Skip to content

VERSALOK ORTHOCORD 210808Recalls & reported problems

Medos International Sàrl · VERSALOK ANCHOR w/ORTHOCORD PEEK & Titanium Anchor (1) Violet (1) Blue Strand, Size 2 (5 metric) ORTHOCORD braided composite suture, 36 Inches (91cm) · Class 2 · Implant

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Quick summary

This Medos VersaLok suture anchor appears in 92 adverse event reports, with no deaths and 10 injuries. Reports mostly describe anchors breaking, pulling out, or moving out of place during shoulder and other repairs, and parts that are hard to advance. Some patients were listed with tissue injury. No recalls name this device.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
89
Deaths reported
0
Injuries reported
10
Recalls naming it
0

Summary

The FDA's MAUDE database contains 89 adverse event reports that list this exact device (DI 10886705001224), received between Mar 14, 2018 and May 16, 2026 — classified as 10 injuries, 79 malfunctions.

The most frequently reported specific device problem is “Break” (24 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 1 reportsDec 2023: 2 reportsJan 2024: 6 reportsFeb 2024: 2 reportsMar 2024: 2 reportsApr 2024: 1 reportsMay 2024: 1 reportsJun 2024: 2 reportsJul 2024: 0 reportsAug 2024: 1 reportsSep 2024: 0 reportsOct 2024: 2 reportsNov 2024: 2 reportsDec 2024: 0 reportsJan 2025: 1 reportsFeb 2025: 1 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 1 reportsJun 2025: 0 reportsJul 2025: 1 reportsAug 2025: 2 reportsSep 2025: 0 reportsOct 2025: 2 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 1 reportsApr 2026: 0 reportsMay 2026: 1 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520266
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Break24
  • Migration18
  • Difficult to Advance15
  • Adverse Event Without Identified Device or Use Problem7
  • Device-Device Incompatibility6
  • Detachment of Device or Device Component3
  • Material Deformation3
  • Appropriate Device Problem Term/Code Not Available2

Patient problems

  • No Clinical Signs, Symptoms or Conditions54
  • No Consequences Or Impact To Patient22
  • Unspecified Tissue Injury6
  • Not Applicable3
  • Injury2
  • No Known Impact Or Consequence To Patient2
  • Foreign Body In Patient1
  • No Code Available1

Recent adverse event reports

The 25 most recent of 89 reports, in the reporter's own words as published by the FDA.

  • MalfunctionMay 16, 2026Difficult to Advance

    IT WAS REPORTED THAT DURING A SHOULDER ARTHROSCOPY PROCEDURE IT WAS OBSERVED THAT THE VERSALOK PRE PACKED W/OC ANCHOR DEVICE COULD NOT BE SECURED. IT WAS REPORTED THAT ANOTHER DEVICE WAS USED TO COMPLETE THE SURGERY. THERE WERE NO ADVERSE CONSEQUENCES TO THE PATIENT. NO ADDITIONAL INFORMATION COULD BE PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25195143 ↗
  • MalfunctionMar 11, 2026Break

    IT WAS REPORTED THAT DURING PATELLAR DISLOCATION SURGERY, THE VERSALOK PRE PACKED W/OC ANCHOR WAS BROKEN OFF. ALL THE BROKEN PARTS WERE REMOVED. ANOTHER DEVICE WAS USED TO COMPLETE THE SURGERY. THERE WERE NO ADVERSE CONSEQUENCES TO THE PATIENT. NO ADDITIONAL INFORMATION WAS PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24569490 ↗
  • MalfunctionOct 20, 2025Migration; Migration

    IT WAS REPORTED THAT DURING A SHOULDER SURGERY, IT WAS OBSERVED THAT THE VERSALOK PRE PACKED W/OC DEVICE ANCHOR PULLED OUT. ANOTHER DEVICE WAS USED TO COMPLETE THE SURGERY. THERE WERE NO ADVERSE CONSEQUENCES TO THE PATIENT. NO ADDITIONAL INFORMATION COULD BE PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23335123 ↗
  • MalfunctionOct 16, 2025Break; Break

