BRISTOW LATARJETRecalls & reported problems across all models
Depuy Mitek, LLC · 15 models in the FDA device registry
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 50
- Deaths reported
- 0
- Injuries reported
- 3
- Recalls
- 0
We found 15 models of BRISTOW LATARJET by Depuy Mitek, LLC in the FDA's device registry (GUDID). Together they are named in 50 adverse event reports in the FDA's MAUDE database, received between Feb 15, 2018 and May 28, 2026 — classified as 3 injuries, 47 malfunctions.
The most frequently reported specific device problem is “Break” (15 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event. Widely used products collect more reports, so totals can't be used to compare products or estimate failure rates. Only reports that name a model's exact device identifier are counted. What reports can and can't tell you · How this data works
Reports received per month, all models
Most reported BRISTOW LATARJET problems
Device problems
- Break15
- Dull, Blunt8
- Degraded6
- Material Deformation6
- Material Twisted/Bent4
- Migration or Expulsion of Device3
- Appropriate Device Problem Term/Code Not Available2
- Detachment of Device or Device Component2
Patient problems
- No Consequences Or Impact To Patient24
- No Clinical Signs, Symptoms or Conditions21
- Not Applicable5
- Pain2
- No Patient Involvement1
- Unspecified Tissue Injury1
BRISTOW LATARJET models
15 models, most reported first. Each model has its own page with its reports, recalls and device details.
- BRISTOW LATARJET 285270ORTHOPEDIC MANUAL SURGICAL INSTRUMENT12 reports
- BRISTOW LATARJET 285240ORTHOPEDIC MANUAL SURGICAL INSTRUMENT11 reports
- BRISTOW LATARJET 285220ORTHOPEDIC MANUAL SURGICAL INSTRUMENT6 reports
- BRISTOW LATARJET 285143BIT, DRILL5 reports
- BRISTOW LATARJET 285255ORTHOPEDIC MANUAL SURGICAL INSTRUMENT2 reports
- BRISTOW LATARJET 285295ORTHOPEDIC MANUAL SURGICAL INSTRUMENT2 reports
- BRISTOW LATARJET 285112Screw, fixation, bone2 reports
- BRISTOW LATARJET 285122Screw, fixation, bone2 reports
- BRISTOW LATARJET 285126Screw, fixation, bone2 reports
- BRISTOW LATARJET 285245ORTHOPEDIC MANUAL SURGICAL INSTRUMENT1 report
- BRISTOW LATARJET 285260ORTHOPEDIC MANUAL SURGICAL INSTRUMENT1 report
- BRISTOW LATARJET 285325ORTHOPEDIC MANUAL SURGICAL INSTRUMENT1 report
- BRISTOW LATARJET 285111Screw, fixation, bone1 report
- BRISTOW LATARJET 285121Screw, fixation, bone1 report
- BRISTOW LATARJET 285142Screw, fixation, bone1 report
Recent BRISTOW LATARJET adverse event reports
The 15 most recent reports across all models, in the reporter's own words as published by the FDA.
MalfunctionMay 28, 2026Detachment of Device or Device Component
IT WAS REPORTED THAT DURING LOAN KIT PRODUCT INSPECTION IT WAS OBSERVED THAT THE LATARJET CURVED OSTEOTOME DEVICE WAS COMPOSED OF 1 PART. IT WAS OBSERVED THAT THE BOTTOM METAL PART OF THE DEVICE HAD COME OFF FROM THE HANDLE. THERE WAS NO PATIENT INVOLVEMENT. NO ADDITIONAL INFORMATION COULD BE PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25322395 ↗MalfunctionJan 16, 2025Degraded; Degraded
IT WAS REPORTED THAT DURING A BONY BANKART SHOULDER REPAIR PROCEDURE IT WAS OBSERVED THAT THE LATARJET CC STEP DRILL 3.2MM DEVICE WAS DULL. ANOTHER DEVICE WAS USED TO COMPLETE THE PROCEDURE. THERE WERE NO ADVERSE CONSEQUENCES TO THE PATIENT, OR SURGICAL DELAY. NO ADDITIONAL INFORMATION COULD BE PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21172970 ↗MalfunctionMar 25, 2024Break; Break
