Skip to content

AGILITY 614-178Recalls & reported problems

Cerenovus, Inc. · AGILITY 10 STEERABLE GUIDEWIRES Guidewire Dia. .010 Inch (.25mm) Usable Length 195cm 10cm Steerable Guidewire with Hydrophilic Coating Tip Flexibility Standard Contents: 1 guidewire, 1 torquing device and 1 introducer. · Class 2

Adverse event reports
1
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 10886704029038), received between Jun 15, 2022 and Jun 15, 2022 — classified as 1 injury.

The most frequently reported specific device problem is “Material Separation” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Most reported problems

Device problems

  • Material Separation1

Patient problems

  • No Clinical Signs, Symptoms or Conditions1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • InjuryJun 15, 2022Material Separation; Material Separation

    IT WAS REPORTED BY A CONSUMER THAT A PATIENT UNDERWENT EMBOLIZATION OF A SPINAL DURAL ARTERIOVENOUS MALFORMATION (AVM) USING A 195 CM AGILITY .010 GUIDEWIRE (PRODUCT CODE: 614178, LOT UNKNOWN), WHICH FRACTURED DURING THE PROCEDURE AND WAS LEFT INSIDE THE PATIENT. THE WIRE WAS TRACKED OUT IN VERY TORTUOUS ANATOMY AND WHILE THE CATHETER WAS ADVANCED THE WIRE FRACTURED. THE SURGERY WAS DELAYED FOR MORE THAN 60 MINUTES. THE GUIDEWIRE WAS NOT ABLE TO BE REMOVED USING A SNARE, BUT THE PROCEDURE WAS COMPLETED SUCCESSFULLY. FRAGMENTS WERE GENERATED AND THEY WERE NOT EASILY REMOVED WITHOUT ADDITIONAL INTERVENTION. PER THE EVENT DESCRIPTION, THE PATIENT EXPERIENCED UNSPECIFIED CONSEQUENCES, BUT THERE WAS NO CLINICAL OUTCOME EXPERIENCED BY THE PATIENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 14696327 ↗

Other GUIDE, WIRE, CATHETER, NEUROVASCULATURE devices

All GUIDE, WIRE, CATHETER, NEUROVASCULATURE devices →

Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.