    IT WAS REPORTED THAT DURING A SHOULDER ARTHROSCOPY SURGICAL PROCEDURE IT WAS OBSERVED THAT THE GRYPHON P BR DS ANCHOR W/OC DEVICE ANCHOR WAS BROKEN OFF. ALL BROKEN PARTS WERE REMOVED. ANOTHER LIKE DEVICE WAS USED TO COMPLETE THE PROCEDURE. THERE WERE NO REPORTS OF INJURIES, MEDICAL INTERVENTION OR PROLONGED HOSPITALIZATION. NO ADDITIONAL INFORMATION WAS PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23311216 ↗
  • MalfunctionAug 26, 2025Migration; Migration

    IT WAS REPORTED THAT DURING A ROTATOR CUFF REPAIR SURGICAL PROCEDURE IT WAS OBSERVED THAT THE VERSALOK ANCHOR W/ORTHOCORD PEEK DEVICE WAS PULLED OUT. ANOTHER DEVICE WAS USED TO COMPLETE THE SURGERY. THERE WAS A 10-MINUTE DELAY IN THE PROCEDURE DUE TO THE EVENT. THERE WERE NO ADVERSE CONSEQUENCES TO THE PATIENT. NO ADDITIONAL INFORMATION COULD BE PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22893736 ↗
  • MalfunctionAug 15, 2025Migration; Migration

    IT WAS REPORTED THAT DURING A PATELLAR DISLOCATION SURGICAL PROCEDURE IT WAS OBSERVED THAT THE VERSALOK PRE PACKED W/OC DEVICE ANCHOR WAS PULLED OUT. ANOTHER LIKE DEVICE WAS USED TO COMPLETE THE PROCEDURE SUCCESSFULLY. THERE WERE NO REPORTS OF INJURIES, MEDICAL INTERVENTION OR PROLONGED HOSPITALIZATION. NO ADDITIONAL INFORMATION WAS PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22816371 ↗
  • MalfunctionJul 25, 2025Migration; Migration

    IT WAS REPORTED THAT DURING AN UNSPECIFIED SURGICAL PROCEDURE, THE VERSALOK PRE PACKED DEVICE ANCHOR WAS PULLED OUT. ANOTHER DEVICE WAS USED TO COMPLETE THE SURGERY. THERE WAS A 5 MIN DELAY. THERE WERE NO ADVERSE CONSEQUENCES TO THE PATIENT. NO ADDITIONAL INFORMATION WAS PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22607793 ↗
  • MalfunctionMay 28, 2025Use of Device Problem; Use of Device Problem

    IT WAS REPORTED THAT DURING AN ANTERIOR CRUCIATE LIGAMENT SURGICAL PROCEDURE, THE VERSALOK PRE PACKED W/OC DEVICE WAS MISSING AN OVERLINE RING. ANOTHER DEVICE WAS USED TO COMPLETE THE SURGERY. THERE WERE NO ADVERSE CONSEQUENCES TO THE PATIENT. NO ADDITIONAL INFORMATION COULD BE PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22094833 ↗
  • MalfunctionFeb 20, 2025Break; Break

    IT WAS REPORTED THAT DURING A ROTATOR CUFF REPAIR SURGICAL PROCEDURE THE VERSALOK PRE PACKED W/OC DEVICE ANCHOR WAS BROKEN WHILE IMPLANTING THE LATERAL ROW ANCHOR. ANOTHER LIKE DEVICE WAS USED TO COMPLETE THE PROCEDURE. THERE WAS PATIENT INVOLVEMENT. THERE WERE NO ADVERSE PATIENT CONSEQUENCES REPORTED. THERE WERE NO REPORTS OF INJURIES, MEDICAL INTERVENTION OR PROLONGED HOSPITALIZATION NO ADDITIONAL INFORMATION WAS PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21426345 ↗
  • MalfunctionJan 7, 2025Break; Break

