BRISTOW LATARJET 285122Recalls & reported problems
Depuy Mitek, LLC · Bristow - Latarjet Instability Shoulder Screw 40mm Titanium Screw & Top Hat; Quanitity: (1) Screw; (1) Top Hat · Class 2 · Implant
All models of BRISTOW LATARJET (15)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 2
- Deaths reported
- 0
- Injuries reported
- 2
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 10886705020676), received between Dec 10, 2018 and Dec 10, 2018 — classified as 2 injuries.
The most frequently reported specific device problem is “Migration or Expulsion of Device” (2 reports).
5 other recalls cover devices cleared under the same FDA submission (K110763); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Most reported problems
Device problems
- Migration or Expulsion of Device2
Patient problems
- Pain2
Recent adverse event reports
The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.
InjuryDec 10, 2018Migration or Expulsion of Device; Migration or Expulsion of Device
THIS IS REPORT 1 OF 2 FOR THE SAME EVENT. IT WAS REPORTED BY THE AFFILIATE IN (B)(6) THAT THERE WAS CONVERSION OF LATARJET PROCEDURE TO REVERSE SHOULDER ARTHROPLASTY. REVISION PERFORMED BECAUSE OF PAIN AND THE GLENOID SCREWS HAD PENETRATED ARTICULATE SURFACE AND AXILLARY NERVE PALSY PERSISTING. PATIENT HAD MULTIPLE SHOULDER SURGERIES PREVIOUSLY PRIOR TO LATARJET PROCEDURE, SUBSEQUENTLY EXPERIENCED PERSISTENT AXILLARY NERVE PALSY POST-OPERATIVELY, AND PAIN IN SHOULDER JOINT. SHOULDER JOINT BECAME UNSTABLE. LATARJET SCREWS AND TOP HATS (X 2 EACH) WERE REMOVED AND DISCARDED. DELTA XTEND RSA WAS IMPLANTED. IT WAS REPORTED THAT THE PATIENT'S BONE QUALITY WAS QUITE GOOD. IT WAS REPORTED THAT THE SCREWS WERE WELL SEATED IN BONE AND REQUIRED SOME EFFORT TO REMOVE. IT WAS REPORTED THAT THE SURGEON WAS NOT OFF-AXIS. IT WAS REPORTED THAT THE PATIENT WAS NOT YOUNG AND EXPERIENCED OA IN HER SHOULDER JOINT QUITE QUICKLY DUE TO ABNORMAL JOINT MOVEMENT, SECONDARY TO THE AXILLARY NERVE PALSY. IT WAS REPORTED THAT THE GLENOID SCREW/S BECAME EXPOSED. IT WAS REPORTED THAT THE SCREWS WERE REMOVED, AND A PATIENT HAD A SHOULDER REPLACEMENT PERFORMED. IT WAS REPORTED THAT ALL METAL WORK REMOVED. IT WAS NO
Patient problems: Pain
MAUDE report 8148350 ↗InjuryDec 10, 2018Migration or Expulsion of Device; Migration or Expulsion of Device
ADDITIONAL INFORMATION RECEIVED FROM THE AFFILIATE ON 12/04/2018 REPORTING THAT PREVIOUS SURGERY WAS PERFORMED ON (B)(6) 2016.
Patient problems: Pain
MAUDE report 8148343 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K110763. One submission often covers many models, so check the product and lot details in each record.
- Z-2616-2017 · May 12, 2017 · TerminatedDePuy Mitek LATARJET EXPERIENCE Coracoid Drill Product Code: 288201 Product Usage: The Latarjet Cortical Screw set is intended to treat recurrent shoulder instability by supporting
- Z-2618-2017 · May 12, 2017 · TerminatedDePuy Mitek LATARJET EXPERIENCE Top Hat Tap Product Code: 288203 Product Usage: The Latarjet Cortical Screw set is intended to treat recurrent shoulder instability by supporting t
- Z-2615-2017 · May 12, 2017 · TerminatedDePuy Mitek LATARJET EXPERIENCE-Combo Screw Driver Product Code: 288211 Product Usage: The Latarjet Cortical Screw set is intended to treat recurrent shoulder instability by suppo
- Z-2619-2017 · May 12, 2017 · TerminatedDePuy Mitek LATARJET EXPERIENCE Extraction Driver Product Code: 288216 Product Usage: The Latarjet Cortical Screw set is intended to treat recurrent shoulder instability by suppor
- Z-2617-2017 · May 12, 2017 · TerminatedDePuy Mitek LATARJET EXPERIENCE Coracoid Top Hat Drill Product Code: 288202 Product Usage: The Latarjet Cortical Screw set is intended to treat recurrent shoulder instability by su
Other Screw, fixation, bone devices
- NA 03.043.029Synthes GmbH · Screw, fixation, bone503 reports
- OSTEORAPTOR 72201995Smith & Nephew, Inc. · Screw, fixation, bone253 reports
- Control 360 SK12Treace Medical Concepts, Inc. · Screw, Fixation, Bone195 reports
- HEALIX PEEK ORTHOCORD 222205Medos International Sàrl · Screw, fixation, bone186 reports
- PRECICE® Bone Transport System STP3.5-400Nuvasive Specialized Orthopedics, Inc. · Screw, Fixation, Bone3 recalls0 reports
- PRECICE® Bone Transport System STP3.5-225Nuvasive Specialized Orthopedics, Inc. · Screw, Fixation, Bone3 recalls0 reports
- PRECICE® Bone Transport System STP3.5-350Nuvasive Specialized Orthopedics, Inc. · Screw, Fixation, Bone3 recalls0 reports
- PRECICE® Bone Transport System STP3.5-300Nuvasive Specialized Orthopedics, Inc. · Screw, Fixation, Bone3 recalls0 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page