NEUROSCOUT 601-314Recalls & reported problems
Cerenovus, Inc. · NEUROSCOUT 14 STEERABLE GUIDEWIRES Guidewire Dia. .014 Inch (.36mm) Usable Length 205cm 10cm Steerable Guidewire with Hydrophilic Coating Tip Flexibility Standard Contents: 1 guidewire, 1 torquing device and 1 introducer. · Class 2
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 3
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 10886704028376), received between Mar 29, 2019 and Apr 21, 2022 — classified as 1 injury, 2 malfunctions.
The most frequently reported specific device problem is “Material Separation” (1 report).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Most reported problems
Device problems
- Material Separation1
- Physical Resistance/Sticking1
- Unsealed Device Packaging1
Patient problems
- No Clinical Signs, Symptoms or Conditions1
- No Code Available1
- No Consequences Or Impact To Patient1
Recent adverse event reports
The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.
MalfunctionApr 21, 2022Unsealed Device Packaging; Unsealed Device Packaging
THE HEALTHCARE PROFESSIONAL REPORTED THAT DURING THE PROCEDURE, THE PHYSICIAN OPENED THE OUTER PACKAGE OF THE 0.014IN. X 205CM NEUROSCOUT 14 STANDARD GUIDEWIRE (601314 / JA2077) FOR INSPECTION AND FOUND THAT THE INNER STERILE PACKAGING WAS OPEN. THE DEVICE WAS NOT USED IN THE PATIENT. A NEW DEVICE WAS USED TO COMPLETE THE PROCEDURE. THERE WAS NO REPORT OF ANY PATIENT INJURY OR COMPLICATION. PHOTOS WERE INCLUDED IN THE COMPLAINT. THE PHOTOS ARE OF THE PRODUCT PACKAGING BOX AND PRODUCT POUCH.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 14177632 ↗InjurySep 14, 2020Material Separation; Material Separation
THE HEALTHCARE PROFESSIONAL REPORTED THAT DURING THE PROCEDURE TO TREAT AN INTRACRANIAL ARTERY STENOSIS IN THE PATIENT WHO WAS IMPLANTED WITH AN ENTERPRISE STENT (ABOUT TWO YEARS PRIOR), THE PHYSICIAN TRIED TO PASS THE 0.014 IN. X 205 CM NEUROSCOUT 14 STANDARD GUIDEWIRE (601314 / J6918N) THROUGH THE PREVIOUSLY IMPLANTED STENT, AND HE FELT A LITTLE FORCE, BUT HE THOUGHT IT WAS WITHIN ¿REASONABLE SCOPE.¿ AFTER ABOUT 6 CM OF THE GUIDEWIRE TIP CAME INTO THE STENT, THE GUIDEWIRE FRACTURED AND SEPARATED 13CM FROM ITS DISTAL TIP. THE DETACHED SEGMENT OF THE GUIDEWIRE EMBOLIZED IN THE MIDDLE CEREBRAL ARTERY (MCA). THIS PROMPTED THE USE OF A COMPETITOR STENT RETRIEVER TO REMOVE THE DETACHED SEGMENT OF THE GUIDEWIRE FROM THE PATIENT. THE EVENT RESULTED IN AN APPROXIMATELY ONE-HOUR PROCEDURAL DELAY. THERE WAS NO OBSERVABLE CLINICAL SYMPTOMS OR CHANGES IN SYMPTOMS IDENTIFIED IN THE PATIENT. THE PROCEDURE WAS COMPLETED WITH A COMPETITOR GUIDEWIRE. IMAGING FROM THE PROCEDURE WAS INCLUDED AND WAS FORWARDED TO INDEPENDENT PHYSICIAN FOR REVIEW ON (B)(6) 2020. ¿THE COMPLAINT FORM IS ACCOMPANIED BY A FLUOROSCOPIC IMAGE OF THE WIRE IN THE PATIENT AND A PHOTO OF THE WIRE FRAGMENT. THE PHYSICIAN REPORTS
Patient problems: No Code Available
MAUDE report 10532235 ↗MalfunctionMar 29, 2019Physical Resistance/Sticking; Physical Resistance/Sticking
AS REPORTED BY A HEALTHCARE PROFESSIONAL, DURING A COIL EMBOLIZATION OF A NON-TORTUOUS PERICALLOSAL ARTERY ANEURYSM, THE 6F ENVOY DA GUIDE CATHETER WAS POSITIONED AND THE PROWLER SELECT LPES (606S155FX/17680646) MICROCATHETER AND NEUROSCOUT 14 STD (601314/HU8939) GUIDEWIRE WERE INTRODUCED, BUT THE GUIDEWIRE BECAME IMPEDED IN THE BODY/SHAFT OF THE MICROCATHETER NEAR THE SITE OF THE ANEURYSM AND THE PHYSICIAN WAS UNABLE TO REMOVE THE GUIDEWIRE FROM THE MICROCATHETER. MANEUVERS WERE MADE IN AN ATTEMPT TO ¿DROP¿ THE GUIDEWIRE FROM INSIDE OF THE MICROCATHETER, BUT THE PHYSICIAN ULTIMATELY DECIDED TO WITHDRAW THE GUIDEWIRE AND MICROCATHETER FROM THE GUIDE CATHETER AS A UNIT. UPON REMOVAL, THE PHYSICIAN NOTICED THAT THE GUIDEWIRE WAS TRAPPED INSIDE OF THE MICROCATHETER AND THE DEVICES WERE ¿FIXED¿. THE GUIDEWIRE AND MICROCATHETER WERE REPLACED, AND THE PROCEDURE WAS COMPLETED SUCCESSFULLY WITH THE SAME ENVOY GUIDE CATHETER. NO PATIENT COMPLICATIONS OCCURRED AS A RESULT OF THE EVENT. THE LENGTH OF TIME THAT THE SURGERY WAS DELAYED DUE TO THE EVENT WAS NOT REPORTED; HOWEVER, THE DELAY WAS NOT CONSIDERED CLINICALLY SIGNIFICANT. THE DEVICES WERE STORED, PREPPED, AND FLUSHED ACCORDING TO THE I
Patient problems: No Consequences Or Impact To Patient
MAUDE report 8464256 ↗
Other GUIDE, WIRE, CATHETER, NEUROVASCULATURE devices
- Synchro SELECT SSTD215STRStryker Corporation · Guide, wire, catheter, neurovasculature1 recall97 reports
- Synchro SELECT SSTD215PREStryker Corporation · Guide, wire, catheter, neurovasculature1 recall16 reports
- Synchro SELECT SSUP215PREStryker Corporation · Guide, wire, catheter, neurovasculature1 recall11 reports
- Synchro2 S2SPS14300Stryker Corporation · GUIDE, WIRE, CATHETER, NEUROVASCULATURE1 recall1 death6 reports
- Synchro SELECT SSUP300STRStryker Corporation · Guide, wire, catheter, neurovasculature1 recall5 reports
- Synchro2 S2SPP14300Stryker Corporation · GUIDE, WIRE, CATHETER, NEUROVASCULATURE1 recall2 reports
- Synchro SELECT SSFT300STRStryker Corporation · Guide, wire, catheter, neurovasculature1 recall0 reports
- Mirage 103-0608Micro Therapeutics, Inc. · Guide, wire, catheter, neurovasculature12 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page