    IT WAS REPORTED THAT DURING A ROTATOR CUFF REPAIR SURGICAL PROCEDURE THE VERSALOK PRE PACKED W/OC DEVICE ANCHOR WAS BROKE OFF. ALL BROKEN PARTS WERE REMOVED. ANOTHER DEVICE WAS USED TO COMPLETE THE SURGERY. THERE WAS PATIENT INVOLVEMENT. THERE WERE NO REPORTS OF INJURIES, MEDICAL INTERVENTION OR PROLONGED HOSPITALIZATION. NO ADDITIONAL INFORMATION WAS PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21092653 ↗
  • MalfunctionNov 21, 2024Break; Break

    IT WAS REPORTED THAT DURING A ROTATOR CUFF REPAIR PROCEDURE, IT WAS OBSERVED THAT THE ANCHOR OF THE VERSALOK WITH ORTHOCORD DEVICE WAS BROKEN OFF. IT WAS REPORTED THAT ALL BROKEN PARTS WERE REMOVED. ANOTHER DEVICE WAS USED TO COMPLETE THE SURGERY. THERE WERE NO ADVERSE CONSEQUENCES TO THE PATIENT. NO ADDITIONAL INFORMATION COULD BE PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20747089 ↗
  • MalfunctionNov 18, 2024Break; Break

    IT WAS REPORTED FROM CHINA THAT DURING A ROTATOR CUFF REPAIR SURGICAL PROCEDURE IT WAS OBSERVED THAT THE VERSALOK PRE PACKED W/OC ANCHOR DEVICE WAS BROKEN OFF. IT WAS REPORTED THAT ALL PARTS WERE REMOVED FROM THE PATIENT. ANOTHER LIKE DEVICE WAS USED TO COMPLETE THE PROCEDURE. THERE WERE NO ADVERSE PATIENT CONSEQUENCES. NO ADDITIONAL INFORMATION WAS PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20712828 ↗
  • MalfunctionOct 22, 2024Migration; Migration

    IT WAS REPORTED BY THE SALES REP IN CHINA THAT DURING AN UNSPECIFIED SURGICAL PROCEDURE ON (B)(6) 2024 THE 4.5 HEALIX ADVANCE BR W/OCORD DEVICE ANCHOR BROKE OFF WHILE IN USE WITH THE VERSALOK PRE PACKED W/OC DEVICE. ALL BROKEN PARTS WERE REMOVED. THE ANCHOR WAS PULLED OUT. ANOTHER LIKE DEVICE WAS USED TO COMPLETE THE PROCEDURE. THERE WERE NO ADVERSE PATIENT CONSEQUENCES REPORTED. NO ADDITIONAL INFORMATION WAS PROVIDED. THIS IS REPORT 2 OF 2 OF THE SAME EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20509464 ↗
  • MalfunctionOct 14, 2024Break; Break

    IT WAS REPORTED BY A SALES REP IN CHINA THAT DURING A ROTATOR CUFF REPAIR PROCEDURE ON (B)(6) 2024, IT WAS OBSERVED THAT THE VERSALOK ANCHOR WITH ORTHOCORD DEVICE WAS BROKEN OFF. ANOTHER LIKE DEVICE WAS USED TO COMPLETE THE PROCEDURE. THERE WERE NO ADVERSE PATIENT CONSEQUENCES NOR SURGICAL DELAY REPORTED. NO ADDITIONAL INFORMATION WAS PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20437563 ↗
  • MalfunctionAug 5, 2024Break; Break