IT WAS REPORTED FROM AUSTRALIA THAT DURING A SHOULDER REPAIR SURGICAL PROCEDURE IT WAS OBSERVED THAT THE LATARJET CURVED OSTEOTOME DEVICE WAS BROKEN INTO TWO PIECES ON THE CONSIGNED LATARJET TRAY. IT WAS REPORTED THAT ANOTHER DEVICE WAS USED TO COMPLETE THE PROCEDURE. THERE WERE NO ADVERSE CONSEQUENCES TO THE PATIENT. NO ADDITIONAL INFORMATION COULD BE PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18970416 ↗MalfunctionOct 4, 2023Break; Break
IT WAS REPORTED BY THE HEALTHCARE PROFESSIONAL IN SWITZERLAND THAT DURING AN UNKNOWN PROCEDURE OF THE SHOULDER ON (B)(6) 2023, IT WAS OBSERVED THAT THE BRISTOW-LATARJET GLENOID 2.9 MM DRILL BIT DEVICE BROKE WHILE TRYING TO DRILL A HOLE IN THE GLENOID. IT WAS UNKNOWN IF AND HOW THE PROCEDURE WAS COMPLETED. THERE WAS A 45 MINUTE DELAY. THERE WERE NO ADVERSE PATIENT CONSEQUENCES REPORTED. NO ADDITIONAL INFORMATION WAS PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17870888 ↗MalfunctionMar 20, 2023Unintended Movement; Unintended Movement
IT WAS REPORTED BY A HEALTHCARE PROFESSIONAL IN AUSTRALIA THAT DURING AN UNKNOWN PROCEDURE ON (B)(6) 2023, IT WAS OBSERVED THAT THE HANDLE ON THE CURVED OSTEOTOME DEVICE APPEARED LOOSE. IT WAS UNKNOWN HOW THE PROCEDURE WAS COMPLETED. THERE WERE NO ADVERSE PATIENT CONSEQUENCES NOR SURGICAL DELAY REPORTED. NO ADDITIONAL INFORMATION WAS PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16574456 ↗MalfunctionNov 14, 2022Material Deformation; Material Deformation
IT WAS REPORTED BY A HEALTHCARE PROFESSIONAL IN AUSTRALIA THAT DURING A SHOULDER REPAIR PROCEDURE ON (B)(6) 2022, IT WAS OBSERVED THAT THE HEXAGONAL END ON THE 2.5MM CANNULATED SCREWDRIVER DEVICE WAS BURRED ROUND AND SPINNING IN THE GLENOID SCREW. ACCORDING TO THE REPORT, THE DEVICE WAS INEFFECTIVE IN TURNING THE SCREW. DURING IN-HOUSE ENGINEERING EVALUATION, IT WAS DETERMINED THAT THE DEVICE WAS DEFORMED ON THE DISTAL PART OF THE TIP. IT WAS REPORTED THAT THE GUIDE WIRE WAS REMOVED AND TIGHTENED SCREW WITH A NON-CANNULATED SCREWDRIVER TO COMPLETE THE PROCEDURE WITH A DELAY OF FIVE MINUTES. THERE WERE NO ADVERSE PATIENT CONSEQUENCES REPORTED. NO ADDITIONAL INFORMATION WAS PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15791025 ↗MalfunctionOct 5, 2022Detachment of Device or Device Component; Detachment of Device or Device Component
IT WAS REPORTED BY THE SALES REP IN AUSTRALIA THAT DURING KIT INSPECTION, IT WAS OBSERVED THAT THE IMPACTING PLATE ON THE CURVED OSTEOTOME DEVICE HAD COME OFF FROM THE HANDLE ON THE DEVICE. THERE WAS NO PROCEDURE NOR PATIENT INVOLVEMENT REPORTED. NO ADDITIONAL INFORMATION WAS PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15545392 ↗MalfunctionSep 20, 2022Use of Device Problem; Use of Device Problem
IT WAS REPORTED BY A HEALTHCARE PROFESSIONAL IN AUSTRALIA THAT DURING A SHOULDER REPAIR PROCEDURE ON (B)(6) 2022, IT WAS OBSERVED THAT THE SILVER PIECE AT THE TOP OF THE CURVED OSTEOTOME DEVICE WAS MISSING. THERE WERE NO ADVERSE PATIENT CONSEQUENCES NOR SURGICAL DELAY REPORTED. NO ADDITIONAL INFORMATION WAS PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15457564 ↗MalfunctionJul 26, 2022Use of Device Problem; Use of Device Problem
IT WAS REPORTED BY THE SALES REP THAT DURING AN UNKNOWN PROCEDURE ON (B)(6) 2022, IT WAS OBSERVED THAT THE CURVED OSTEOTOME DEVICE WAS MISSING THE SILVER CAP ON THE BACK. ANOTHER LIKE DEVICE WAS USED TO COMPLETE THE PROCEDURE. THERE WERE NO ADVERSE PATIENT CONSEQUENCES NOR SURGICAL DELAY REPORTED. NO ADDITIONAL INFORMATION WAS PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15103340 ↗MalfunctionJun 6, 2022Break; Break
