BRISTOW LATARJET 285270Recalls & reported problems
Depuy Mitek, LLC · Bristow - Latarjet Instability Shoulder System Curved Osteotome · Class 1
All models of BRISTOW LATARJET (15)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 12
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 12 adverse event reports that list this exact device (DI 10886705018000), received between Mar 13, 2020 and May 28, 2026 — classified as 12 malfunctions.
The most frequently reported specific device problem is “Break” (3 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Break3
- Degraded3
- Detachment of Device or Device Component2
- Use of Device Problem2
- Material Twisted/Bent1
- Unintended Movement1
Patient problems
- No Clinical Signs, Symptoms or Conditions10
- No Consequences Or Impact To Patient2
Recent adverse event reports
The 12 most recent of 12 reports, in the reporter's own words as published by the FDA.
MalfunctionMay 28, 2026Detachment of Device or Device Component
IT WAS REPORTED THAT DURING LOAN KIT PRODUCT INSPECTION IT WAS OBSERVED THAT THE LATARJET CURVED OSTEOTOME DEVICE WAS COMPOSED OF 1 PART. IT WAS OBSERVED THAT THE BOTTOM METAL PART OF THE DEVICE HAD COME OFF FROM THE HANDLE. THERE WAS NO PATIENT INVOLVEMENT. NO ADDITIONAL INFORMATION COULD BE PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25322395 ↗MalfunctionMar 25, 2024Break; Break
IT WAS REPORTED FROM AUSTRALIA THAT DURING A SHOULDER REPAIR SURGICAL PROCEDURE IT WAS OBSERVED THAT THE LATARJET CURVED OSTEOTOME DEVICE WAS BROKEN INTO TWO PIECES ON THE CONSIGNED LATARJET TRAY. IT WAS REPORTED THAT ANOTHER DEVICE WAS USED TO COMPLETE THE PROCEDURE. THERE WERE NO ADVERSE CONSEQUENCES TO THE PATIENT. NO ADDITIONAL INFORMATION COULD BE PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18970416 ↗MalfunctionMar 20, 2023Unintended Movement; Unintended Movement
IT WAS REPORTED BY A HEALTHCARE PROFESSIONAL IN AUSTRALIA THAT DURING AN UNKNOWN PROCEDURE ON (B)(6) 2023, IT WAS OBSERVED THAT THE HANDLE ON THE CURVED OSTEOTOME DEVICE APPEARED LOOSE. IT WAS UNKNOWN HOW THE PROCEDURE WAS COMPLETED. THERE WERE NO ADVERSE PATIENT CONSEQUENCES NOR SURGICAL DELAY REPORTED. NO ADDITIONAL INFORMATION WAS PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16574456 ↗MalfunctionOct 5, 2022Detachment of Device or Device Component; Detachment of Device or Device Component
IT WAS REPORTED BY THE SALES REP IN AUSTRALIA THAT DURING KIT INSPECTION, IT WAS OBSERVED THAT THE IMPACTING PLATE ON THE CURVED OSTEOTOME DEVICE HAD COME OFF FROM THE HANDLE ON THE DEVICE. THERE WAS NO PROCEDURE NOR PATIENT INVOLVEMENT REPORTED. NO ADDITIONAL INFORMATION WAS PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15545392 ↗MalfunctionSep 20, 2022Use of Device Problem; Use of Device Problem
IT WAS REPORTED BY A HEALTHCARE PROFESSIONAL IN AUSTRALIA THAT DURING A SHOULDER REPAIR PROCEDURE ON (B)(6) 2022, IT WAS OBSERVED THAT THE SILVER PIECE AT THE TOP OF THE CURVED OSTEOTOME DEVICE WAS MISSING. THERE WERE NO ADVERSE PATIENT CONSEQUENCES NOR SURGICAL DELAY REPORTED. NO ADDITIONAL INFORMATION WAS PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15457564 ↗MalfunctionJul 26, 2022Use of Device Problem; Use of Device Problem
IT WAS REPORTED BY THE SALES REP THAT DURING AN UNKNOWN PROCEDURE ON (B)(6) 2022, IT WAS OBSERVED THAT THE CURVED OSTEOTOME DEVICE WAS MISSING THE SILVER CAP ON THE BACK. ANOTHER LIKE DEVICE WAS USED TO COMPLETE THE PROCEDURE. THERE WERE NO ADVERSE PATIENT CONSEQUENCES NOR SURGICAL DELAY REPORTED. NO ADDITIONAL INFORMATION WAS PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15103340 ↗MalfunctionJun 6, 2022Break; Break