    IT WAS REPORTED BY THE SALES REP IN CHINA THAT DURING A ROTATOR CUFF REPAIR SURGICAL PROCEDURE ON 2024 THE VERSALOK PRE PACKED W/OC DEVICE WAS BROKEN AND THE OUTER SHEATH WAS RUPTURED. ANOTHER LIKE DEVICE WAS USED TO COMPLETE THE PROCEDURE. THERE WERE NO ADVERSE PATIENT CONSEQUENCES REPORTED. NO ADDITIONAL INFORMATION WAS PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19911621 ↗
  • MalfunctionJun 5, 2024Migration; Migration

    IT WAS REPORTED FROM CHINA THAT DURING A ROTATOR CUFF REPAIR PROCEDURE, IT WAS OBSERVED THAT THE ANCHOR ON THE VERSALOK PRE PACKED W/OC DEVICE PULLED OUT. ANOTHER LIKE DEVICE WAS USED TO COMPLETE THE PROCEDURE. THERE WERE NO ADVERSE CONSEQUENCES TO THE PATIENT NOR SURGICAL DELAY REPORTED. NO ADDITIONAL INFORMATION WAS PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19472322 ↗
  • MalfunctionJun 3, 2024Difficult to Advance; Difficult to Advance

    IT WAS REPORTED BY THE SALES REP IN CHINA THAT DURING A ROTATOR CUFF REPAIR SURGICAL PROCEDURE ON (B)(6)2024 THE VERSALOK PRE PACKED W/OC DEVICE COULD NOT SECURE THE ANCHOR. ANOTHER LIKE DEVICE WAS USED TO COMPLETE THE PROCEDURE. THERE WERE NO ADVERSE PATIENT CONSEQUENCES REPORTED. NO ADDITIONAL INFORMATION WAS PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19448710 ↗
  • MalfunctionMay 22, 2024Difficult to Advance; Difficult to Advance

    IT WAS REPORTED BY THE HEALTHCARE PROFESSIONAL IN CHINA THAT DURING A ROTATOR CUFF REPAIR SURGICAL PROCEDURE ON (B)(6) 2024 THE VERSALOK PRE PACKED W/OC DEVICE COULD NOT SECURE THE ANCHOR. ANOTHER LIKE DEVICE WAS USED TO COMPLETE THE PROCEDURE. THERE WAS NO DELAY IN THE PROCEDURE REPORTED. THERE WERE NO ADVERSE PATIENT CONSEQUENCES REPORTED. NO ADDITIONAL INFORMATION WAS PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19372320 ↗
  • MalfunctionApr 3, 2024Difficult to Advance; Difficult to Advance

    IT WAS REPORTED BY THE HEALTHCARE PROFESSIONAL IN CHINA THAT DURING A ROTATOR CUFF REPAIR PROCEDURE THE VERSALOK PRE PACKED W/OC DEVICE COULD NOT SECURE THE ANCHOR. ANOTHER LIKE DEVICE WAS USED TO COMPLETE THE PROCEDURE. THERE WAS NO DELAY IN THE PROCEDURE REPORTED. THERE WERE NO ADVERSE PATIENT CONSEQUENCES REPORTED. NO ADDITIONAL INFORMATION WAS PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19032827 ↗
  • MalfunctionMar 25, 2024Device-Device Incompatibility; Device-Device Incompatibility

    IT WAS REPORTED FROM CHINA THAT DURING A ROTATOR CUFF REPAIR SURGICAL PROCEDURE, IT WAS OBSERVED THAT THE VERSALOK PRE PACKED W/OC SCREW DEVICE COULD NOT FIRE. IT WAS REPORTED THAT ANOTHER LIKE DEVICE WAS USED TO COMPLETE THE PROCEDURE. THERE WERE NO ADVERSE CONSEQUENCES TO THE PATIENT. NO ADDITIONAL INFORMATION COULD BE PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18970580 ↗
  • MalfunctionMar 20, 2024Migration; Migration