IT WAS REPORTED BY THE AFFILIATE IN (B)(6) THAT DURING AN UNKNOWN SURGERY ON (B)(6) 2022, IT WAS OBSERVED THAT THE CURVED OSTEOTOME DEVICE WAS BROKEN. THERE WERE NO ADVERSE PATIENT CONSEQUENCES NOR SURGICAL DELAY REPORTED. NO ADDITIONAL INFORMATION WAS PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 14605161 ↗MalfunctionNov 7, 2021Appropriate Device Problem Term/Code Not Available; Appropriate Device Problem Term/Code Not Available
IT WAS REPORTED BY THE AFFILIATE IN (B)(6) THAT THE LATARJET CC STEP DRILL 3.2MM DEVICE HAD AN UNSPECIFIED MALFUNCTION. DURING IN-HOUSE ENGINEERING EVALUATION, IT WAS DETERMINED THAT THE EDGES ON THE DEVICE WERE DULL UNDER MAGNIFICATION. THERE WAS NO PROCEDURE NOR PATIENT INVOLVEMENT REPORTED. NO ADDITIONAL INFORMATION WAS PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12758970 ↗MalfunctionOct 15, 2021Material Deformation; Material Deformation
IT WAS REPORTED BY THE SALES REP THAT DURING AN UNKNOWN SURGERY ON (B)(6) 2021, IT WAS OBSERVED THAT THE LATARJET CANN. SCREWDRIVER DEVICE WAS STRIPPED. DURING IN-HOUSE ENGINEERING EVALUATION, IT WAS DETERMINED THAT THE DISTAL TIP ON THE DEVICE WAS DEFORMED. ANOTHER LIKE DEVICE WAS USED TO COMPLETE THE PROCEDURE. THERE WERE NO ADVERSE PATIENT CONSEQUENCES NOR SURGICAL DELAY REPORTED. NO ADDITIONAL INFORMATION WAS PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12638982 ↗MalfunctionJul 29, 2021Degraded; Degraded
IT WAS REPORTED BY THE AFFILIATE IN (B)(6). THAT DURING A SHOULDER REPAIR PROCEDURE ON (B)(6) 2021, IT WAS OBSERVED THAT THE LATARJET CURVED OSTEOTOME DEVICE WAS BLUNT. ANOTHER LIKE DEVICE WAS USED TO COMPLETE THE PROCEDURE WITH A DELAY OF ONE MINUTE. THERE WERE NO ADVERSE PATIENT CONSEQUENCES REPORTED. NO ADDITIONAL INFORMATION WAS PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12247077 ↗MalfunctionJul 19, 2021Material Twisted/Bent; Material Twisted/Bent
IT WAS REPORTED FROM THE AFFILIATE THAT THE LATARJET SOLID SCREWDRIVER DEIVICE BOLTS THAT HELD THE DRIVER SHAFTS IN THE HANDLES CAME LOOSE OR HELD DEBRIS UNDER THEM. DURING AN IN-HOUSE ENGINEERING EVALUATION, IT WAS DETERMINED THAT THE DISTAL TIP OF THE SCREWDRIVER WAS DEFORMED AND BENT. IT WAS NOT REPORTED IF THE DEVICE WAS USED IN SURGERY, OR IF THERE WAS PATIENT INVOLVEMENT. IT WAS NOT REPORTED IF THERE WERE ANY DELAYS IN THE PROCEDURE OR WHETHER A LIKE DEVICE WAS AVAILABLE FOR USE. THERE WERE NO ADVERSE PATIENT CONSEQUENCES REPORTED. NO ADDITIONAL INFORMATION WAS PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12191739 ↗MalfunctionJul 19, 2021Corroded; Corroded
IT WAS REPORTED FROM THE AFFILIATE THAT THE LATARJET TOP HAT SCREW DRIVER DEVICE BOLTS THAT HELD THE DRIVER SHAFT IN THE HANDLES CAME LOOSE AND HELD DEBRIS UNDER THEM. DURING AN IN-HOUSE ENGINEERING EVALUATION, IT WAS DETERMINED THAT THE DEVICE HAD RUST ON THE ETCH OF THE DEVICE. IT WAS NOT REPORTED IF THE DEVICE WAS USED IN SURGERY, OR IF THERE WAS PATIENT INVOLVEMENT. IT WAS NOT REPORTED IF THERE WERE ANY DELAYS IN THE PROCEDURE OR WHETHER A LIKE DEVICE WAS AVAILABLE FOR USE. THERE WERE NO ADVERSE PATIENT CONSEQUENCES REPORTED. NO ADDITIONAL INFORMATION WAS PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12191665 ↗
Other Depuy Mitek, LLC products
- VAPR30 models · 1,986 reports
- HEALIX ADVANCE6 models · 873 reports · 2 recalls
- VAPR COOLPULSE2 models · 779 reports
- FMS9 models · 434 reports
- BIOKNOTLESS PANACRYL ETHIBOND3 models · 35 reports · 2 recalls
- LUPINE ORTHOCORD4 models · 112 reports
- QUICKANCHOR ETHIBOND TAPERCUT2 models · 104 reports
- INTRAFIX14 models · 91 reports
Sources and updates
- Models: FDA GUDID, file dated Oct 13, 2026. Models are grouped by the brand name the manufacturer registered.
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page