Skip to content

Vantage 350-22-01Recalls & reported problems

Exactech, Inc. · Class 2 · Implant

All models of Vantage (66)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
3
Deaths reported
0
Injuries reported
3
Recalls naming it
1

Summary

The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 10885862276339), received between Dec 20, 2022 and May 17, 2024 — classified as 3 injuries.

The most frequently reported specific device problem is “Naturally Worn” (3 reports).

1 FDA recall record name this device's identifier, 1 still open.

5 other recalls cover devices cleared under the same FDA submission (K152217); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 1 reportsMay 2024: 1 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Aug 30, 2021Open, Classified

    Inserts were packaged in vacuum bags that lacked an additional oxygen barrier layer.

    Z-0024-2022 · Root cause: Process control · Recall details · FDA record ↗

Most reported problems

Device problems

  • Naturally Worn3
  • Adverse Event Without Identified Device or Use Problem1

Patient problems

  • Failure of Implant3
  • Cyst(s)1
  • Implant Pain1
  • Pain1
  • Scar Tissue1

Recent adverse event reports

The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.

  • InjuryMay 17, 2024Naturally Worn; Naturally Worn

    IT WAS REPORTED VIA LEGAL DOCUMENTATION THAT APPROXIMATELY 37 MONTHS AFTER A RIGHT TOTAL ANKLE REPLACEMENT PROCEDURE, THE PATIENT UNDERWENT A REVISION PROCEDURE TO ADDRESS PROSTHESIS WEAR. NO FURTHER ISSUES OR COMPLICATIONS WERE REPORTED. NO ADDITIONAL INFORMATION IS AVAILABLE.

    Patient problems: Failure of Implant; Implant Pain; Pain

    MAUDE report 19341474 ↗
  • InjuryApr 25, 2024Naturally Worn; Naturally Worn

    AS REPORTED VIA LEGAL DOCUMENTATION, A PATIENT HAD A RIGHT ANKLE REPLACEMENT PROCEDURE ON (B)(6) 2019 AND THEN WAS REVISED ON (B)(6) 2024. THE PATIENT REQUIRED REVISION SURGERY FOR ISSUES INCLUDING BUT NOT LIMITED TO POLYETHYLENE PROSTHESIS WEAR. THE MOSTLY LIKELY CAUSE FOR THE REVISION REPORTED DUE TO PROSTHESIS WEAR IS A COMBINATION OF RISK FACTORS INCLUDING USE ERROR, IMPLANT POSITIONING, IMPLANT SIZE SELECTION, AND PATIENT FACTORS. HOWEVER, THIS CANNOT BE CONCLUSIVELY DETERMINED WITH THE INFORMATION PROVIDED.

    Patient problems: Cyst(s); Failure of Implant; Scar Tissue

    MAUDE report 19182308 ↗
  • InjuryDec 20, 2022Naturally Worn; Adverse Event Without Identified Device or Use Problem; Naturally Worn; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED VIA LEGAL THAT A 59 YO FEMALE PATIENT HAD AN RIGHT ANKLE REVISION PROCEDURE ON (B)(6) 2021 AND HAD A SECOND REVISION ON (B)(6) 2022. REVISION WAS NOTED TO BE FOR FLEXOR DIGITORUM LONGUS TEAR; MEDIAL GUTTER IMPINGEMENT; PERIPROSTHETIC CYST AT THE TALAR COMPONENT; INSTABILITY; DIFFICULTY GETTING TO NEUTRAL DORSIFLEXION; DIFFICULTY WALKING AND WEIGHT-BEARING; STIFFNESS, LOSS OF MOVEMENT AND FLEXIBILITY; PAIN AT THE POSTERIOR ASPECT, TIGHTNESS IN THE POSTERIOR ASPECT; SWELLING OF THE ANKLE. CURRENT REPORTED CONDITION OF PATIENT-RIGHT ANKLE PAIN AND LOSS OF RANGE OF MOTION, NON-WEIGHT BEARING, USE OF CRUTCHES, PAIN MEDICATION, BACK AND NECK PAIN DUE TO ABNORMAL GAIT. NO ADDITIONAL INFORMATION.

    Patient problems: Failure of Implant

    MAUDE report 16016689 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K152217. One submission often covers many models, so check the product and lot details in each record.

  • Z-0587-2024 · Nov 21, 2023 · Open, Classified
    Exactech Vantage Ankle System, Catalog Numbers: a) 350-21-04, b) 350-22-03
  • Z-2926-2020 · Aug 10, 2020 · Open, Classified
    The Vantage Total Ankle Tibial Insert, Fixed Bearing, Size 4, Right 10mm.
  • Z-2966-2018 · Jul 13, 2018 · Terminated
    Tibial Alignment Guide, REF 351-10-00 Vantage Fixed Ankle system components for ankle replacements.
  • Z-0200-2018 · May 16, 2017 · Terminated
    Vantage Total Ankle Tibial Insert, Fixed Bearing, 7 mm, Left, Size 3. Exactech Inc., Intended for patients with ankle joints damage by severe rheumatoid, post-traumatic, or dege
  • Z-0201-2018 · May 16, 2017 · Terminated
    Vantage Total Ankle Tibial Insert, Fixed Bearing, 10 mm, Left, Size 3. Exactech Inc. Intended for patients with ankle joints damage by severe rheumatoid, post-traumatic, or degene

Other Prosthesis, Ankle, Semi-Constrained, Cemented, Metal/Polymer devices

All Prosthesis, Ankle, Semi-Constrained, Cemented, Metal/Polymer devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page