    IT WAS REPORTED BY THE HEALTHCARE PROFESSIONAL IN CHINA THAT DURING A ROTATOR CUFF REPAIR PROCEDURE ON (B)(6) 2024 THE VERSALOK PRE PACKED W/OC DEVICE ANCHOR WAS PULLED OUT TOGETHER WHEN THE SHAFT WAS REMOVED. ANOTHER LIKE DEVICE WAS USED TO COMPLETE THE PROCEDURE. THERE WAS NO DELAY IN THE PROCEDURE REPORTED. THERE WERE NO ADVERSE PATIENT CONSEQUENCES REPORTED. NO ADDITIONAL INFORMATION WAS PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18939837 ↗
  • MalfunctionFeb 8, 2024Migration; Migration

    IT WAS REPORTED BY THE SALES REP IN CHINA THAT DURING A ROTATOR CUFF REPAIR SURGICAL PROCEDURE PERFORMED ON (B)(6) 2024 IT WAS OBSERVED THAT THE VERSALOK PRE PACKED W/OC ANCHOR DEVICE PULLED OUT. ANOTHER LIKE DEVICE WAS USED TO COMPLETE THE PROCEDURE. THERE WERE NO ADVERSE PATIENT CONSEQUENCES NOR SURGICAL DELAY REPORTED. NO ADDITIONAL INFORMATION WAS PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18672636 ↗
  • MalfunctionFeb 7, 2024Break; Break

    IT WAS REPORTED BY THE HEALTHCARE PROFESSIONAL IN CHINA THAT DURING AN ANTERIOR CRUCIATE LIGAMENT REPAIR PROCEDURE ON (B)(6) 2024, IT WAS OBSERVED THAT THE VERSALOK PRE PACKED W/OC DEVICE ANCHOR WAS BROKEN OFF, REMOVED ALL THE BROKEN PARTS. ANOTHER LIKE DEVICE WAS USED TO COMPLETE THE PROCEDURE. THERE WERE NO ADVERSE PATIENT CONSEQUENCES NOR SURGICAL DELAY REPORTED. NO ADDITIONAL INFORMATION WAS PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18657730 ↗
  • MalfunctionJan 29, 2024Difficult to Advance; Difficult to Advance

    IT WAS REPORTED BY THE SALES REP FROM CHINA THAT DURING A ROTATOR CUFF REPAIR SURGICAL PROCEDURE ON (B)(6) 2024 THE 3X VERSALOK PRE PACKED DEVICE COULD NOT SECURE THE ANCHOR. CHANGED ANOTHER TWO TO CONTINUE THE SURGERY, THE SAME PROBLEM HAPPENED AGAIN. ANOTHER DEVICE WAS USED TO COMPLETE THE SURGERY. THERE WERE NO ADVERSE CONSEQUENCES TO THE PATIENT. NO ADDITIONAL INFORMATION COULD BE PROVIDED. THIS IS REPORT 3 OF 3 OF THE SAME EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18594784 ↗
  • MalfunctionJan 26, 2024Difficult to Advance; Difficult to Advance

    IT WAS REPORTED BY THE SALES REP FROM CHINA THAT DURING A ROTATOR CUFF REPAIR SURGICAL PROCEDURE ON (B)(6) 2024 THE 3X VERSALOK PRE PACKED DEVICE COULD NOT SECURE THE ANCHOR. CHANGED ANOTHER TWO TO CONTINUE THE SURGERY, THE SAME PROBLEM HAPPENED AGAIN. ANOTHER DEVICE WAS USED TO COMPLETE THE SURGERY. THERE WERE NO ADVERSE CONSEQUENCES TO THE PATIENT. NO ADDITIONAL INFORMATION COULD BE PROVIDED. THIS IS REPORT 2 OF 3 OF THE SAME EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18587244 ↗

Other SUTURE, NONABSORBABLE, SYNTHETIC, POLYETHYLENE devices

All SUTURE, NONABSORBABLE, SYNTHETIC, POLYETHYLENE devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page