IT WAS REPORTED BY THE AFFILIATE IN (B)(6) THAT DURING AN UNKNOWN SURGERY ON (B)(6) 2022, IT WAS OBSERVED THAT THE CURVED OSTEOTOME DEVICE WAS BROKEN. THERE WERE NO ADVERSE PATIENT CONSEQUENCES NOR SURGICAL DELAY REPORTED. NO ADDITIONAL INFORMATION WAS PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 14605161 ↗MalfunctionJul 29, 2021Degraded; Degraded
IT WAS REPORTED BY THE AFFILIATE IN (B)(6). THAT DURING A SHOULDER REPAIR PROCEDURE ON (B)(6) 2021, IT WAS OBSERVED THAT THE LATARJET CURVED OSTEOTOME DEVICE WAS BLUNT. ANOTHER LIKE DEVICE WAS USED TO COMPLETE THE PROCEDURE WITH A DELAY OF ONE MINUTE. THERE WERE NO ADVERSE PATIENT CONSEQUENCES REPORTED. NO ADDITIONAL INFORMATION WAS PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12247077 ↗MalfunctionJul 13, 2021Degraded; Degraded
IT WAS REPORTED BY THE AFFILIATE IN (B)(6) THAT DURING A SHOULDER REPAIR PROCEDURE ON (B)(6 2021, IT WAS OBSERVED THAT THE LATARJET CURVED OSTEOTOME DEVICE WAS BLUNT. THE SAME DEVICE WAS USED TO COMPLETE THE PROCEDURE WITHOUT DELAY. THERE WERE NO ADVERSE PATIENT CONSEQUENCES REPORTED. NO ADDITIONAL INFORMATION WAS PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12160775 ↗MalfunctionMay 26, 2021Degraded; Degraded
IT WAS REPORTED BY THE SALES REP IN AUSTRALIA THAT DURING A LATARJET PROCEDURE ON (B)(6) 2021, IT WAS OBSERVED THAT THE CURVED OSTEOTOME WAS BLUNT. ANOTHER LIKE DEVICE WAS USED TO COMPLETE THE PROCEDURE. THERE WERE NO ADVERSE PATIENT CONSEQUENCES NOR SURGICAL DELAY REPORTED. NO ADDITIONAL INFORMATION WAS PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11886218 ↗MalfunctionSep 29, 2020Break; Break
IT WAS REPORTED VIA COMPLAINT SUBMISSION TOOL THAT THE LOAN KIT TECHNICIAN REPORTED THAT DURING LOAN KIT INSPECTION, WHEN IMPACTING HEAD/AREA OF A CURVED OSTEOTOME HANDLE HAS BROKEN OFF (NOT RETURNED). IT HAS BEEN FORWARDED TO DEPUY ENGINEERING FOR ASSESSMENT. NO FURTHER INFORMATION CAN BE OBTAINED AS THE CASE WAS NOT REPORTED BY THE CUSTOMER.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 10603516 ↗MalfunctionMar 13, 2020Material Twisted/Bent; Material Twisted/Bent
IT WAS REPORTED BY THE AFFILIATE VIA EMAIL THAT BEFORE A LATARJET PROCEDURE THE SCRUB NURSE NOTED THAT THE CUTTING TIP OF THE LATARJET CURVED OSTEOTOME WAS DAMAGED. NO PATIENT CONSEQUENCE AND NO SURGICAL DELAY WAS REPORTED. NO ADDITIONAL INFORMATION WAS PROVIDED.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 9832463 ↗
Other ORTHOPEDIC MANUAL SURGICAL INSTRUMENT devices
- AML EXCEL 2001-42-000Depuy (Ireland) · ORTHOPEDIC MANUAL SURGICAL INSTRUMENT2,009 reports
- NA 03.010.523Synthes GmbH · ORTHOPEDIC MANUAL SURGICAL INSTRUMENT1,208 reports
- EXPRESSEW 214141Medos International Sàrl · ORTHOPEDIC MANUAL SURGICAL INSTRUMENT1,166 reports
- N/A 9560524Medtronic Sofamor Danek, Inc. · ORTHOPEDIC MANUAL SURGICAL INSTRUMENT866 reports
- INHANCE 6500-11-104Depuy (Ireland) · Orthopedic manual surgical instrument831 reports
- INHANCE 6500-11-102Depuy (Ireland) · Orthopedic manual surgical instrument567 reports
- R3 71364450Smith & Nephew, Inc. · ORTHOPEDIC MANUAL SURGICAL INSTRUMENT520 reports
- N/A 9561524Medtronic Sofamor Danek, Inc. · ORTHOPEDIC MANUAL SURGICAL INSTRUMENT494 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page