semi-Constrained, Cemented, Metal/Polymer1 recall0 reports
- Vantage 350-21-61Exactech, Inc. · Prosthesis, Ankle, Semi-Constrained, Cemented, Metal/Polymer1 recall0 reports
- Vantage 350-21-62Exactech, Inc. · Prosthesis, Ankle, Semi-Constrained, Cemented, Metal/Polymer1 recall0 reports
- Vantage 350-21-63Exactech, Inc. · Prosthesis, Ankle, Semi-Constrained, Cemented, Metal/Polymer1 recall0 reports
- Vantage 350-21-64Exactech, Inc. · Prosthesis, Ankle, Semi-Constrained, Cemented, Metal/Polymer1 recall0 reports
- Vantage 350-22-62Exactech, Inc. · Prosthesis, Ankle, Semi-Constrained, Cemented, Metal/Polymer1 recall0 reports
- Vantage 350-22-63Exactech, Inc. · Prosthesis, Ankle, Semi-Constrained, Cemented, Metal/Polymer1 recall0 reports
Product lines
Other devices under the same FDA submission
21 devices cleared under K152217 are not named by identifier in this recall. One submission often covers many models, so they may or may not be affected.
- Vantage 350-12-03Exactech, Inc. · Prosthesis, Ankle, Semi-Constrained, Cemented, Metal/Polymer11 reports
- Vantage 351-07-04Exactech, Inc. · PROSTHESIS, ANKLE, SEMI-CONSTRAINED, CEMENTED, METAL/POLYMER10 reports
- Vantage 350-11-04Exactech, Inc. · Prosthesis, Ankle, Semi-Constrained, Cemented, Metal/Polymer5 reports
- Vantage 351-10-26Exactech, Inc. · PROSTHESIS, ANKLE, SEMI-CONSTRAINED, CEMENTED, METAL/POLYMER5 reports
- Vantage 350-11-01Exactech, Inc. · Prosthesis, Ankle, Semi-Constrained, Cemented, Metal/Polymer4 reports
- Vantage 351-17-02Exactech, Inc. · PROSTHESIS, ANKLE, SEMI-CONSTRAINED, CEMENTED, METAL/POLYMER4 reports
- Vantage 350-02-02Exactech, Inc. · Prosthesis, Ankle, Semi-Constrained, Cemented, Metal/Polymer3 reports
- Vantage 350-10-04Exactech, Inc. · Prosthesis, Ankle, Semi-Constrained, Cemented, Metal/Polymer3 reports
- Vantage 350-11-02Exactech, Inc. · Prosthesis, Ankle, Semi-Constrained, Cemented, Metal/Polymer2 reports
- Vantage 350-12-02Exactech, Inc. · Prosthesis, Ankle, Semi-Constrained, Cemented, Metal/Polymer2 reports
- Vantage 350-12-04Exactech, Inc. · Prosthesis, Ankle, Semi-Constrained, Cemented, Metal/Polymer2 reports
- Vantage 350-01-02Exactech, Inc. · Prosthesis, Ankle, Semi-Constrained, Cemented, Metal/Polymer2 reports
Other products in the same recall event
The FDA groups related products into one recall event (event 88570).
- Z-0019-2022 · Open, ClassifiedOPTETRAK Comprehensive Knee System, labeled as the following: a. OPTETRAK All-polyethylene CR Tibial Components, with Product Line Numbers: 200-11-XX, 200-12-XX, 200-13-XX, 200…
- Z-0020-2022 · Open, ClassifiedOPTETRAK Comprehensive Knee System including OPTETRAK All-polyethylene UNI Tibial Components, Product Lines, 252-12-XX, 252-13-XX, 252-22-XX, 252-23-XX. NOTE: Information refl…
- Z-0021-2022 · Open, ClassifiedOPTETRAK LOGIC Polyethylene Tibial Inserts, labeled as the following: a. OPTETRAK Logic PS Tibial Inserts, Product Line 02-012-35-XXXX; b. OPTETRAK Logic PSC Tibial Inser…
- Z-0023-2022 · Open, ClassifiedTRULIANT Knee System including TRULIANT Tibial Inserts, also labeled as the following: a. TRULIANT CR Tibial Inserts; b. TRULIANT CR Slope + Tibial Inserts; c. TRULIA…
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page